Advertisement

Blood Pressure Monitor
User Manual
EBP-020

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EBP-020 and is the answer not in the manual?

Questions and answers

Summary of Contents for EASYHOME EBP-020

  • Page 1 Blood Pressure Monitor User Manual EBP-020...
  • Page 2 Dear user Thank you for choosing the easy@Home blood pressure monitor (EBP-020). Please read this manual carefully before use and keep for future reference. The blood pressure monitor is suitable for adults. People 12~18 years old should be accompanied by an adult when taking measurements.
  • Page 3: Table Of Contents

    Contents 1. Indications for Use..............1 2. Monitor Components ..............1 3. Before Use ................3 4. Setting Up Your Device ............. 4 5. Measurement ................7 6. About Blood Pressure ............. 11 7. Contraindications, Precautions, Warnings and Prompt Instructions ......................
  • Page 4: Indications For Use

    1. Indications for Use The blood pressure monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of an adult via the non-invasive oscillometric method(Oscillometric Method is measured by establishing the relationship between systolic blood pressure, diastolic blood pressure, mean blood pressure and cuff pressure oscillation wave.) at medical facilities or at home.
  • Page 5 Display Screen User icon Low battery icon Date / time display Unit of Measurement Systolic mmHg pressure Diastolic Blood Pressure pressure Level Indicator Heartbeat icon Memory icon Pulse Memory Number Arrhythmia Indicator Average value Symbol Note: Arrhythmia Indicator will be displayed when the user has tested for arrhythmia.
  • Page 6: Before Use

    3. Before Use (1) Install the Battery : Open the battery cover and install the batteries in the battery comparement. The battery compartment is positioned at the bottom of the electronic blood pressure monitor. (a) Open the battery cover according to the method shown in the figure above.
  • Page 7: Setting Up Your Device

    Do not use any expired batteries. If the product is not used for a long time (over 3 months), please take out the batteries to avoid battery corrosion and danger to device. (3) USB connection for power supply Optional (Not Included): The product’s power can be supplied through a DC 5V external power supply through the USB port.
  • Page 8 While the device is off, press the "User " key for 3 seconds, the screen displays "User 1 " or "User 2 ", and flickers continuou- sly. Press the "Memory " key to switch to the other user. (System default user: User 1 press the “Memory”...
  • Page 9 e.Press the "Memory " key to increase the value, between 01 and 31. Press the "User " key for confirmation and enter the time setting mode. Adjust Adjust Setting Setting Confirm Confirm Adjust Setting Confirm press the Adjust Setting “Memory”key Press the Confirm "User"...
  • Page 10: Measurement

    (system default unit: mmHg). At this time, pressing the "Memory " key will switch the unit selection to the unit kPa. After selecting the unit, press the "User " key again to confirm. mmHg mode KPa mode 5. Measurement Requirement : Please read the user manuals carefully before use.
  • Page 11 --Exerting your arm may increase your blood pressure, so make sure you are in a comfortable and relaxed state before measurement. Do not move or exercise the muscles of your arm during the measure- ment. If necessary, a cushion can be used to support your arm. --Be sure the cuff is at the level of your heart.
  • Page 12 side of the arm. 4. Tighten the cuff and secure it with the Velcro. 5. The cuff should be comfortably wrapped around the upper arm, with the elasticity space of two fingers. Remove any clothing (e.g pullover) that restricts your arm before measurement. 6.
  • Page 13 (4)Starting measurement Start measurement after wearing the cuff already in position: 1. Press the "On / Off " key. The machine will automatically return to zero. The air pump starts to inflate the cuff, and the change of air pressure in the cuff will be displayed on the screen. 2.
  • Page 14: About Blood Pressure

    (5)The memory function The blood pressure monitor can automatically store 99 measure- ments for each user, totaling 198 for 2 users, in each for two users, and automatically delete the oldest ones when the memory is full. After taking several measurements, press the "Memory "...
  • Page 15 the body. When the heart beats, it creates pressure that pushes blood through a network of tube-shaped blood vessels, which include arteries, veins and capillaries. This pressure — blood pressure — is the result of two forces: The first force (systolic pressure) occurs as blood pumps out of the heart and into the arteries that are part of the circulatory system.
  • Page 16: Contraindications, Precautions, Warnings And Prompt Instructions

