Dual Pro is a fully adjustable proportional head
array with "On Board Programming". This gives
the user the ability to operate their wheelchair
using the full range of the wheelchair's speed
and acceleration parameters. The Dual Pro
contains a combination of proximity and force
sensor technologies. This technology allows the
end user to customize the head array in the
field. Traditional switched head arrays do not
allow this level of customization. For set up
instructions please see the set up guides for 3 or
4 direction proportional located on our website:
www.switchit-inc.com.

On Board Programming

Each pad has a force and a proximity (crawl)
sensitivity setting, and both settings can be
adjusted individually through the "On Board
Programmer". To configure the different pads
follow these easy steps.
3250 Williamsburg Lane—Missouri City, Texas 77459
1-800-376-988—713-896-1699—Fax:713-937-7197
www.switchit-inc.com
DUAL PRO™ HEAD ARRAY SYSTEMS
1. Press the "Set" & "Force" buttons at the
same time for 1 second.
2. This will allow you to select the pad you
would like to configure as indicated by the light
corresponding with each pad.
3. Press the set button to scroll through and
select the pads you would like to configure.
4. You can set the "Force" by pressing the
"Force" button. More Force lights means
more force is required to reach full speed. This is
indicated through the series of lights on the
right side of the keypad.
.
5. You can set the Proximity setting by pressing
the "Crawl" button. More "Crawl" lights means
more speed when within range of the proximity
sensor. This is indicated through the series of
lights on the right side of the keypad.
Note: Once the maximum sensitivity is reached,
the lights will reset to the left indicating you
have reached the minimum setting. The
sensitivity will cycle through the lights until the
desired setting is reached. Once the sensitivity
of both switches is set as desired, scroll through
the pads until no pad indicators are lit.

External Switch Options

The Dual Pro has three inputs for external
switches: mode, reverse, and an assignable user
switch. These are located on the bottom of the
pad.
FlexSwitch on Swing Away
Piko Switch on Gooseneck
Swing away Egg Switch

Technical Information

Dimensions:
Back pad: 3" x 6" (76 x 152 mm)
Side pads: 1.5" x 5.5" (38 x 140 mm)
Whitmyer Hardware: highly adjustable
Displacement: up to 1/4" (3 mm)
Force: 0.3 0z to 1lb (10g to 0,5kg)
Doc: 09.2015 –DP-IGRevA

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DUAL PRO DP-A and is the answer not in the manual?

Questions and answers

Summary of Contents for Switch-it DUAL PRO DP-A

  • Page 1 DUAL PRO™ HEAD ARRAY SYSTEMS Note: Once the maximum sensitivity is reached, 1. Press the “Set” & "Force" buttons at the the lights will reset to the left indicating you same time for 1 second. have reached the minimum setting. The sensitivity will cycle through the lights until the desired setting is reached.
  • Page 2 Head Pad Styles Offered Programming Interface Examples Detailed Example 1. 40% speed within proximity (crawl) and light The Dual Pro will send a signal to the chair as a force for full speed. percentage of full speed. For instance, if the chair is set to 100% forward speed and the proximity (crawl) switch is set to 20% then the chair will only move at 20% of full speed in the...

This manual is also suitable for:

Dual pro dp-pDual pro dp-3-aDual pro dp-3-p

Table of Contents