3 Connect the waveguide filter and amplifier to the WR12 waveguide port of the
4 Connect the provided power adaptor to the N9029ACST-U12 Pwr port.
Software installation requirements
The following software is required and should have been installed or preinstalled
in the M9537A controller.
– M8195A Soft Front Panel (SFP) is required for manual or programmatic
– Keysight Connection Expert, which is preinstalled with the IO Libraries suite.
– IQTools is used to create the FMCW signal required for automotive radar and
Installing standalone version of IQTools
IQTools is a free MATLAB application suitable for use with the M8195A AWG and
other Keysight AWGs. It is capable of synthesizing and generating many types of
waveforms, some which are not available in the M8915A SFP. It also contains
many example applications.
We will use IQTools to calibrate the system and create radar FMCW signals. The
Signal Generation Radar Assistant is also an example application in IQTools. If
you have purchased the M8195A with the N6171A MATLAB license option, you
will receive access to the modifiable version of IQTools. Keysight has also made a
Keysight E8740A-070 Start-up Guide
output amplitude range (75 mVppk to 1 Vppk) to within the linear operating
range of the N9029s' IF input port.
Terminate the unused complement Data Out port with a 50-ohm termination.
Connect the RF output from the N5183B to the LO input of the upconverter
with a 2.4 mm coaxial cable. A connector saver (11900B 2.4 mm(f) to 2.4
mm(f) adaptor) is recommended for the N5183B RF output.
N9029ACST-U12. The N9029ACST-U12 provides a +9 V DC port that can be
used to bias compatible amplifiers from Virginia Diodes.
control of the M8195A Arbitrary Waveform Generator (AWG). The SFP also
has some built in signal creation functions. If the software has not been
preinstalled in the embedded controller, please consult Chapter 2 of the
Keysight M8195A Arbitrary Waveform Generator Getting Started Guide.
This application is required to identify the VISA address of instruments.
to calibrate the Signal Generation system. The following section provides
instruction to install IQTools.
11
Need help?