Thank you for your assistance. For your convenience, please use your mobile phone to scan the QR code on the right. 3 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Service action required The relevant problem cannot be solved by the user. Attention is required by an engineer suitably qualified with this equipment. 5 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
This disclaimer also applies if the system installation does not comply with the requirements of the installation drawings and other specifications that may be stated by the manufacturers. 6 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Training should be organized and monitored so that evidence of understanding is demonstrated by the trainee. * (INTERNATIONALLY HARMONISED) standard 61010-1-1-ISSUE 3 section 5.4.101.1 and Pressure Equipment Directive. P.E.D 7 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Any technical specifications given on the rating plate or installation drawing overrules any information or suggested values or parameters that may be given as an example in this and associated manuals. 9 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
If the safety valve does not operate at the stated pressure it must be checked and tested by a qualified engineer. 10 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Astell Scientific and approved agents are able to offer advice and assistance in setting up and commissioning this steriliser. Please consult Astell Scientific or an approved agent if necessary. 11 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
If handle is rotated to locked position with door open, press [door] button and return handle to “unlocked” position. 12 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
When a screen shows a button, then a press on the button will select that item. Typical view of controller in standby mode 13 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
In Standby/Ready state - with door closed- the system displays the current measured variables and the state of switches, interlocks and services. Any failures are reported as ALARM conditions. 14 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Alphanumeric keyboard Menu button Takes you to the next menu. Return button Return from selected screens back to the start screen. 15 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
To restore normal operation Press the [DOOR] button to unlock Then rotate the handle to the unlocked position. 16 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
The display changes to show that a cycle has been your machine. selected and loaded 17 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Type the value you want using the numeric keys, then press . When you are happy with the date and time shown press the clock button to save the changes. 18 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
For all loads involving pathogens, the load • Press button to start cycle must be sterilized again before opening. • The system starts the cycle process 19 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 20
If the sterilize stage has been set to “sterilize retry” then then you cannot open the door but you can restart the cycle. 20 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
• Type your password. Enter 6 characters followed by “CONTINUE” Note: If you enter the incorrect password you are returned to the Main Menu. 21 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Developers can change all codes. • Log in as a supervisor or higher using an appropriate password • Touch the password bar 22 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 23
• Touch a blank line on the screen • Enter the new password for that new user • Confirm the new password when asked • Press “CONTINUE” 23 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
By pressing the button. The next screen will allow you to set up a new users details. 24 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 25
Note: Only a user with a role of ‘Administrator’ is permitted to delete another user. 25 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 26
If the ‘Export after each cycle’ is not selected. A ‘Printer Archive’ password protected PDF file will be automatically generated each day at 16:00 hrs. 26 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 27
User login and logout attempts. Any value that has been changed and by which User. The time and date that an event occurred. 27 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
• Stage changes will result in a “failed” cycle MANUALLY JUMP STAGES At any point during the cycle you can press the MENU button 28 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
(the same temperature as appears on the temperature display). 29 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Control of temperature does not rely on steam pressure. The actual temperature that the control system attempts to attain is varied automatically at different parts of the 30 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 31
Health and safety requirements in the EC require that generally such a load is not handled at a temperature above 80°C. 31 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
The filling level is correct when the tip of the water level probe dips about 3 mm under the water surface 32 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
97-103°C depending on altitude and the boiling point of water. After free steam time ends, the system goes to heat to sterilize stage. 33 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
The stage continues for set sterilisation time. Sterilise temperature - Set sterilization temperature Sterilisation time - Set sterilization time 34 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
• Then press continue • When you are happy with the date and time shown press the time set button to save the changes 35 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 38
E.g. the chamber might be above 100°C but the pressure might be zero or much lower than normal. Service attention will be required. 38 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
This may be too stiff or solid to use for some loads. Alternative probes may be available. Please consult service department or your dealer. 39 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
The printer is mounted in the top cover. Feed the cut edge of paper into the mechanism and operate the paper feed button 40 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 41
The ribbon should run neatly in the slot provided. If the ribbon is distorted and above the slot, remove, retention the ribbon and replace • Re-load the paper 41 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 42
Service calls during the warranty periods which (in view of the manufacturer or agent) are caused by the user and/or these reasons may incur service charges. 42 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
To comply with HSE requirements, instructions regarding testing of these interlocks must be permanently displayed close to the sterilizer operator position 44 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Hold the tool at the angle shown. Insert it carefully into the hole, holding it straight so that it points toward the front edge of the door. 45 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 46
• You can either place the handle loosely on the end of the square shaft or use a spanner to turn the square shaft. 46 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
• Door interlock tests Carry out the door interlock test described on previous pages. If necessary lubricate locking pins with disulphide high-temp grease. 47 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Difficult to close door against gasket (2). or contaminated. See faults section “water level sensors”. If gasket is more than 6 months old consider a replacement. 50 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Off if not required, when fitted. Same as sterilise temperature. Requires hold warm to be enabled in settings/system/ door, hold warm. 51 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Test certificates for Electrical safety and Calibration of the control system are supplied as standard, full test analysis results are available to special request. 53 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Instruments specified as individually accurate to +/- 1%. It should not be used as a calibration standard for Pressure or Temperature indicators, which must be calibrated in accordance with the Manufacturer’s Specifications. 54 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
NOTES SPACE This space is for your own notes on the system 60 Astell Autoclaves/Sterilisers – AMB220 | AMB230 | AMB240 | AMB420 | AMB430 | AMB440 | AMA240 | AMA440...
Page 61
Astell Scientific Ltd, Powerscroft Road, Sidcup, Kent DA14 5DT, UK Tel: +44 (0)20 8309 2031 E-mail: info@astell.com Astell Scientific Ltd is a company registered in England. Registered no. 487717.