piCO Operating Manual Intended Use Correlation of CO (ppm) and The piCO Smokerlyzer is a breath carbon ® carboxyhaemoglobin (%COHb) monoxide monitor intended for multi-patient Breath carbon monoxide is measured use by healthcare professionals in smoking in parts per million (ppm CO) and blood cessation programmes and research.
To repeat the breath test, double click the button to return to main display and continue from step 2. To switch the monitor off, hold the button down for three seconds. The unit will auto power off after five minutes of inactivity. www.bedfont.com...
Apply calibration gas at 20ppm: Calibration pass/fail: Pass: Fail: Calibration reminder: User profile selection: Select adult user profile: Select adolescent user profile: Select custom user profile (if set via COdata software on PC): Contact Bedfont or distributor for help: D-piece reminder Return:...
CO some time. sensor. Additional technical information is available Under no circumstances should the monitor on request from Bedfont or your local be immersed in or splashed with liquid. distributor. www.bedfont.com...
piCO Operating Manual Operation Ensure that two AA batteries are correctly Attach a breath sampling D-piece and a inserted in the battery compartment. new cardboard mouthpiece to the piCO Check that all connections are pushed The single button controls all functions: firmly together.
If the zeroing fails, a ● The calibration gas will be displayed required is Bedfont 20ppm carbon (see Troubleshooting – monoxide in air. page 15). Double click the button to return to the ● Ensure the fine control valve is in the ‘off’...
Page 12
● The piCO is now calibrated and ready reading will be shown and for use. sampling will cease. ● The piCO will prompt the user to calibrate the unit again in six months. www.bedfont.com...
piCO Operating Manual Set-up: User Profiles In the set-up mode, the user can change the To change the selected profile, scroll the selected user profile from adult to adolescent pointer to the set-up symbol on the (or to a third custom profile, if one has been main display using single clicks of the set).
● Scroll the pointer until it is opposite the sounder symbol, then double click to switch the sounder on or off The large sounder symbol and the small symbol on the top line of the display will change to show whether the sounder is on or off. www.bedfont.com...
COdata icon on the PC to start the programme. ● Refer to the supplied documentation for information on how to operate the COdata software. ● The custom user profile can only be accessed from COdata www.bedfont.com...
piCO Operating Manual Frequently Asked Questions (FAQ) Frequently asked questions about breath carbon monoxide (CO) monitoring in smoking cessation: Q How quickly does CO disappear from the Q What is carbon monoxide? body when someone stops smoking? A Carbon monoxide (CO) is a poisonous A It takes about 5-6 hours for the original gas that you cannot smell or see.
Returns Procedure If your equipment requires servicing, please contact Bedfont’s Customer Repairs Depatment before returning any goods. If you did not purchase your monitor directly from Bedfont, please contact your local distributor or supplier. ● If the monitor is out of warranty and ●...
At the end of the under this warranty when such an item product’s life, contact Bedfont or your local is returned intact, prepaid, to Bedfont distributor for disposal instructions.