    OXYGEN CONCENTRATOR USER MANUAL CATALOGUE PREFACE·········································································2 IMPORTANT SAFEGUARDS················································3 RELATED USE SYMBLE DESCRIPTION································7 CHECK PACKAGE AND PACKING LIST·································8 PRODUCT OPERATING PROCEDURE··································8 PRODUCTION INTRODUCTION··········································11 STRUCTURE···································································· 11 SCOPE OF APPLICATION ··················································11 CONTRINDICATION··························································· 11 ENVIRONMENT REQUIREMENT ·········································11 TRANSPORTATION AND STORAGE CONDITIONS·················12 ADDITIONAL REMARKS····················································· 12 PURITY VALUE AND FLOW FUNTIONS CHART·····················13 TECHNICAL PARAMETERS ················································14 COMPONENTS NAME AND FUNTIONS DESCRIPTION···········15...
  • Page 4: Preface

    OXYGEN CONCENTRATOR USER MANUAL PREFACE Dear user: Thank you for your support!Hope you can be our satisfied user. This manual describes the notices, operation steps, basic functions , technical parameters, basic troubleshooting, return repair instructions and other content, so that you will be familiar with the products and operate the machine well.
  • Page 5: Important Safeguards

    OXYGEN CONCENTRATOR USER MANUAL IMPORTANT SAFEGUARDS DANGER ( If broke the regular operation , may arise serious injury or death.) The measures to decrease burn, electric shock, fire or human injury risk. Avoid using the machine when take bath.If needed , Please follow up the doctor, and the oxygen machine must be used in another room 2.5 meters away outside the bathroom.Suggested that the oxygen p nasal oxygen tube is less than 11 meters and keep...
  • Page 6 OXYGEN CONCENTRATOR USER MANUAL oxygen supply device also configure other spare (such as oxygen bottles, oxygen bag etc.). ◆ This product can not be used to maintain any life, the patients with aerobic treatment are suggested to follow up your doctor's guidance to choose the oxygen flow and time when using this machine.
  • Page 7 OXYGEN CONCENTRATOR USER MANUAL CAUTIONS(Notice to the following content to avoid breaking the machine) ◆ Do not open the machine shell and internal boss to do the maintenance. If users found some problems, or found abnormal alarm, do not disassemble the repair, should contact the dealer or manufacturer immediately.
  • Page 8 OXYGEN CONCENTRATOR USER MANUAL gas in the rated flow, the purity can up to 90%. ◆ Rotary flow control knob can not be forced too hard, or easy to damage the spool. When the flow control knob is turned on to the highest, but the flow indicator is zero, please turn off the power and then check the fault.
  • Page 9: Related Use Symble Description

    OXYGEN CONCENTRATOR USER MANUAL ◆ Please select the suitable humidification bottle , rotate and fix in the machine as required. Do not use under the revolving situation.(Refer to the humidified bottle connection indicator diagram). ◆ When use the humidification bottle , pay attention to add water as the highest and lowest water level marked on the bottle body.
  • Page 10: Check Package And Packing List

    OXYGEN CONCENTRATOR USER MANUAL CHECK PACKAGE AND PACKING LIST When you receive the product, please open the package carefully. This product is equipped with upper and lower foam protection cover, such as the protective cover is damaged, please immediately check whether the product is damaged or not.
  • Page 11 OXYGEN CONCENTRATOR USER MANUAL Intake Air Connector Safety Valve(Inside) Oxygen Exit Highest Water Level Bottle Body Lowest Water Level (Figure1) a. Put into the distilled water as the water level instructions.Fix the cover to the body (Figure 1) ; Fix the humidifer by screw, please take off when don’t (Figure 2)...
  • Page 12 OXYGEN CONCENTRATOR USER MANUAL the purity .When the machine is on,the instruction light will shin for three time. Several minuates later, the yellow low oxygen purity alarm light will got,leaving the power supply and normal light shinning. 5、 According to the output flow, the flow control knob is turned clock wise to adjust the large flow rate and turn clockwise to turn down the flow.
  • Page 13: Production Introduction

    OXYGEN CONCENTRATOR USER MANUAL PRODUCTION INTRODUCTION Oxygen concentrator refers to an oxygen concentrator used at home, nursing homes or patient care center to provide low flow oxygen therapy. Our oxygen generator product is adopted the pressure swing adsorption principle, air pressure and pressure, after the molecular sieve nitrogen and oxygen molecules in the air separation, oxygen retention, nitrogen discharge.
  • Page 14: Transportation And Storage Conditions

