Lauper Instruments LIPPKE 2000 SL User Manual

Package test system
Table of Contents

Advertisement

User Manual
Package Test System
LIPPKE 2000 SL
Lauper Instruments AG
Gas-Detection
Irisweg 16B
CH-3280 Murten
Tel. +41 26 672 30 50
info@lauper-instruments.ch
www.lauper-instruments.ch
www.lauper-instruments.ch

Advertisement

Table of Contents
loading

Summary of Contents for Lauper Instruments LIPPKE 2000 SL

  • Page 1 User Manual Package Test System LIPPKE 2000 SL Lauper Instruments AG Gas-Detection Irisweg 16B CH-3280 Murten Tel. +41 26 672 30 50 info@lauper-instruments.ch www.lauper-instruments.ch www.lauper-instruments.ch...
  • Page 2: Table Of Contents

    Contents Safety instructions 5 - 6 Introduction 7 - 11 System survey 7 -9 2.1.1 The LIPPKE 2000 SL control unit 2.1.2 Holding stand and needle heads 8 - 9 2.1.3 Clamping device General operating instructions 10 - 11 Commissioning...
  • Page 3 Package test system LIPPKE 2000 SL Introduction Pressure loss test 21 - 25 General instructions for pressure loss test 21 - 22 5.1.1 Safety instruction 5.1.2 Standard settings 5.1.3 Instructions for clamping device 5.1.4 Instructions for holding stand 21 - 22 5.1.5 Measurement...
  • Page 4 Package test system LIPPKE 2000 SL Introduction Index of figures Figure 2.1: Package test system Figure 2.2: Holding device Figure 2.3: Needle No. 1B Figure 2.4: Needle No. 2B Figure 2.5: Needle No. 3B Figure 2.6: Clamping device Figure 2.7: Control unit, front panel Figure 3.1: Holding stand...
  • Page 5: Safety Instructions

    Package test system LIPPKE 2000SL Safety instructions 1 Safety instructions This operator’s manual has to be read, understood and observed before commissioning the • test system. If one of the following safety instructions is not or only partially observed, the Paul Lippke •...
  • Page 6 Package test system LIPPKE 2000SL Safety instructions The following symbols are used in order to obtain more clarity: • Caution: Please observe instruction by all means. • • Important note, may lead to malfunctions or damage of test system if •...
  • Page 7: Introduction

    The serial interface permits the output of test reports to a printer or PC. 2.1.1 The LIPPKE 2000 SL control unit The LIPPKE 2000 SL control unit houses the micro controller and the pneumatic system. Here the air supply is connected and guided through the package holding device to the package in a controlled way.
  • Page 8: Holding Stand And Needle Heads

    Package test system LIPPKE 2000 SL Introduction 2.1.2 Holding stand and needle heads A closed package is connected to the control unit by means of the package holding stand (see figure 2.2). Figure 2.2: Package holding stand Two different types of needle heads are used in order to conduct compressed air into the package: the mono needles (type 1B and type 2B) and the double needle type 3B.
  • Page 9: Clamping Device

    Package test system LIPPKE 2000 SL Introduction Figure 2.4: Needle Type 2B In case of the double needle (see figure 2.5), the sensor line is lead separately into the package. If very large air supply is required for measuring, back-pressure is possible when using a mono needle.
  • Page 10: General Operating Instructions

    Introduction 2.2 General operating instructions After switching on (power switch on the rear panel, see figure 3.3, page 16), the LIPPKE 2000 SL control unit first displays the parameter last used (see figure 2.7). Figure 2.7: Control unit, front panel...
  • Page 11 Package test system LIPPKE 2000 SL Introduction Setting of test time (only in LOSS Mode!) is performed on the right panel. The function is acti- vated by pressing one of these three key, i.e. a time value is now displayed. ”SEC MIN” serves for changing to the desired unit of time (seconds or minutes).
  • Page 12: Commissioning

    Package test system LIPPKE 2000 SL Commissioning 3 Commissioning Conditions 3.1.1. Electrical supply Important note: Connection to a mains power supply unit different from the nameplate can lead to device damage. A mains power supply of 230 V∼ ± 10 % at a frequency of 50 Hz ± 1 Hz is required for the sys- tem (other options upon request).
  • Page 13: Compressed Air Supply

    Package test system LIPPKE 2000 SL Commissioning 3.1.3. Compressed air supply Caution: If it is not ensured that maximum pressure is kept in any case, there is danger of bursting of hoses and other parts. Important note: Perfect function is not guaranteed if the pressure value remains under the above-mentioned input pressure.
  • Page 14: Mechanical Construction Of Holding Stand

