Axion STIM-PRO T-400 Instruction Manual
Hide thumbs Also See for STIM-PRO T-400:

Advertisement

Quick Links

INSTRUCTION MANUAL
STIM-PRO T-400
axion GmbH
Mollenbachstr. 13
71229 Leonberg
Germany
2460
V 1.1

Advertisement

Table of Contents
loading

Summary of Contents for Axion STIM-PRO T-400

  • Page 1 INSTRUCTION MANUAL STIM-PRO T-400 axion GmbH Mollenbachstr. 13 71229 Leonberg Germany 2460 V 1.1...
  • Page 2: What Is Pain

    AXION INTRODUCTION INTRODUCTION The primary functions of the STIM-PRO T-400 are pain relief and pain treatment with the TENS mode. What is pain? Pain is a warning system and the body’s method of telling us that something is wrong. Without it, abnormal conditions and injuries to vital parts of the body may go undetected.
  • Page 3: Scope Of Delivery

    For further details please refer to your doctor. Scope of delivery Each STIM-PRO T-400 is fully equipped with the following standard equipment. 1 piece TENS machine 4 pieces...
  • Page 4: General Description

    These impulses reach the nerves and muscles through the skin. The electronics of the STIM-PRO T-400 create electrical impulses that can be modified in intensity, pulse duration and frequency in the manual modes. All functions are easy to handle and the big display shows all parameter values clearly readable.
  • Page 5: Operating The Device

    STIM-PRO T-400 OPERATING THE DEVICE ON / OFF switch Open the front lid of the device to reach the controls. The device can be switched on inreasing the intensity with the regulators on top of the machine. Regulating the intensity The intensity can be adjusted stepless from 0-80mA by turning the intensity regulators on top of the device.
  • Page 6 AXION Modus Auswahl-Taste There are four TENS modes available: (B) Burst, (N) Nor- mal, (M) / (SD1) / (SD2) Modulated, (P) Program mode. The modes are switched by pressing the “Mode” button. In the program mode P there are 12 pre-defined programs with fixed settings availabe (more information regarding the programs on pages 22-23).
  • Page 7: Program Mode

    STIM-PRO T-400 Regulating the pulse width The pulse width can be adjusted from 30 µs to 300 µs. It regulates the time for which the electrical signal is applied. The current value is shown on the LCD display. Timer The device comes with a timer that can be adjusted from 1 to 60 minutes.
  • Page 8: Program Mode (P)

    AXION Program mode (P) The programs have proven to be helpful in most cases for the suggested applications. However, they might be used for the treatment of other injuries or discomforts as well. Every patient responds individually to stimulation current.
  • Page 9 STIM-PRO T-400 P 07 For sensible parts of the body Due to the reduced duration of pulse these applications are to be specifically used in cases of tension headache, face pain, neck pain, herpes zoster, migraine. Frequency: 80 Hz Pulse width: 60 μs...
  • Page 10: Construction Of The Device

    AXION CONSTRUCTION OF THE DEVICE Front Back...
  • Page 11: Changing Batteries

    STIM-PRO T-400 CHANGING BATTERIES When the low battery indicator is shown, the battery needs to be replaced. Remove the battery if the device is not likely to be used for some time. Make sure the power is off. Open the battery compartment by sliding down and removing the front cover.
  • Page 12: Safety Checks

    AXION MAINTENANCE, TRANSPORTATION AND STORAGE OF THE DEVICE Non-flammable cleaning solution (containing 70% of alcohol) is suitable for cleaning the device. Stains and spots can be removed with a cleaning agent. Do not submerge the device in liquids or expose it to large amounts of water.
  • Page 13: Conformity To Safety Standards

    There is no stimulation current? Check if the electrodes are correctly connected to the same channel and if the intensity is high enough. CONFORMITY TO SAFETY STANDARDS The STIM-PRO T-400 device is in compliance with the EN 60601-1-2:2007 and 60601-1:2006 safety standards.
  • Page 14: Technical Description

    AXION TECHNICAL DESCRIPTION Channel 2 channels with adjustable intensity Intensity Adjustabl 0 - 80 mA with a load of 500 Ohm on each channel Pulse amplitude Asymmetrical, bi-phasic square pulse Output Voltage 0 - 50 V Power Source 1 pc. 9,0V battery...
  • Page 15: Warranty

    STIM-PRO T-400 Operating Temperature 0° C to 40° C Relative humiditiy 30 % - 75 % condition Air pressure 700 hPa - 1060 hPa Notice All technical values include a tolerance of +/- 5 % WARRANTY All TENS models carry a warranty of 24 months from the date of delivery.

Table of Contents