Chattanooga MONTANE Series User Manual
Chattanooga MONTANE Series User Manual

Chattanooga MONTANE Series User Manual

2/3/5/7 section tables, examination tables, bobath tables

Advertisement

MONTANE TABLE LINE
User Manual
Moving Rehabilitation Forward™
2 SECTION TABLES
3520 Tatras 2 Section 120V (2 wheels/2 legs)
3521 Tatras 2 Section 120V (4 wheels)
3522 Tatras 2 Section 230V (2 wheels/2 legs)
3525 Tatras Gerlach 2 Section 230V (4 wheels)
3 SECTION TABLES
3530 Atlas 3 Section 120V (2 wheels/2 legs)
3531 Atlas 3 Section 120V (4 wheels)
3532 Atlas 3 Section 120V (Postural Drainage–2 wheels/2 legs)
3533 Atlas 3 Section 120V (Postural Drainage–4 wheels)
3535 Atlas 3 Section 230V (2 wheels/2 legs)
3536 Atlas 3 Section 230V (4 wheels)
3537 Atlas 3 Section 230V (Postural Drainage–2 wheels/2 legs)
3538 Atlas 3 Section 230V (Postural Drainage–4 wheels)
5 SECTION TABLES
3550 Alps 5 Section 120V (2 wheels/2 legs)
3551 Alps 5 Section 120V (4 wheels)
3552 Alps 5 Section 120V (Postural Drainage–2 wheels/2 legs)
3553 Alps 5 Section 120V (Postural Drainage–4 wheels)
3555 Alps 5 Section 230V (2 wheels/2 legs)
3556 Alps 5 Section 230V (4 wheels)
3557 Alps 5 Section 230V (Postural Drainage–2 wheels/2 legs)
3558 Alps 5 Section 230V (Postural Drainage–4 wheels)
7 SECTION TABLES
3570 Andes 7 Section 120V (2 wheels/2 legs)
3571 Andes 7 Section 120V (4 wheels)
3572 Andes 7 Section 120V (Postural Drainage–2 wheels/2 legs)
3573 Andes 7 Section 120V (Postural Drainage–4 wheels)
3575 Andes 7 Section 230V (2 wheels/2 legs)
3576 Andes 7 Section 230V (4 wheels)
3577 Andes 7 Section 230V (Postural Drainage–2 wheels/2 legs)
3578 Andes 7 Section 230V (Postural Drainage–4 wheels)
EXAMINATION TABLES
3590 Dolomites 2 Section 120V Examination
3591 Dolomites 2 Section 120V Examination
3592 Dolomites 2 Section 230V Examination
3593 Dolomites 2 Section 230V Examination
BOBATH TABLES
3580 Taurus 1 Section 120V (4 Wheels)
3581 Taurus 2 Section 120V (4 Wheels)
3582 Taurus 1 Section 230V (4 Wheels)
3583 Taurus 2 Section 230V (4 Wheels)
DJO is an ISO 13485 Certified Company

Advertisement

Table of Contents
loading

Summary of Contents for Chattanooga MONTANE Series

  • Page 1 MONTANE TABLE LINE User Manual Moving Rehabilitation Forward™ 2 SECTION TABLES 3520 Tatras 2 Section 120V (2 wheels/2 legs) 3521 Tatras 2 Section 120V (4 wheels) 3522 Tatras 2 Section 230V (2 wheels/2 legs) 3525 Tatras Gerlach 2 Section 230V (4 wheels) 3 SECTION TABLES 3530 Atlas 3 Section 120V (2 wheels/2 legs) 3531 Atlas 3 Section 120V (4 wheels)
  • Page 2: Table Of Contents

    TABLE OF CONTENTS 1. FOREWORD ........1 6.
  • Page 3: Foreword

    FOREWORD Montane Table Line This manual has been written for the owners and operators of the Montane Table Line . It contains general instructions for operation, precautionary instructions, and maintenance recommendations . In order to obtain maximum life and efficiency from your Montane Table, and to assist in the proper operation of the unit, read and understand this manual thoroughly .
  • Page 4: Precautions