    Package test system LIPPKE 2000 SL Commissioning 3.2.2. Mechanical construction of holding stand In general, the package holding stand is delivered already completely assembled, so that you don’t have to read this chapter. Assembly according to 3.1 only requires a few movements of the hand. The bar must be screwed into the base plate.
  • Page 15: Pneumatic Connections

    Package test system LIPPKE 2000 SL Commissioning O-Ring O-Ring Filterröhrchen Filterschaum- Filtereinsatz stoff Nadelkopf Prüfkopf Filter Figure 3.2: Construction of filter First screw the filter element with foam filter into the test head for this purpose. Then the small filter tube and the needle head can be attached. For the positions of the individual O-rings please see figure 3.2.
  • Page 16: Operation

    Package test system LIPPKE 2000 SL Commissioning The connections for holding stand and clamping device are also on the rear panel of this module. Control unit Holding device Clamping device Connection Mark Connection Mark SENSE thin hose thin hose FEED...
  • Page 17: Burst Test

    Package test system LIPPKE 2000 SL Burst test 4 Burst test General instructions for burst test (LOSS mode) The general information about burst tests is presented in brief outlines and sorted in the order valid for carrying out a measurement procedure. This will facilitate subsequent retrieving of special information during operation.
  • Page 18: Measurement

    Package test system LIPPKE 2000 SL Burst test For small-sized packages which will deform under pressure also in the area of the sealing dia- phragm, you should use the grey diaphragm ( 1mm thin and very flexible). However, when performing burst tests it is not necessary in many cases to seal the package with a diaphragm at the prick-in point.
  • Page 19: After The Measurement

    Package test system LIPPKE 2000 SL Burst test 4.1.6. After the measurement Caution: If the package did not burst, it is still under pressure! Remove the pack- age carefully, since pressure falls rather quickly through the prick-in point. In case of package bursting, the burst pressure is displayed by the 2000 SL and you can take out the package.
  • Page 20: Figure 4.1: Print-Out Of A Burst Test

    Package test system LIPPKE 2000 SL Burst test PACKAGE TEST SYSTEM LIPPKE 2000 SL Date: 28.08.98 Time: 22:51 Mode: Burst test Product: Batch: Tester: Remark: Pressure rise rate: 10 mbar/s Burst pressure: 265 mbar Figure 4.1: Print-out of a burst test You can now enter product, batch, tester and a remark if necessary.
  • Page 21: Pressure Loss Test

    Package test system LIPPKE 2000 SL Pressure loss test 5 Pressure loss test 5.1 General instructions for pressure loss test The general information about pressure loss tests is presented in brief outlines and sorted in the order valid for carrying out a measurement procedure. This will facilitate subsequent retrieving of special information during operation.
  • Page 22: Measurement

    Package test system LIPPKE 2000 SL Pressure loss test Testing flat packages: Here, there is the danger that the needle sticks through the package. In this case it is safer to glue on a black diaphragm and to carefully prick the package with a sharp object (e.g.
  • Page 23: Carrying Out The Measurement

    Package test system LIPPKE 2000 SL Pressure loss test 5.3 Carrying out the measurement You may start the measurement by the ”START” key. After that pressure is built up in the pack- age. The first 50% of the desired test pressure are built up relatively quick, whereas the second 50% are built up more slowly in order to exactly reach the nominal value.
  • Page 24: Protocol

    Package test system LIPPKE 2000 SL Pressure loss test 5.5 Protocol In any case, test data will be output on the serial interface (e.g. to a printer). You can read details about formatting this output in chapter 6, page 26. The standard type of data output at the end of...
  • Page 25: Figure 5.1: Print-Out Of A Regular Pressure Loss Test

    Package test system LIPPKE 2000 SL Pressure loss test PACKAGE TEST SYSTEM LIPPKE 2000 SL Date: 29.08.98 Time: 20:41 Mode: Pressure loss test Product: Batch: Tester: Remark: Test pressure: 500 mbar Compensation time: 0 seconds Test time: 300 seconds Pressure loss: 5 mbar Figure 5.1: Print out of a regular pressure loss test...
  • Page 26: Configuration

    Package test system LIPPKE 2000 SL Configuration 6 Configuration Important note : Wrong settings may lead to test system malfunctions. If in doubt, ask a service technician from Paul Lippke Handels-GmbH. The configuration serves for matching the following items: • Number of print-outs •...
  • Page 27: Auto-Stop-Threshold

    Setting of the Auto-Stop-threshold in the range from 0 to 99. If the current pressure drops within 1/10 by this percentage or more, LIPPKE 2000 SL judges this as ”bursting of package” and fin- ishes the test. The display will change as follows: “Auto-StoP 20”...
  • Page 28: Figure 6.1: Test Report, Burst Test