    OXYGEN CONCENTRATOR USER MANUAL 3. Atmospheric Pressure:86 kPa~106 kPa ; 4. No corrosive gas and strong magnetic field in the surrounding environment. TRANSPORTATION AND STORAGE CONDITIONS 1.Range of Environmental Temperature:- 20℃ ~ +55℃; 2.Range of Relative Humidity : ≤ 93% ; without no condensation phenomenon;...
  • Page 15: Purity Value And Flow Funtions Chart

    Oxygen Concentrator is for non permanent installation equipment. Altitude: 0 meter to 2000 meter,oxygen purity ≥90%, 2001 meter to 4000 meter, ≤90% PURITY VALUE AND FLOW FUNTIONS CHART CP101 output pressure=0,Purity and value flow funtions chart(Figure 4) Figure 4...
  • Page 16: Technical Parameters

    OXYGEN CONCENTRATOR USER MANUAL TECHNICAL PARAMETERS Model CP101 Working Voltage 230 V~ ± 10% / 50 Hz ± 1 Hz Rated Power (VA) Output Pressure=0/7kPa, 1~10 the flow range(L/min) Output Pressure=0,the purity value(within 10min, Oxygen flow=1~10 L/min,purity≥90% up to the standard)...
  • Page 17: Components Name And Funtions Description

    OXYGEN CONCENTRATOR USER MANUAL Compressor safety valve release pressure:250kPa±50 kPa Oxygen output temperature:≤46℃ COMPONENTS NAME AND FUNTIONS DESCRIPTION (Figure5) (Figure 6) * The following Component description refer to figure 5, figure6 * 1. Oxygen outlet:Connect to the humidified bottle or nasal oxygen tube by connecting tube 2.
  • Page 18 Notice : Test the oxygen flow meter and make sure the knot in the regular tick mark(CP101 Max flow 5L/ min,do not let the knot over 5liter). The oxygen flow is so important so do not add or down the flow, just refer to the flow the doctor required.
  • Page 19 OXYGEN CONCENTRATOR USER MANUAL * The following parts are illustrated with reference to Figure 8 * 10.LED Display(Reference figure 8) 11a 11b a. Green power indicator light-Start operating b. Green indicator light-Oxygen output normally c. Green indicator light-Low Purity d. Red indicator light-Contact the manufacturer e.
  • Page 20: Cleaning And Maintenance Instructions

    OXYGEN CONCENTRATOR USER MANUAL Note:All the alarm state of the oxygen making machine belongs to the low priority. The alarm system have been set up in the factory, the user cannot change the alarm system settings. CLEANING AND MAINTENANCE INSTRUCTIONS 1、...
  • Page 21: Fault And Repairing Comparison Table

    OXYGEN CONCENTRATOR USER MANUAL 2、 Do not accept the the failure machine due to unauthorized modification or installation of other features. The following conditions are not free of maintenance: 1、 Products over the warranty period; 2、 Failure to operate not as manual requirements; 3、...
  • Page 22 OXYGEN CONCENTRATOR USER MANUAL supplier. Please don't try any other repairs. Symptom Possible Cause Remedy 1. Power cord not Check power connection at the wall outlet. On 230 properly inserted voltage units, check the back of the unit. Unit does not into wall outlet.
  • Page 23: Electromagnetic Immunity

    Guidance and manufacturer’s declaration – electromagnetic immunity The CP101 OXYGEN CONCENTRATOR is intended for use in the electromagnetic environment specified below. The customer or the user of the CP101 OXYGEN CONCENTRATOR should assure that it is used in such an environment.
  • Page 24 OXYGEN CONCENTRATOR USER MANUAL Portable and mobile RF communications equipment should be use no closer to any part of the CP101 OXYGEN CONCENTRATOR including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 25: Environment Protection

    RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CP101 OXYGEN CONCENTRATOR is used exceeds the applicable RF compliance level above, the CP101 OXYGEN CONCENTRATOR should be observed to verify normal operation.
  • Page 26: Manufacturer Information

    Building A7,5th Floor, Hantian Science City, Shenhai Road, Nanhai, Foshan,Guangdong, China 528200 +86-757-86678238, +86-757-86678231 Shenzhen Yamind Medical Technology Co.,Ltd ECE TIBBİ CİHAZLAR VE MEDİKAL SAN. TİC. A.Ş İTHALATÇI: Sümer Mahallesi Prof. Dr. Turan Güneş Cad. The İstanbul Veliefendi Sit. G1 Ticaret Apt.