SCI 260 Series Quick Start Manual

SCI 260 Series Quick Start Manual

Ground bond testers

Advertisement

Quick Links

QUICK START GUIDE
GROUND BOND TESTERS
SAFETY CHECKLIST
S
urvey the test station. Make sure it is safe & orderly.
A
lways keep unqualified/unauthorized personnel away from the test area.
F
amiliarize yourself with safety protocols in the event of a problem.
E
xercise caution and never touch products or connections during a test.
T
rain operators. Connect the return lead first and never touch clips directly.
Y
ou should always know when a test is being performed.
WARNING:
ELECTRICAL SAFETY TESTING. AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND
CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK. TO PREVENT ACCIDENTAL
INJURY OR DEATH, THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HANDLING
AND USING A TEST INSTRUMENT.
260 Series
M od el s: 264, 266
THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH
See How Simple
Using Your Tester
Can Be
Watch the Quick
Start Video

Advertisement

loading

Summary of Contents for SCI 260 Series

  • Page 1 QUICK START GUIDE 260 Series GROUND BOND TESTERS M od el s: 264, 266 SAFETY CHECKLIST See How Simple urvey the test station. Make sure it is safe & orderly. Using Your Tester Can Be lways keep unqualified/unauthorized personnel away from the test area.
  • Page 2 TESTER SETUP WARNING: LOCATE A SUITABLE TESTING AREA WITH A THREE-PRONG, GROUNDED OUTLET. BE SURE THAT YOUR THREE-PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING. READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST. Model 266 Model 264 266 Rear Connect the female end of the standard NEMA style line power cord into the...
  • Page 3 SETTING TEST MEMORIES If you wish to have multiple test setups, you will need to begin by choosing a Memory Location PROGRAM A MEMORY (M1 - M5) for each test. Home screen. Turn the yellow rotary knob to scroll to M1 to edit Memory 1. (M2 = Memory 2, etc.) When M1 is selected.
  • Page 4: Test Connection

    DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST. Setting up a test connection on the 260 Series is simple. Follow the instructions below to connect your DUT to a tester. Hook the black return test lead (P/N 99-10724-01) to the black RETURN terminal on the front panel of the tester.
  • Page 5: Test Results

    CONDUCT A TEST WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST. Connect the Interlock Disable Key (P/N 99-10040-01) to the Signal Input connector on the rear panel of the tester. If you’re not utilizing a DUT enclosure (P/N 99-10715-01 or P/N 99-10599-01) or other safety device, the Interlock Disable Key is required in order to run a test.
  • Page 6: Personal Protection Equipment

    KEEP YOUR OPERATORS SAFE WITH OUR PERSONAL PROTECTION EQUIPMENT One of the best ways to prevent injury is to ensure that your test station is set up safely and securely. Test Verification Box TVB-2 WHY IS THIS SAFE? Ensures that the failure detectors of a Slaughter electrical safety tester are functioning properly.
  • Page 7 SHIP IT Due for Calibration? Our new ship it safe program provides SAFE the packaging and peace of mind you need to ensure your tester arrives in perfect condition. Call us when you’re PROGRAM ready to return your tester for calibration or repair and we’ll overnight a custom shipping box for a flat fee of $100.00.
  • Page 8 See How Simple Using Your Tester Can Be Watch the Quick Start Video Testers For Electrical Safety Compliance 28105 N. Keith Drive, Lake Forest, IL 60045 USA +1-847-932-3662 • +1-800-504-0055 • Fax +1-847-932-3665 • hipot.com • info@hipot.com © 2019 SCI V1.0 5/19...