    Hypotension (low blood pressure) has not been defined. It is general- ly considered that high pressure lower than l00mmHg (13. 3kPa) is defined as hypotension. 7. Contraindications, Precautions, Warnings and Prompt Instructions (1)Warnings: Do not let a child below 12 years old and the people who can’t express their intention use the device.
  • Page 17 Make sure there is no connection tube kinking before starting measurement to avoid potential injuries. For any patient, do not measure more than 3 times continuously; there should be at least 5 minutes of interval rest between any two measurements, otherwise it may cause extravasated blood. Do not apply the cuff over a wound, as this can cause further injury.
  • Page 18 Do not make measurements 1h after meals or immediately after strenuous exercise, showering, smoking, drinking alcohol, or drinking coffee or black tea. Do not use the blood pressure monitor for any purpose other than blood pressure measurement. Please use the blood pressure monitor under the guidance of doctor.
  • Page 19 Please maintain the device regularly for a longer shelf-life. (2)Precautions: Do not repair, disassemble or modify the Blood pressure monitor without permission. Do not drop or knock into the machine. Do not use the blood pressure monitor immediately after it has been in temperatures below 32℉.
  • Page 20: Q & A

    (3)Indicative description The patient is an intended operator. During blood pressure measurement, make sure the measured arm is at the same level at the heart. Indoor temperature, environment, noise, subject's body position, speech or exercise might have an effect on the blood pressure measurement.
  • Page 21 Q2.Why do I feel pain or numbness due to tightening of the cuff during measurement? During blood pressure measurement, the cuff shall tighten until the arterial blood flow is temporarily stopped. So, you might feel painful or numbness for a few seconds, which is not harmful for your body. And you can always stop the device from inflating the cuff by pressing the on/off button, or by manually removing the cuff from the arm and try again.
  • Page 22: Troubleshooting

    the blood pressure measurement. 2.Cuff position. The measured value varies with the cuff size. Partic- ularly, if the cuff is twisted round the elbow, you cannot obtain the correct measured value. --Please use the correct cuff and tube positioning for measurement. The arm circumference range of the enclosed cuff is 22~42cm (center of the upper arm).
  • Page 23 P is displayed in the DIA area. The inflation pressure is greater than HI is displayed in the 307±8 mmHg ( 41 kPa ±1 kPa ), DIA area Please measure again. *Trouble shooting Abnormal Measure Item to be checked phenomenon Replace the batteries with new ones or plug Check whether the...
  • Page 24: Cleaning And Disinfection

    Is the cuff wrapped Please tighten the too loose? cuff The cuff leaks The inflatable air cuff Replace with a bag of the cuff is broken new cuff If blood pressure still cannot be measured normally with the above methods, please contact the distributor and DO NOT disassemble it by yourself.
  • Page 25 Disinfection steps: 1)Wipe the machine body with a clean soft cloth lightly dampened with a small amount of the above disinfectant, and dry it immediately. 2)The body of the Blood pressure monitor can be wiped with a clean cloth lightly dampened with a small amount of 75% medical alcohol for disinfection.
  • Page 26: Storage

    -Do not clean while using the device. Do not allow water or other liquids to enter the main body 11、Storage - Do not wipe the host and cuff - Do not clean or moisten the with volatile oil, diluent or cuff.
  • Page 27 Product name Blood Pressure Monitor M odel EBP-020 Display mode Digital display Measuring Oscillographic method method Measuring Upper arm part Blood 30mmHg-255mmHg Pressure Measuring (5.3kPa-37.3kPa) value range Pulse rate 40-199 pulse rates / minute Static Pressure ±3mmHg(±0. 4kPa) value measurement...
  • Page 28: Precautions

    Service life 5 years Date of See the label environment Temperatur If the product is 41℉-104℉ e condition stored or used outside the Humidity 15%-90%RH specified Working condition temperature and environment humidity range, it A tmospheri 70kPa-106k Pa will not work c condition properly.
  • Page 29: Statement Of Electromagnetic Compatibility

    Guidance and manufacturer´s declaration – Electromagnetic emission The Model EBP-020 Blood pressure monitor is intended for use in the electromagnetic environment specified below.The customer or the user of the Model EBP-020 Blood pressure monitor should assure that it is used in such an environment. Emissions...
  • Page 30 The Model EBP-020 Blood pressure monitor is intended for use in the electromagnetic environment specified below.The customer or the user of The Model EBP-020 Blood pressure monitor should assure that it is used in such an environment. Immunity test IEC 60601...
  • Page 31 80MHz to 800MHz 800MHz to 2.7GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters(m). 10 V/m Radiated RF Field strengths from fixed RF 80 MHz to 10 V/m IEC 61000-4-3...
  • Page 32 RF communications equipment and Model EBP-020 Electronic Blood Pressure Monitor The Model EBP-020 Series Blood pressure monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Model EBP-020 Series Blood pressure monitor...
  • Page 33 Manufactured for Easy Healthcare Corporation 360 Shore Dr.Unit B Burr Ridge, Illinois, 60527, UNITED STATES Questions or comments? Please call toll-free: 1-855-822-6999 M-F 9 a.m-5 p.m. CST E-mail: service@healthcare-manager.com...

Table of Contents