Solar/185 Mobile hoist User instruction Important information This user instruction must be read before the Solar/185 is used. This user instruction is intended for technicians, installers, therapists and other personnel, in connection with the use, servicing and maintenance of the Solar/185.
IM AND USE 1.1 Purpose The Solar/185 mobile hoist/lift, referred to in this manual as Solar, is designed to transfer persons in institutions such as nursing homes, hospitals, care centres and in homecare. Solar can be used for transfer between bed and chair/wheelchair, from chair to another chair and from chair to a toilet.
1.3 Before use The carer/helper is responsible for ensuring that a daily check is carried out BEFORE the Solar and accessories are used, including the following: • Check that the Solar is complete, and no parts are missing. • Check that there are no signs of rust, wear and tear or other defects. If there is any sign of a defect the Solar must be taken out of use.
1.5 Unpacking Check that all the parts are present and undamaged. Any suspicion of damage or shortage should be reported to your dealer or Winncare Nordic. We recommend that the box packaging is kept if the Solar is to be used on other sites or for return shipment in case of repair or annual check.
Page 7
3. Assembly the actuators piston rod to the lifting arm. Use the Allen key and the spanner to tighten the bolt/nut. (Fig. 1.6D). Put the cap on the nut. Fig. 1.6D 4. Assembly the handles on the mast with two screws and tighten with the Allen key.
Page 8
Before first time use and after each service check the control box must be reset push up/down buttons simultaneously on the hand control (Fig 1.6G). See section 3.1 – Using the control box. Fig. 1.6G Attach the spreader bar. Rotate the attachment point so it is pointing upwards, insert the spreader bar and rotate 90°...
1.7 Spreader bar and lifting slings The Solar is equipped with a spreader bar, which is designed for use with Winncare Nordic lifting slings. Winncare Nordic´s lifting slings have 2-5 lifting straps and are designed for lifting hooks larger than ø8 mm ( in).
2. F UNCTIONS 2.1 Pictograms and functions – control-box (1) Connecting plug - hand control (2) Connecting plug - lifting actuator (3) Connecting plug - electrical leg spreading (4) Connecting plug - power supply (5) ON – indicates that main supply is on (6) Charge –...
2.2 Pictograms and functions – hand control (1) Service – See section 3.2 (2) Battery – indicates the state of the battery (see section 2.4) (3) Max Kg – indicates an overload (see section 3.2) (4) Up/Down – These functions operates the lifting arm to move up or down.
2.4 Charging The Solar is equipped with a battery that requires regular charging. Winncare Nordic recommends that the battery is charged when the Solar/185 is not in use or every night. The Solar can be charged from the mains (wall outlet).
2.5 Transport The Solar is most safely transported using the original packing. Symbols are printed on the packaging to ensure the best information to carrier companies. 3. U 3.1 Using the Solar Solar is ready for use if all assembly instructions have been followed and the emergency stop has been released.
4. M AINTENANCE 4.1 Cleaning The Solar may be cleaned using a slightly damp cloth with a standard household cleaning agent. Chemicals and other strong/abrasive agents must not be used. It is recommended that the Solar be cleaned after each use to avoid infection. The Solar can be disinfected with household alcohol.
Lifting mechanism: Actuator runs both ways. Actuator is mounted correct. Lifts from • bottom to top with max SWL 185 kg. (408 lb). Ratchet works when lowering. Hand Control: All functions work as specified in the user guide. • • Control Box and battery: Correctly mounted.
6. T ECHNICAL DATA AND DIMENSIONS 6.1 Technical specifications Lifting properties: Load capacity: 185 kg (408 Ib) Number of lift movements: 40-60 with 80 kg (176 Ib) / 0.5 m (1.6 ft) lift Duty cycle: 10%, max, 2 min/18 min Manually emergency lowering: Electronically lowering: Electrical details:...
6.2 Dimensions Highest point Maximum reach point Lowest position a) Max. height of CSP* 1980 mm (75 in) b) Min. Height of CSP* 680 mm (27 in) c) Lifting range 1300 mm (51 in) d) Min. Dist. from wall to CSP* at max.
7. E NVIRONMENT AND QUALITY Solar has an expected lifetime of minimum 10 years under the assumption that all recommended service checks have been carried out and that the normal maintenance and care instructions have been followed. 7.1 Disposal Disposal of this Solar can be made through Winncare Nordic. After useful lifetime Winncare Nordic will receive the Solar back and split the individual fractions in an environmentally friendly way either through reuse or recycling.
8. A CCESSORIES Spreader bars 015-02128 2-points 58 cm. spreader bar 015-02126 2-points 48 cm. spreader bar 015-02106 3-points spreader bar Spreader bars with weigh scale 017-00117 2-points 58 cm. spreader bar with weigh scale 017-00144 2-points 48 cm. spreader bar with weigh scale 017-00143 3-points spreader bar with weigh scale Wall charger...
Page 20
Manufacturer: Winncare Nordic ApS Phone: + 45 70 27 37 20 9 3 / 4 2 / E Ø F C L A S S I Taarnborgvej 12 C Fax: + 45 70 27 37 19 E N I S O 1 0 5 3 5 4220 Korsoer Email: Info@winncare.dk...