Calibrating An Ac Voltage Source - Fluke 540B Instruction Manual

Thermal transfer standard
Table of Contents

Advertisement

540B
STEP
CONTROL
POSITION
OPERATION
a
POWER
ON
Set the RANGE switch in a position that
will
cause
GALV
OPEN
the PERCENT INPUT meter to indicate close to,
SENSITIVITY
LOW
but not exceed, 100% for the level
to
be calibrated.
RANGE
Adjust the de source voltage to the exact rms value
MODE
DC SEARCH
of
the
ac
voltage
to be calibrated. (Note PERCENT
INPUT meter indication. )
b
MODE
DC TRANSFER
Tap the GALV switch against the spring tension to
GALV
MOMENTARY just long enough
to
see
if
the
GAL-
REFERENCE
COARSE
VANOMETER needle will stay on
scale.
Adjust
ADJUST
MEDIUM
the COARSE and MEDIUM controls until a
zero
reading is obtained.
c
GALV
OPEN
Adjust
the
GALV ZERO control for zero reading
SENSITIVITY
MED
on the GALVANOMETER.
GALV ZERO
d
GALV
LOCK
Adjust the MEDIUM and FINE controls for zero
REFERENCE
MEDIUM
reading on the GALVANOMETER.
ADJUST
FINE
e
GALV
OPEN
Adjust the GALV ZERO control for a zero reading
SENSITIVITY
HIGH.
on the GALVANOMETER.
GALV ZERO
--
f
GALV
LOCK
Adjust the FINE control for a zero reading on the
REFERENCE
GALVANOMETER.
ADJUST
FINE
g
GALV
OPEN
SENSITIVITY
LOW
h
MODE
AC SEARCH
Adjust the ac level to be calibrated until a reading
is obtained on the PERCENT INPUT meter that is
equal to the reading obtained and noted in step a.
i
MODE
AC TRANSFER
NOTE: DO
NOT
ADJUST THE REFERENCE
ADJUST CONTROLS IN THIS
MODE.
j
SENSITIVITY
LOW
Tap
the
GALV switch against
the
spring tension to
GALV
MOMENTARY just long enough to see
if
the
GAL-
VANOMETER needle will stay on scale while
adjust-
ing the level of the ac
to
be calibrated. When an on
scale reading is obtained adjust the
ac
level for a
zero reading on the GALVANOMETER.
k
GALV
LOCK
Further refine the adjustment of the ac voltage until
SENSITIVITY
MED
a GALVANOMETER reading of zero is obtained in
HIGH
the MED and HIGH SENSITIVITY positions. Record
the ac voltage source setting.
1
Repeat steps b through k at least three times or until repeatibility is attained.
Make
note of the
ac source setting.
m
Reverse the input polarity to the thermocouple by depressing the POLARITY switch. Repeat steps
b through k at least three times or until repeatibility is attained. Make note of the ac source setting.
n
Set the ac source voltage adjustment in the center of the two settings recorded in steps
l
and m.
The rms value of the ac voltage now equals that desired.
Figure
2-
7.
CALIBRATING AN AC VOLTAGE SOURCE
2-9