Summary of Contents for Arthrex AR-3200-1040
Page 1
Cable Kit, Arthrex Shaver and Pump HD3™ The Synergy Cable Kit, Arthrex Shaver and Pump Instructions for Use provides important information for the safe operation and installation of the accessory kit. Read this User’s Guide thoroughly prior to using this system and keep it in an easily accessible place for use by all operating personnel.
Page 2
Arthrex. Arthrex reserves the right to revise this publication and to make changes from time to time in the contents hereof without obligation to notify any person of such revision or changes, unless otherwise required by law.
Synergy manual before utilizing Cable be read carefully. Kit, Arthrex Pump and Shaver. The safe and effective use of this system requires the WARNING: The safety and/or health understanding of and compliance with all of the patient, user, or a third party is warnings, precautionary notices, and at risk.
Please make sure that all each of the System Components prior to required information has been supplied. Call use. Arthrex for a RMA Number for the device return for service. 2. The electrical installation of the operating •...
Requirements: To utilize the pump and shaver setting display a facility needs to: • Have the most recent hardware version of the Arthrex pump, shaver, insufflator, and Apollo HD3™ UHD4™...
Page 11
Figure 2. System tab Once the auxiliary device use has been enabled the preferences for each surgeon must be configured. Enter System Set-Up by pressing the Maintenance Icon located in the upper left corner of Synergy tablet (as shown in Figure 1 above) and then select Surgeon Management. On the Surgeons screen as show in Figure 3 touch the name of the surgeon that wants to auxiliary device settings display and then select the “I”...
Page 12
Figure 4. Surgeon screen The screen shown in Figure 5 will appear. Select the Pump tab (as shown in Figure 5) if the pump will be used with the procedure. Figure 5. Pump tab 950-0052-01r0_fmt_en-US 12 of 20...
Page 13
Adjust the Set Pressure slider to the surgeon’s desired value (note: as you adjust the slider the numerical value appears to the right of the slider). Finally, choose one of the three options for the surgical display. If “On” is selected the pump pressure appears on the surgical monitor at all times. If “Off”...
Page 14
Enable the RF/. Adjust the Apollo probe slider to the surgeon’s desired value (note: as you adjust the slider the numerical value appears to the right of the slider). Touch Save after the changes have been made. Figure 8. Insufflator tab Select the Insufflator tab (as shown in Figure 6) if the Insufflator will be used with the procedure.
Page 15
Adjust the Set Pressure and Set Flow slider to the surgeon’s desired value (note: as you adjust the slider the numerical value appears to the right of the slider). Finally, choose one of the three options for the surgical display. If “On” is selected the insufflator flow and pressure appears on the surgical monitor at all times.
Page 16
A surgeon cannot have the pump and insufflator enabled at the same time. Based on the type of surgeon (general or orthopedic) the user needs to choose one or the other. Both cannot be enabled at the same time for a specific surgeon. After you have made the necessary pump, RF, insufflator, and shaver configuration changes and have hit Save you will return to the Surgeon screen shown in Figure 4.
Page 17
Figure 10. Auxiliary device settings on surgical display If pump, shaver, RF, or insufflator settings are not configured to automatically display on the monitor but the surgeon wants to see those settings during a procedure or if the settings display on the screen and the surgeon wants them removed, then this can be done via the Synergy tablet.
Pump, Shaver Display System, Synergy Insufflator, and Synergy Apollo devices. Your authorized Arthrex service department is the most knowledgeable about the Arthrex Medical Camera Systems and/or camera heads and will provide competent and efficient service. Any services and/or repairs done by any unauthorized repair facility may result in reduced performance of the instruments or instrument failure.
5.2 Troubleshooting Symptom Possible Cause Corrective Action • Ensure • Settings do not Pump/shaver/Apollo/Insufflator pump/shaver/Apollo/Insufflator are appear on surgical not powered on powered on • monitor Pump/shaver/Apollo/Insufflator • Confirm display not enabled pump/shaver/Apollo/Insufflator • Surgeon preferences set to not enabled in settings display settings •...
Page 20
Arthrex, Inc. Arthrex GmbH 1370 Creekside Blvd. Erwin-Hielscher-Strasse 9 81249 München, Germany Naples, FL 34108, USA +49 89 909005-0 +1 (800) 934-4404 www.arthrex.de www.arthrex.com Technical Support 1-888-420-9393 950-0052-01 Rev 0 09/2020 950-0052-01r0_fmt_en-US 20 of 20...