    PRECAUTIONS Montane Table Line PRECAUTIONARY INSTRUCTIONS The precautionary instructions found in this section and throughout this manual are indicated by specific symbols . Understand these symbols and their definitions before operating this equipment . The definitions of these symbols are as follows: BIOHAZARD CAUTION Text with a “BIOHAZARD"...
  • Page 5 PRECAUTIONS Montane Table Line CAUTION CAUTION • This table should be operated, transported, and stored in • Read, understand, and practice the precautionary and operating instructions temperatures between -10° C to+ 45° C, with max humidity of 75% found in this manual . Know the limitations and hazards associated with your (no precipitation) .
  • Page 6 PRECAUTIONS Montane Table Line WARNING WARNING • Never place your hands and feet near the working mechanism of the table • This table should be operated by a clinician that has read and fully understands when making any and all adjustments to height or table sections . Moving parts this document .
  • Page 7: Safety Precautions

    SAFETY PRECAUTIONS Montane Table Line CAUTION Do not put a load on the head and foot sections as shown below. Do not overload Montane Table . Patient weight may not exceed 150 kg. In case the electrical system or the actuator cover comes in contact with liquid of any kind, immediately pull the power supply plug out of the mains socket and contact the manufacturer’s service department .
  • Page 8 SAFETY PRECAUTIONS Montane Table Line CAUTION Do not lower the table with its front section folded down! Lowering the table to the lowest possible position while its head section is deflected or in the down position may cause a collision between the head section and the lower frame elements.
  • Page 9: Montane Table Line

    MONTANE TABLE LINE Montane Table Line TATRAS 2 SECTION TABLE 69 cm ( 27 in ) FOOT SECTION 46 cm ( 18 in ) HEAD SECTION 201 cm ( 79 in ) 155 cm ( 61 in ) HEAD REST RELEASE HANDLE CASTER CASTER UP / DOWN CONTROL BAR...
  • Page 10: Atlas 3 Section Table

    MONTANE TABLE LINE Montane Table Line ATLAS 3 SECTION TABLE 69 cm (27.2 in) 46 cm (18.1 in) POSTURAL DRAIN SWITCH FOOT SECTION CHEST SECTION 54 cm (21.3 in) HEAD SECTION 201 cm (79.1 in) FOOT SECTION ELEVATION HANDLE 101 cm (39.8 in) HEAD REST RELEASE HANDLE CASTER...
  • Page 11: Alps 5 Section Table

    MONTANE TABLE LINE Montane Table Line ALPS 5 SECTION TABLE 69 cm (27.2 in) 46 cm (18.1 in) FOOT SECTION CHEST SECTION HEAD REST RELEASE HANDLE 54 cm (21.3 in) HEAD SECTION 201 cm FOOT SECTION ELEVATION HANDLE (79.1 in) HEAD REST RELEASE HANDLE 101 cm (39.8 in)
  • Page 12: Andes 7 Section Table

    MONTANE TABLE LINE Montane Table Line ANDES 7 SECTION TABLE 69 cm (27.2 in) 46 cm FOOT SECTION (18.1 in) CHEST SECTION POSTURAL DRAIN SWITCH SHOULDER WING 54 cm (21.3 in) ARMREST SECTION 201 cm (79.1 in) HEAD SECTION FOOT SECTION ELEVATION HANDLE 101 cm (39.8 in)
  • Page 13: Dolomites 2 Section Examination Table

    MONTANE TABLE LINE Montane Table Line DOLOMITES 2 SECTION EXAMINATION TABLE 69 cm ( 27 in ) 69 cm ( 27 in ) SIDE RAILINGS HEAD SECTION FOOT SECTION 201 cm ( 79 in ) PAPER TOWEL HOLDER 129 cm ( 51 in ) HEAD REST RELEASE HANDLE CASTER...
  • Page 14: Taurus 1 Section Bobath Table

