X-Z Lab RadPavise User Manual

X-Z Lab RadPavise User Manual

Personal radiation detector

Advertisement

Quick Links

Speci c Settings (continued)
Date & Time Settings
Locate DATE & TIME sub-menu and press OK to enter the date & time settings
Interface. Use the navigation keys to select desired date & time. Press OK to
confirm and return to the previous screen. Press OK, and the cursor will move to
the right one. Use the navigation keys to select the appropriate value. Once
cycled through all options the time will be set. Press RETURN at anytime to start
over.
Ports
Ports
Locate and lift the tab on the right-hand side of the device to
expose the ports.
Lemo Port
Connect different types of high-performance SiPM radiation
detectors to the Lemo port to extend the dynamic range and
sensitivity of the device.
MicroUSB Port
The MicroUSB port can be used to charge the device or
transfer data to a computer through RadSuite-Dose.
LEMO PORT
MICROUSB
Radiological
Detector
YSO(Ce) Scintillator + SiPM
Type of Radiation Detected
Gamma & X-Ray
Energy Range
20 keV ~ 3 MeV
Measurement Range
Dose Rate: 0.01 Sv/h ~ 1 mSv/h
Measurement Range
Integrated Dose: 0.01 Sv ~ 100 Sv
Sensitivity
34 cps/ Sv/h (∝ Cs-137)
Energy Response
20% (∝ Cs-137)
±
Dose Rate Linearity
≤ 10% up to 1 mSv/h
Accuracy
5% (∝ Cs-137)
±
Alarm Threshold
User-Set Values for Dose Rate: 1 Sv/h ~ 1 mSv/h
Alert Options
Audible (80 dB at a distance of 12 in / 30 cm),
Visual (LED & Display) & Vibrating
Alarm Response Time
< 2 Seconds
Overload Display
Activation When > 1 mSv/h
Overload Indication up to 10 Sv/h
Electrical
Mechanical
&
Communications
Via MicroUSB and
PC-Based Software (RadSuite-Dose)
Ergonomics
Upward-Facing, Tilted Screen
Power Supply
Rechargeable Lithium-Ion Battery
Standard Battery Life
Typically 240 hours in Background Field
Display
Backlit LCD
Weight
2.1 oz / 60 g
Operating Environmental
Temperature: -4 °F to 122 °F / -20 °C to 50 °C
Tolerances
Humidity: 90% RH
Accessories
MicroUSB Cable
Dimensions
2.7 x 1.8 x 0.7 in / 69 x 46 x 17 mm
FCC ID
2AC7P-110
EMI/EMC
Complies with the IEC 61526
X-Z LAB, Inc.
Bishop Ranch 6
2440 Camino Ramon
Suite #264
San Ramon, CA 94583
United States
Phone
(925) 359-6908
Fax
(925) 380-6784
Email
contact@x-zlab.com
www.x-zlab.com
All Digital
Modularity in Design
Transformative
Personal Radiation Detector
User Guide

Advertisement

loading

Summary of Contents for X-Z Lab RadPavise

  • Page 1 Typically 240 hours in Background Field Personal Radiation Detector Display Backlit LCD Weight 2.1 oz / 60 g X-Z LAB, Inc. Operating Environmental Temperature: -4 °F to 122 °F / -20 °C to 50 °C Tolerances Humidity: 90% RH Bishop Ranch 6...
  • Page 2 Introduction Operation RadPavise is a next-generation All Digital Personal Radiation Cumulative Dose Interface Speci c Settings Power On Detector (PRD) which combines four functions in one: Press the OK key for two seconds until the screen lights up and The rst line shows the cumulative measured dose for the Alarm Options •...