GIMA. GIMA reserves the right to change, cancel or otherwise amend the data contained in this document at any time and for any reason without prior notice inasmuch as GIMA is constantly seeking new solutions which lead to product evolution.
Page 3
7. After one hour, tighten the tie-rods again using the prescribed tightening torque. Chemical anchoring: drill the wall using the template as indicated above. Insert the resin inside the 8 drilled holes and fill the hole, carefully following the manufacturer’s instructions. GIMA recommends using HILTI HIT-HY 270 resin or similar products.
Boxes containing the Product together with operator’s manual Transport is made by GIMA or any road-hauler as long as in compliance with the following characteristics: Temperature (°C): -15 / +60; Humidity: 10 / 75 %; Atmospheric pressure (hPa): 500 / 1060.
Page 5
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 5 of 20 Do not stack packaging Weight of packaging Protect from rain Weight of packaging ‘I’ ON power Weight of packaging ‘O’ Stand-By OFF power ‘L’ ‘N’ Line lead connection point Neutral lead connection point...
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 6 of 20 1.4 EC Declaration of Conformity of the Manufacturer The company: RIMSA P. LONGONI S.r.l. Via Monterosa, 18/20/22 - 20831 SEREGNO (MB) – ITALY declares under its own responsibility that the Product (Medical device for observation and diagnosis): APPLY LABEL made by RIMSA P.LONGONI S.r.l., complies with Annex VII of Directive 93/42/EEC dated 14/05/1993, enforced in...
13. In case of failure to return a part requested by GIMA, the cost of the component part will be charged. 14. GIMA cannot accept returns from end users or in any case from parties other than the buyer.
The product is supplied complete with lamp, sterilisable handpiece, bar tube 960 mm long, bar plate, counter-plate for fastening to the ceiling, bar cover with relative safety ring and nuts for bar fastening. GIMA does not provide any kind of anchoring for fastening the plate to the ceiling. Such equipment must all be provided by the installer.
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 9 of 20 in the terminal marked with the ground symbol by means of the nut. Product power protection is ensured by an input fuse (L, N) of the T1AH 250V 5X20 type, which are already connected in the switchboard.
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 10 of 20 3.3 Safety conditions (secondary effects) - Do not direct the light source into the patient’s and/or operator’s eyes. - Obligation to adequately protect the patient’s eyes. Failure to follow such precautions could cause glare and potential damage to the retina. - Never place and/or hang anything on the Product.
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 11 of 20 5 Cleaning and disinfecting 5.1 Cleaning the Product Switch off the Product by detaching the main switch and make sure it cannot be switched back on. Protect the Product from water spray and detergents and do not clean it with liquids. Leave the lamp body to cool down.
If the product still does not work, contact the assistance service. See par. 10. If, after adjustment, the product still fails to remain in The Product does not remain in position position, contact the GIMA after-sales service. The light flickers Contact the after sales service...
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 13 of 20 6.6 Spare parts list only use original GIMA parts WARNING: Description Order code Fuse T1AH 250V ‘5x20’ Z400208 Sterilisable grip Z100848 Drawing 190 To access the fuses, loosen the 3 dowels (1) of the ring (2). Remove the bar cover (3) by pulling downwards.
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 16 of 20 9 EMC Declaration The Product has been tested according to EN60601-1-2 standard to ensure correct electromagnetic compatibility. Portable and mobile RF-communications equipment can affect the Product. The Product should not be used adjacent with other equipment and that if adjacent use is necessary the Product should be observed to verify normal operation.
Page 17
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 17 of 20 Test level to Electromagnetic environment - Immunity test Conformity level IEC 60601-1-2 directives Electrostatic discharge +/- 6 kV contact +/- 6 kV contact Floors should be wood, concrete or ceramic (ESD) +/- 8 kV air +/- 8 kV air tile.
Page 18
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 18 of 20 Test level Conformity level Immunity test Electromagnetic environment - directives to IEC 60601-1-2 Portable and mobile RF communications equipment should be used no closer to any part of the Product, included cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Page 19
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 19 of 20 Recommended separation distance between portable an mobile RF communications equipment and the Product The Product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Product can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Product as recommended below, according to the maximum output power of the communications equipment.
MO080-IT 12/09/2017 Operator’s Manual Rev.3 Page 20 of 20 10 Clutch adjustment The Product is sold balanced and does not require further adjustment. Nevertheless, if the movements of the arms around the rotation joints becomes too stiff or too loose over time, such as to prevent the device remaining in position, the different clutch systems can be adjusted to restore correct stability.