Battery & Install Batteries are not included with the 163-DIG product. Refer to Scale Specifications on page 6 for the type of batteries to use. A display of Lo or the battery symbol on the 163-DIG readout indicates low batteries and replacements are required.
Independent Failure Mode Warning Do not use results from the 163-DIG to make changes to your system, do use the results to perform additional system checks. Section 1. 163-DIG Scale Overview...
Components Table 2-1. Parts and Accessories Part Number Description 163-DIG Digital Compression Scale 603061 Compression Foam Block 805972 Soft Case for 163-DIG (optional) Platform lined with thin foam Display Removable & Backlit LCD panel Capacity Units & Accuracy Select kg, lbs...
Soft Case Features Initialization Sequence Upon power-on, 163-DIG performs a startup sequence countdown. Keep the unit on a flat and stable surface during initialization. At the end of initialization 163-DIG automatically zeroes. Zero/Tare Zero/Tare allows users to zero the scale for relative measurements, or measurements in different orientations.
Wriggle and slide the Display Panel upwards in the direction of the Platform while holding the release tab. The separated Display Panel is corded to the Platform Body and ready for use in desired orientation. Tare the scale when changing orientation. Figure 2-2. Display panel removed from 163-DIG Features...
Raise compression paddle (plate) to maximum height and center it. Turn off Flex-mode. Place the Model 163-DIG on the breast support approximately centered on the compression paddle. Optionally place the Foam Compression Block on top of the 163-DIG Platform, which is also useful for testing curved paddles.
Power up and zero the 163-DIG flat and stable surface. Rotate the 163-DIG onto its narrow side on a flat and stable surface and zero the scale. Bring the 163-DIG to the Mammography system and very carefully apply pressure to clamp the scale between the compression paddle and detector.
10 kg certified weight or combination of certified weights. Process While the 163-DIG is off press and hold the Unit button while holding the Power button until the display reads CAL. Immediately after, a random set of numbers appear (raw load cell value).
• Avoid contact with alcohol. • Wipe after use with a damp cloth, if necessary. Contacting Gammex Support Technical support is available from Gammex Inc. Contact the Support department: • By telephone: +1 (321) 259-6862, option 3, then 3. •...
In addition to the regulatory information contained in the body of this manual, the following supplemental regulatory information is provided. Gammex Symbols The following symbols are used in this guide and in Gammex’s product labels. WARNING: This symbol indicates a hazard that could result in major injury or equipment damage.
Gammex products, contact Gammex directly. Modifications to Equipment WARNING: No modifications may be made to this unit without the expressed written consent of Gammex. Unauthorized changes may present a risk to user or patients and will void the warranty. Appendix A: Regulatory Supplement...