Summary of Contents for Zimmer Z Wave Q

  • Page 1 User Manual Z Wave...
  • Page 3 Illustrations Front of the device Fig. 1 Encoder Handpiece...
  • Page 4 Illustrations Rear of the Device / Rear of the SysCart Fig. 2 Switches / 1 Socket for mains cable Connector 2 Main fuse sockets 3 Identification plate 4 Socket for footswitch Fig. 3 Switches / 1 Socket for handpiece Connector 2 Identification plate 3 Socket for water filling / emptying...
  • Page 5 Illustrations Screens and Displays Fig. 4 Start Screen 1 Settings 2 Protocols – direct activation of application screen Fig. 5 Therapy Screen 1 Pulse Energy 2 Frequency 3 Start button 4 Treatment Menu 5 Pulse Mode...
  • Page 6 Illustrations Screens and Displays Fig. 6: Maintenance Screen 1 Handpiece Information 2 Footswitch Connection 3 Fill / Empty the water cooling circuit 4 Test Handpiece Button 5 General Settings 6 Maintenance Button...
  • Page 7 Illustrations Accessories Slim Handpiece Accessories Aesthetic handpiece Accessories Aesthetic handpiece Slim handpiece Silicone cap 15mm Silicon cap 39mm Applicator head 15mm Applicator head 39 mm Filling funnel with hose and Funnel Holder...
  • Page 8: Explanation Of Symbols

    Explanation of Symbols In the instructions for use this symbol indicates “Caution”. In the instructions for use this symbol indicates "Attention" with regard to possible Attention damage of the device. Application Part Typ BF Instructions for use Follow instructions for use. Serial number Article number Manufacturer...
  • Page 9 Explanation of Symbols Connection for filling the cooling unit with water (max.1.5L) Max. 1.5L The device must not be tilted. Pump symbol indicating the location of the pump for water evacuation. Do not reuse. If used more than once, cross infection may occur. The transport package shall not be exposed to sunlight.
  • Page 10: Table Of Contents

    Contents Illustrations Front of the Device Rear of the Device Screens / Displays Explanation of Symbols Page Intended Purpose Side Effects Application Information Warnings Device Setup Self-Test Basic Settings Operation Instructions Technical Information...
  • Page 11 Contents Cleaning / Disinfection Scope of Delivery and Accessories Device Combinations Safety and Maintenance Setup Water cooling Circuit Error Messages Troubleshooting Disposal Manufacturer´s EMC declaration Valid for ZWave devices. This instructions for use is an integral part of the device. It must be stored with the device and kept accessible at all times for anyone authorized to operate this device.
  • Page 12: Intended Purpose

    Intended Purpose Intended Use is a device for the external application of radial sound waves to ZWave relieve minor muscle aches and pain, for the temporary increase in local blood circulation and temporary reduction in the appearance of cellulite. Treatment Patient a) Age: Only adults, Population...
  • Page 13: Side Effects

    Side Effects Treatments with ZWave can occasionally lead to dermatitis, petechiae, haematoma, swelling or pain. Page 2...
  • Page 14: Application Information

    Attention any time. Only accessories provided by Zimmer MedizinSysteme GmbH must be used. Attention The silicon protection cap (15mm) for the handpiece H O Slim is only for single use.
  • Page 15: Warnings

    Warnings Users of the device must be instructed in the correct use of the system and have appropriate skills. Treatment instructions concerning the place of treatment, duration and extent of treatment require medical knowledge and may only be given by licensed physicians, therapists and health professionals.
  • Page 16: Device Setup

    Device Setup Note: Make sure that the power switch on the device is set to "0". Mains cable Connect the power cable to the appropriate socket on the device and connection connect the cable to the mains. Note: The device may only be connected to power outlets with grounding contact.
  • Page 17: Self-Test

    Self-Test Self-Test After the device is switched on, the ZWave performs a self-test, which checks all internal components. In the event of a fault, an error message appears. In addition, an extended function test can be performed as described below. This test should be performed monthly or if there are any doubts about the viability of the device.
  • Page 18: Basic Settings

    Basic Settings Setting Screen Note: Changes in the basic settings are only possible from the start-up screen Settings menu Via the Start Screen you get into the settings menu (general), the factory settings can be individually changed and adjusted The setting options are described below. The default settings are pre-programmed as shown in the screen.
  • Page 19 Basic Settings (9) Colour Scheme Activation of the “Colour scheme” button changes the display colour from dark to bright and vice versa. (10) Handpiece In this display panel, the counter status for the connected handpiece is illustrated. By touching the button with the green arrow in the right bottom of the display the settings are saved and the configuration menu is closed Maintenance Screen...
  • Page 20 Basic Settings Protocol Screen (1) Energy Displays the set pulse energy. When the application is active, the bar graph is filled in. The pulse energy can be set before and during the delivery of pulses. The pulse energy is adjustable between 60–185 mJ in 10 mJ steps. (2) Frequency Displays the set frequency.
  • Page 21: Operation Instructions

    The pulse generator in the handpiece is a wear part and needs to be replaced after a certain period as its ability to function decreases with time. Zimmer MedizinSysteme GmbH guarantees unrestricted application of at least two million pulses per one pulse generator. Depending on the power and frequency, more than two million pulses can be generated.
  • Page 22 Operation Instructions Application Please hold the handpiece as shown below. The ZWave operates with mechanical energy; it is transferred to the patient through a handpiece. For this purpose, the handpiece with the applicator is placed vertically on the application area. While applying pulses, you can work dynamically on one area as well as stationary at one point.
  • Page 23 Technical Data Power supply 110-240 Vac/ 50-60 Hz Mains fuse 2 x T3.15AL, 250V Power consumption 250 VA Protection class Applied part class IP protection class Device IPX0, Footswitch IPX5, Handpiece IPX0 Operating Mode Continuous operation Dimensions 20.87” x 21.26” x 54.72” (530 x 540x 1390 mm) (L*W*H) Weight 56 lbs.
  • Page 24: Cleaning / Disinfection