    MONTANE TABLE LINE Montane Table Line TAURUS 1 SECTION BOBATH TABLE 120 cm TABLE TOP ( 47 in ) 200 cm ( 79 in ) CASTER CASTER...
  • Page 15: Taurus 2 Section Bobath Table

    MONTANE TABLE LINE Montane Table Line TAURUS 2 SECTION BOBATH TABLE HEAD SECTION 120 cm ( 47 in ) HEAD REST RELEASE HANDLE FOOT SECTION 80 cm ( 31 in ) 200 cm ( 79 in ) CASTER CASTER...
  • Page 16: Design And Function

    DESIGN AND FUNCTION Montane Table Line COMPONENTS OF THE MONTANE TABLE The frame of the table is made of welded steel sections coated with powder paint . It includes the following components: 1. Bottom Frame—A support base for the whole unit with a foot-operated height adjustment switch assembly –...
  • Page 17: Nomenclature

    NOMENCLATURE Montane Table Line DESCRIPTION OF DEVICE MARKINGS The markings on the Montane Table Line are assurance of its conformity to the highest applicable standards of medical equipment safety and electromagnetic compatibility . One or more of the following markings may appear on the device: Class II Equipment Meets Directive 93/42/EEC...
  • Page 18: Specifications

    SPECIFICATIONS Montane Table Line TABLE SPECIFICATIONS Parameter Voltage Frequency, 100-240/50-60 Hz/3-1,5 A Current Consumption: Fuses: None Duty Cycle: 2 minutes "on"/18 minutes "off" Table Length (Head to Toe): 201 cm (79 in) Table Height 49- 101 cm(19-40in) Electrical Safety Classification Class I Functionally Earthed Electrical Type:...
  • Page 19 SPECIFICATIONS Montane Table Line TABLE SPECIFICATIONS Technical Parameters of Montane Table Line Montane Table Line Technical Data Tatras Atlas Alps Andes Atlas PD Alps PD Andes PD Head Section Length Chest – of the Section Table Top Foot [cm] Section Total Head Angle...
  • Page 20 SPECIFICATIONS Montane Table Line TABLE SPECIFICATIONS – (continued) Montane Table Line Technical Data Tatras Atlas Alps Andes Atlas PD Alps PD Andes PD Width of the Standard: Approx 69cm (27in) Wide: Approx 76cm (30in) Table [cm] Total Load [kg] Not More Than 300Kg (660LBs) Not More Than 225Kg (495LBs)
  • Page 21: Technical Specifications

    TECHNICAL SPECIFICATIONS Montane Table Line TABLE 1: GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS The Montane Tables are intended for use in the electromagnetic environment specified below . The customer or the user of the Montane Tables should assure that it is used in such an environment .
  • Page 22 TECHNICAL SPECIFICATIONS Montane Table Line TABLE 2 & 3: GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS The Montane Tables are intended for use in the electromagnetic environment specified below . The customer or the user of equipment* should assure that it is used in such an environment .
  • Page 23 TECHNICAL SPECIFICATIONS Montane Table Line TABLE 2 & 3: GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY – (continued) Immunity IEC 60601 Compliance Electromagnetic Environment – Guidance Test Test Level Level Portable and mobile RF communications equipment should be used no closer to any part of the equipment*, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 24 TECHNICAL SPECIFICATIONS Montane Table Line NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies . NOTE 2: These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures, objects and people . a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy .
  • Page 25: Setup

