MICROCON® EXC7 and EXC7-UV User Manual Read and Save these Instructions Manufactured by 749 Hope Road, Suite A- Eatontown, NJ 07724 USA Tel: 800-224-9768, 732-389-8922 Fax: 732-389-8821 www.biologicalcontrols.com 8/19/2010...
Page 3
CAUTIONS IN THIS MANUAL AND ATTACHED TO THE UNIT BEFORE STARTING INSTALLATION OR MAINTENANCE ACTIVITIES. IMPROPER INSTALLATION, ADJUSTMENT, ALTERATION, SERVICE, MAINTENANCE, OR USE CAN CAUSE FIRE, ELECTRICAL SHOCK, OR OTHER CONDITIONS THAT MAY CAUSE PERSONAL INJURY OR PROPERTY DAMAGE. Copyright 2010 Biological Controls 8/19/2010...
Interconnect Cable Assembly. When a controlled negative room-pressure is required, the system will include a ceiling diffuser and an exhaust control-duct (transition box with damper). Several low voltage (24V) remote power control and monitoring options are available. Copyright 2010 Biological Controls 8/19/2010...
T-bar channels. Four (4) small brackets located on the top of the Filter Module chassis must be used to level and support the weight of the unit using approved cable, wire, or chain hardware. Copyright 2010 Biological Controls 8/19/2010...
They can be accessed and installed through the 12” duct opening in the side of the unit. If not, the bulbs can be accessed by removing the HEPA filter. Copyright 2010 Biological Controls 8/19/2010...
AC power source that will not be interrupted during normal activity. Note that the polarity ) of the power-module 24VDC must be observed as shown in the schematic for proper operation. The module positive (+) wire is marked with a white-ink dashed line. Copyright 2010 Biological Controls 8/19/2010...
For negative-pressure environments (see the System Applications Section), adjust the damper in the transition-box to achieve the desired room-pressure. (Note: A pressure monitor will be necessary for this procedure. Wall mount and portable units are available from Biological Controls.) Copyright 2010 Biological Controls 8/19/2010...
The UV lamps require change-out every 8000 hrs (approximately 1 year) of continued use. Log the Hour-Meter readings to determine the UV maintenance schedule. The UV view-port mounted on the control panel will glow blue when the UV bulbs are active. Copyright 2010 Biological Controls 8/19/2010...
When handling a new HEPA filter, hold it by its external frame without touching the media. Temporally place cardboard over the entire room-side face of filter to protect the media during the procedure. Remove the protective cardboard just before the filter clamps are fully tightened. Copyright 2010 Biological Controls 8/19/2010...
Never attempt to replace lighted UV lamps. When changing UV lamps, use approved UV eye and skin protection. UV lamps require renewal every 8000 hours (approx. 1 year) of operation. Check the Hour-Meter Log to determine a routine maintenance schedule. Copyright 2010 Biological Controls 8/19/2010...
12 feet and the total duct-rise should not exceed 10 feet. Since the exhaust-air has been filtered it is possible to exhaust directly into an air-return area above a sealed ceiling. Copyright 2010 Biological Controls 8/19/2010...
For units out of warranty, a written purchase order must be issued to Biological Controls prior to the return. The factory will NOT accept and will refuse any merchandise returned without proper authorization. The factory is not responsible for any damage to or loss of merchandise during return shipping.