4.3.4 Adjust the Auto Zero Range (F15)……………………………13 4.3.5 Adjust DA (F25) ………………………………………………..13 5. RS485 Communication Port………………………………………15 5.1 Connecting the RS485 output……………………………….15 5.2 Communication Mode of RS485 ……………………………15 5.3 Data Format (TI-1500VC Vision: V3.5C) ………………….16 5.4 Command Format ……………………………………………17 Hardware Vision Explain……………………………………………..18...
1. TI-1500VC Specifications 2. Dimensions & Installation 2.1 TI-1500VC Dimension: (1) Analog Signal Input: The transmitter can drive 4 x 350Ωload cell. The signals input every channel can be adjusted. The excitation voltage is +5V. (2) Interface: a. 4—20mA analog output (with digital adjust) b.
0.5d input/output are in one network, the last Band Note that the scale must be in TI-1500VC’s J4 should be in standstill to automatically zero. 8d 9d RS485 re-phase short circuit, RS-485 Selections are in Display...
Page 5
2 √ Selects the range within which Determines location of the 0√ Zero Range the scale may be zeroed. Note Decimal Pt. decimal point. 0.00 0.000 that the indicator must be in 0.0000 00 standstill to zero the scale. Allows the lb/kg key to be Cr-DIS√...
3.2 User menu Changing communication mode and formula is in this menu. Allows you to key-in known zero Press ZERO Key-in Zero calibration value in case of key to begin A menu memory loss in the field. sequence Name/Code Description Code/Value Allows you to key-in a known Press ZERO...
2. In F or A menu mode, press and hold “↑” key, back to normal weigh mode. 4.2 Display and Annunciator Definitions of keys function TI-1500VC is utilized a blue 7-segments LCD display. Follow is the display detail: Keyboard Setting Mode...
A/D count. When the stabile annunciator turns on, press “↓” key to set zero point, Note: Every TI-1500VC has Factory Reset (F21) when it out of then press “SET” key to save the zero value.
4.3.5 Adjust DA (F25) The analog output 4-20mA and 0-10V has been adjusted, when the TI-1500VC out of the factory. The parameter should not be adjusted again generally. If necessary, please follow the step: (1) Enter F23 to select the type of analog output.
Weight State Parity Data Units TI-1500VC adr:00 5.3 Data Format ( TI-1500VC Vision:V3.5c) b. Point to Multi-point Start Transmission (1 bit): STX; TI-1500VC Device Address (2 bits): Address information; adr:01 I/O port state (3 bits): I/O port state information; Weight Data (8 bits): Symbol (1 bit) + Decimal Point (1 bit) + xxxxxx TI-1500VC (6 bits, integer type);...
Show as follow figure. There is a hardware’s mark label paste near the power terminal. So the different PCB board which marks Operational command (2 bits): different sign can not be changed. Otherwise the TI-1500VC will be GD: Get weight data; destroyed. If the transmitter out of operate because of the GC: Get calibration information;...