    Package test system LIPPKE 2000 SL Configuration Test report in French language (corresponding to figures 6.1 and 6.2), ”OUtput French” Test report in Spanish Language (corresponding to figures 6.1 and 6.2), ”OUtput Spanish” Test report in Italian Language (corresponding to figures 6.1 and 6.2), ”OUtput Italian”...
  • Page 29: Figure 6.2: Test Report, Pressure Loss Test

    Package test system LIPPKE 2000 SL Configuration PACKAGE TEST SYSTEM LIPPKE 2000 SL Date: 29.08.98 Time: 20:41 Mode: Pressure loss test Product: Batch: Tester: Remark: Test pressure: 600 mbar Compensation time: 0 seconds Test time: 300 seconds Pressure loss: 5 mbar Figure 6.2: Test report, pressure loss test...
  • Page 30: Figure 6.4: Data Output At The End Of Pressure Loss Test

    Package test system LIPPKE 2000 SL Configuration 01.08.98 (Date) 16:55 (Time) (Mode L = Loss Mode) mbar (Pressure unit) (Test pressure) (Compensation time) (Test time) (Pressure loss (burst after x seconds Figure 6.4: Data output at the end of pressure loss test Empty line if package bursts within compensation time.
  • Page 31: Figure 6.5: Data Output, Burst Test

    Package test system LIPPKE 2000 SL Configuration ”OUtput Continuous”. Upon the start of the test, the test parameter data ”Date and time”, mode (“B”urst or “L”oss), pressure rise rate (burst test) or test pressure, test time and compensation time (pressure loss test) will be output.
  • Page 32: Figure 6.6: Data Output, Pressure Loss Test

    Package test system LIPPKE 2000 SL Configuration 01.08.98 (Date) 16:55 (Time) (Mode) 00100 (Test pressure) 00015 (Compensation time) 00030 (Test time) Start 00000 (Measured pressure) 00001 00001 00002 00002 00002 Readjustment time 00100 00100 00101 00100 Start Test 00100 00100...
  • Page 33: Printer Configuration

    Package test system LIPPKE 2000 SL Configuration 6.5 Printer configuration When using a printer, the following settings are relevant: Number of line feeds at the end of a print-out ("LineFeed 6") within a setting range of 1 to 99; in this way, you can e.g. match the print- out to label size Form feed at the end of print-out ("FormFeed 0") yes (=1) or no...
  • Page 34: Date And Time

    Package test system LIPPKE 2000 SL Configuration • By pressing the ”STOP” key, you can store all changes and quit this item. • By pressing the ”TIME” key, the standard settings for the STAR TSP 442 will be reactivated (see ...).
  • Page 35: Acoustic Signal

    Display in ”mbar” (“mbar” LED lights up) “Pressure Unit mbar” Display in ”PSI” (“mbar” LED doesn’t light up) “Pressure Unit PSI” In case of pressure unit change, the LIPPKE 2000 SL control unit is set to the following default values: Pressure unit mbar:...
  • Page 36: Maintenance

    Package test system LIPPKE 2000 SL Maintenance 7 Maintenance The LIPPKE 2000 SL control unit is maintenance-free. This particularly applies to pneumatic system and electronic control system. After switching off the system, you can clean its outside surface using a moist cloth, if it becomes necessary.
  • Page 37: Error Causes

    Package test system LIPPKE 2000 SL Error causes 8 Error causes The following list of possible error causes and their remedies is not necessarily complete. Should your system reveal any errors which are not mentioned at all or just incidentally, please do not hesitate to contact us.
  • Page 38: Problems During Measurements

    Package test system LIPPKE 2000 SL Error causes 8.3 Problems during measurements If the system doesn’t build up pressure during measurement, this may have various causes which can easily be removed: prick-in point leaking pressure rise too high very porous package...
  • Page 39: Appendix

    Package test system LIPPKE 2000 SL Appendix 9 Appendix 9.1 Operating principles The following diagram shows the operating principle of the LIPPKE 2000 SL package test system: Figure 9.1: Pneumatic system Paul Lippke Handels-GmbH...
  • Page 40: Specifications And Certificate

    Package test system LIPPKE 2000 SL Appendix If pressure shall be built up in a package, both valves open, i.e. passing of air through both of them is possible. The ”SENSE” line is connected to the pressure sensor in the control unit so that the control circuit is closed by this.
  • Page 41: Figure 9.2: Sample Of Works Certificate

    Package test system LIPPKE 2000 SL Appendix Figure 9.2: Sample of works certificate Paul Lippke Handels-GmbH...
  • Page 42: Dimensions And Weights

    Width / mm Height/ mm Depth / mm Weight / kg Control unit Holding stand Table 9.1: Dimensions and weights 9.4 Spare parts list (extract) The following spare parts are available for the LIPPKE 2000 SL: Part number Unit Name of item SK 765...