Page 1
GG2 COMMUNICATIONS 4-20mA / IMAC / RS485 User Manual Version: 1, July 2020 Designed and manufactured in Australia by Ampcontrol Pty Ltd...
Page 2
The note symbol highlights key information. Please share these notes with other operators. ENVIRO The enviro (environmental) symbol highlights areas which may have an impact on the surrounding fauna and/or flora. Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 2 of 22...
Page 3
GG2B032 Version 1 – JULY/2020 Copyright Notice The Ampcontrol GG2 Display described in this document is the property of AMPCONTROL PTY LTD. It is furnished under a license agreement and is to be used only in accordance with the terms of the agreement.
(using ISO/IEC 17025), testing laboratories (using ISO/IEC 17025), medical testing laboratories (using ISO 15189), inspection bodies (using ISO/IEC 17020) and proficiency testing providers using ISO/IEC 17043. Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 5 of 22...
The GG2 Display User Manual is expected to be read in conjunction with the following documentation: GG2B011 GG2 Detector User Manual GG2B023 GG2 Display User Manual GG2B033 GG2 Design, Installation and Maintenance Application Guide Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 6 of 22...
If there is evidence of physical damage, notify Ampcontrol immediately. Notify Ampcontrol immediately in the case of any discrepancies to the packing list. Keep a record of any claims and correspondence. Photographs are recommended.
This bit is set when it Power traverses the setpoint supply < 1 = Trip i.e. typical used for Tripping Setpoint B in RTGMS / Gas monitoring (trip) SLP codes. Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 9 of 22...
Page 10
A fault which Hardware 1 = Hardware impedes/prevents gas fault fault detection/other critical function has been detected. Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 10 of 22...
Page 11
0 - 24 VDC (LSB GG2 Mode : 0 - 240 for AIM mode) 7 - 0 (0.0 - 24.0 V) 0 - 24 VDC (GG2 mode) Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 11 of 22...
Setpoint 1 (Gas) 4000 - 20000 fault i.e. loss of analogue loop or negative drift alarm Setpoint 2 (Gas) 4000 - 20000 RTGMS used for Warn Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 12 of 22...
Page 13
(excluding the last sys clock data bus at this address (and address of each data +1), so that the GG2 can use it block, which would Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 13 of 22...
Page 14
Result data (from be modified in this step) module): Bit 7 : 0 = Complete without error Bit 6 - 0 : Future use (command error codes) Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 14 of 22...
Page 15
Bit 4: 1 = Yes applied (for > 20s) Bit 3: 14 days until Linearity 1 = Yes test due The 6 monthly calibration check is due. Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 15 of 22...
Page 16
32 bits (for calibration Detector Serial number tracking) low word Future use MSB : Future use Detector Software LSB : 8 bit version Version number (unsigned) Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 16 of 22...
10500 20800 HIGH WARMUP 20.8 < x <= 21.1 10625 21000 HIGH SOFT FAULT 21.2 21.1 < x <=21.4 10750 21200 HARD FAULT 21.4 < x Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 17 of 22...
6 – bit 5 days) 6 – bit 6 Detector present 1 = detector present 6 – bit 7 4-20mA loop activation 1 = 4-20mA loop active Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 18 of 22...
Page 19
FACTORY USE ONLY Detector pressure Hectopascals (hPa) Signed Int Detector temperature Degrees Celsius(°C) * Signed Int Detector humidity Permyriad Signed Int Detector bus voltage Millivolts Signed Int Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 19 of 22...
Page 20
4: healthy ≥ SP3 RTC date and time Time in seconds since 01-01-2000 00:00 32 bit All zeros if time lost or Unsigned Int RTC date and time faulty Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 20 of 22...
Page 21
RESERVED Detector Last linearity test Date (in Ampcontrol date 32 bit Time in seconds since format) Unsigned Int 01-01-2000 00:00 Detector Last linearity test Date Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 21 of 22...
Page 22
Unsigned Int in bypass bypass Display bus voltage Millivolts (mV) Signed Int µA. Intended 4-20mA 4-20mA conversion value Unsigned Int analog signal value Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version Page 22 of 22...