    Cleaning / Disinfection Before starting any maintenance and cleaning measures, the device must always be turned off with the main switch, and the mains cable must be disconnected. Make sure that during the cleaning and disinfection the labelling of the device (such as warnings, label of control devices and identification plate) are not damaged.
  • Page 25 Cleaning / Disinfection Silicone protection cap Cleaning, manually: Tools: Drinking water, lukewarm  Vessel, e.g. kidney dish  Brush, e.g. medium hard toothbrush  Alcohol-free plastic cleaner (e.g. cleaner for medical devices)  Remove the silicone protection cap from the applicator head prior to cleaning. Prepare a solution of the cleaning agent following the instructions of the manufacturer.
  • Page 26 Storage tray for accessories 67250130 Mains cable 10102909 Assembly Instruction Subject to change! The device may only be operated with original accessories from Zimmer MedizinSysteme GmbH. Otherwise, the function and safety of patients, users and other persons cannot be guaranteed. Page 15...
  • Page 27: Device Combinations

    Device Combination Device Combinations For the ZWave no combination units are provided by the manufacturer. Anyone who contrary to these specifications combines devices and thus operates a medical system does so at its own risk. Whoever combines the devices and thus operates a medical system is responsible for the correctness of the combination.
  • Page 28: Safety And Maintenance

    The ZWave is manufactured according to the IEC 60601-1 safety regulations. a) Zimmer MedizinSysteme GmbH can only be held responsible for safety and reliability of the device if  the device is operated using a proper power outlet with earth contact and the electrical installation complies with DIN VDE 0100 part 710, ...
  • Page 29: Setup Water Cooling Circuit

    Setup Water cooling Circuit Filling the water If the user presses the button to fill the water cooling circuit the following steps have cooling circuit to be done: The message gives the user the instruction to plug in the filling tube. Then the user has to fill in the distilled water and monitor the fill level in the images shown below.
  • Page 30 Setup Water cooling Circuit Emptying the water To empty the water cooling circuit the user has to press the button in the configuration menu. After that, a message occurs that instructs the user to plug in the cooling circuit tube and set up a collection container to empty the cooling circuit. (See figure below). After that step a message occurs that shows that the pumping process.
  • Page 31 Setup Water cooling Circuit Emptying the water If the process was stopped a message occurs to show the user that there could be still some water left in the system. cooling circuit Disconnect the To separate the handpiece, the user has to press the right button in the configuration menu.
  • Page 32 Setup Water cooling Circuit If process failed, an error message occurs and gives the user the instructions to repeat the process. If the process was successful, a message occurs to show the user that the handpiece can now be disconnected. Attention The handpiece must not be disconnected from the device during active therapy.
  • Page 33: Error Messages Troubleshooting Disposal

    Check the total pulse count of the handpieces in the settings menu. If the total is less than two million shocks, it may be subject to replacement under warranty. To replace the sound wave generator, please contact Zimmer MedizinSystems at (800)327-3576...
  • Page 34 Error Messages Troubleshooting Disposal Device Malfunction No response to the main switch / display remains dark. Possible cause 1 Mains power connection Fix for cause 1 Check whether the power cord is properly plugged into the electrical outlet and whether the plug is firmly inserted into the socket of the device. Check the power cord for damage.
  • Page 35 Error Messages Troubleshooting Disposal Empty tank by water evacuation pump In case the software is not able to empty the water out of the device automaticly after pushing the “empty tank” button in the maintanance software, use the manual water evacuation pump from inside the SysCart. (see picture above). To access the pump, open the front flap and unlock the four screws in •...
  • Page 36 Distributor 3 Goodyear, Suite B Irvine, CA. 92618 Phone (800) 327-3576 Fax (949) 727-2154 www.zimmerusa.com info@zimmerusa.com Manufacturer Zimmer MedizinSysteme GmbH Junkersstraße 9 89231 Neu-Ulm, Germany Tel. +49 7 31. 97 61-291 Fax. +49 7 31. 97 61-299 export@zimmer.de www.zimmer.de Disposal The device may only be returned to the factory in the original packaging.
  • Page 37: Manufacturer´s Emc Declaration

    Manufacturer´s EMC declaration ZWave is developed according to the recognized standards of technology; the information on the intended use of the components is taken into account. ZWave must not be operated near active HF surgery devices or magnetic resonance tomography that can cause high levels of electromagnetic interference.
  • Page 38 Manufacturer´s EMC declaration Table 2 Guidance and Manufacturing Declaration- Electromagnetic Immunity The device ZWave is intended for use in the electromagnetic environment specified below. The customer or user of the device ZWave should ensure that it is used in such environment. Immunity Tests IEC 60601- Test Level Compliance Level...
  • Page 39 Manufacturer´s EMC declaration Table 3 Guidance and Manufacturing Declaration- Electromagnetic Immunity The device ZWave is intended for use in the electromagnetic environment specified below. The customer or user of the device ZWave should ensure that it is used in such environment. Compliance Level Immunity Test IEC 60601-Test...
  • Page 40 Z Wave User Manual Zimmer MedizinSystems 3 Goodyear, Suite B Irvine, CA. 92618 800 327 3576 949 727 2154 fax www.zimmerusa.com info@zimmerusa.com...