    SETUP Montane Table Line CONTENTS OF CARTON Contents of Carton: • Montane Table After removing top and side panels of shipping carton, you may notice metal bands or plastic straps . These are used to secure various sections of the table •...
  • Page 26 SETUP Montane Table Line 1 . Select the place for the use of the table with a minimum of 80 cm 2 . A 230 VAC/50 Hz, 120 VAC/60 Hz, 100 VAC/60 Hz power supply (31 . 5 inches) of clearance on each side of the unit . connection must be available in the vicinity of the table .
  • Page 27 SETUP Montane Table Line 5 . Access lock . CAUTION Always remove the key from the lock before leaving the workstation, preventing unauthorized persons from starting the table control function . Then check it is locked, by testing the height movement of the table . Connecting the power cord Before disconnecting the power cord from the controller, make sure that the MONTANE tables are equipped with a locking mechanism for access to the height...
  • Page 28: Operation

    OPERATION Montane Table Line RELOCATING THE TABLE WARNING Moving the Table 1 . The Montane Table with the optional 4 caster system includes a foot lever (see C in Figure 3) which can be set in two positions: • Never transport a patient on the Montane Table . This table is not de- signed to support a patient during transport .
  • Page 29: Adjusting The Table

    OPERATION Montane Table Line ADJUSTING THE TABLE The height adjustment is done by means of a foot-controlled frame located at the bottom of the table . To lift the table, push the foot controller frame in the direction of the head rest (marked as B in the picture below) . To lower it, push the foot controller frame in the direction of the leg rest (marked as A in the same picture) . The table will continue to rise or fall as long as the foot controller frame is kept in its left-most or right-most position .
  • Page 30: Adjusting Table Sections

    OPERATION Montane Table Line ADJUSTING TABLE SECTIONS WARNING Adjusting the Postural Drainage Press the Postural Drainage Switch (located on both sides of the unit) to raise the Chest and Back Sections to the desired angle for Postural Drainage (illustration below) . •...
  • Page 31: Maintenance

    MAINTENANCE Montane Table Line CLEANING THE UPHOLSTERY 4 . Never use: CAUTION – cleaning paste, wax, sprays – strong detergents, solvents and cleaning agents containing solvents, cleaning preparations for natural leather . We recommend to clean and/or disinfect the upholstered surfaces and Use of such preparations may cause stiffening and breakages of the cover other parts of the tables after each use to ensure proper hygiene level .
  • Page 32: Lubricating Points

    MAINTENANCE Montane Table Line LUBRICATING POINTS Clean the metal parts with a soft, damp sponge . Wipe the parts dry immediately after cleaning . All movable joints must be lubricated every six months or when experiencing squeaks or other noise . We recommend using any available aftermarket lubricants .
  • Page 33: Service

    MAINTENANCE Montane Table Line At least once every 2 years, and each time after the breakdown/ repair of the actuator/ table automation, the user's technical service must perform or commission a table review for electrical safety . The minimum scope of the review should include: check for mechanical damage to the wiring of the table;...
  • Page 34 TROUBLESHOOTING Montane Table Line Failure indication Possible cause Action to be taken If necessary, take the patient off the table with the user's help. To restart the controller, remove the mains plug from the mains electrical outlet in the room and Exceeded the continuous work cycle The adjustment mechanism the access lock from the socket and wait for a minimum 1 h.
  • Page 35: Warranty

    WARRANTY Montane Table Line DJO, LLC (“Company”), warrants that the Montane Tables (“Products”) are free of defects in material and workmanship . This warranty shall remain in effect for two years (24 months) for the cushions and three years (36 months) for the base frame and motor from the date of original consumer purchase . If this Product fails to function during the warranty period due to a defect in material or workmanship, at the Company's option, the Company or selling dealer will repair or replace this Product without charge within a period of thirty days from the date on which the Product is returned to the Company or the dealer .
  • Page 36 DJO is an ISO 13485 Certified Company Meden-Inmed ul. Wenedow 2 75-847 Koszalin Poland T: 1.800.494.3395 USA DJOGLOBAL.COM © 2019 DJO, LLC. All rights reserved. Made for DJO, LLC DJO is an ISO 13485 Certified Company 13-4322_E English...