Summary of Contents for SDT SDT200
Page 1
SDT International sa-nv • Bd de l’Humanité 415 • B-1190 Brussels (Belgium) • Tel: +32(0)2 332 32 25 • info@sdt.be • www.sdtultrasound.com SDT North America • 1532 Ontario St, Cobourg, ON K9A 4J7 • Toll Free North America: 1-800-667-5325 • info@SDTHearMore.com...
Page 2
You can download the latest version of this manual by visiting the SDT International website: sdtultrasound.com Due to typical self-discharge of NiMH batteries, it is recommended to charge the SDT200 battery at least every 3 months, even if it is not used.
Table of contents Overview ............................. 4 Laser safety information ......................5 Safety precautions for the SDT200 ATEX version ............... 5 Getting started Charging the battery ........................6 Turning on & turning off the SDT200 ..................7 Plug & unplug an external sensor ....................7 Functions reference guide Taking and storing an ultrasonic measurement .................
Overview The main side The back side Marking Serial Number CE marking The rear side of the SDT200 ATEX version has a specific ATEX marking The sensor side Built-in pyrometer Activation of this functionality is optional The bottom side Reset button Insert a needle to make a reset.
Laser safety information The SDT200 is equipped with a laser class 2: Risk of injury to the eye. Never look directly to the laser beam. Never point the laser beam at a person' eye. Do not aim the laser at specular reflection surfaces.
Charging the battery The SDT200 uses an internal NiMH battery. The battery charger outlet will connect to a socket on the bottom side of the SDT200. Charging the battery of the SDT200 ATEX version must exclusively be performed outside potentially explosive environments.
Line up the red dot on the plug with the red mark on the connector. Insert the plug into the connector without any rotating movement. When an external sensor is connected to the SDT200, it is automatically selected. Howerver, you can switch between available sensors using the button.
Use the F3 button to go back to the ultrasound measurement screen using the modified settings. Note that the preferred configuration is reloaded at startup of the SDT200. Getting data Press the M button: ...
Use the F3 button to go back to the temperature measurement screen using the modified settings. Note that the preferred configuration is reloaded at startup of the SDT200. Freeze measurement Press the M button to freeze the measurement.
Review measurements stored in the SDT200 From the Measurement screen press the Enter key to access the Main Menu. Use the arrow buttons to highlight the Node Tree icon: Press the Enter button to display the list of memory locations: Note: the “+”...
Before continuing this section you must have installed and activated the DataDump software. Your SDT200 must be switched on with the Measurement screen loaded. Then connect the SDT200 to a USB socket of your computer using the Mini-USB/USB cable provided by SDT Start the SDT DataDump application on your computer.
Auto power down icon: adjust the time before device auto power down when it is unused. Reset icon: force a factory reset of the SDT200 device. Note that this does not erase saved measurements. Escape icon: to get back to the main menu. In the main menu, to get back to the...
SDT200 technical specifications Technical specifications for all SDT200 versions For SDT200 ATEX version specific technical specifications, please see next page Function Multifunction detector Display Graphic LCD with backlighting (128 x 64) Keyboard 12 function keys Ultrasonic sensor Built-in sensors Pyrometer...
Page 14
Specific technical specifications for the SDT200 ATEX version The SDT200 ATEX version is certified ATEX II 1 G Ex ia IIC T3/T2 Ga, according to IEC 60079-0, 60079-11 and 60079-26. Check room temperature for use of the SDT200 ATEX version in potentially explosive environments.
0 and 50°C – for emissivity = 1 ) For measured temperature between 120°C and 180°C : ±2°C For measured temperature between 180°C and 240°C : ±3°C For measured temperature between 240°C and 380 : ±4°C Field of view at 50% of 10° attenuation MAN.200.EN--05--SDT200-user-manual-Eng.docx 15/19...
Battery charger Charger type: Specific for SDT170, SDT200 and SDT270 NiMH battery pack Indoor use only; ambient temperature: min 0°C, max 45°C. Mode of operation: continuous. Use: Environment pollution : max degree 2. Altitude during operation: max 2000 m. 230 VAC or 110 VAC +6 % / -10 %, 50/60 Hz, 25 VA,...
For these reasons, SDT Calibration Certificates do no more mention a “calibration due date”, but simply the “last calibration date”. For user convenience, the date of the last calibration of an SDT200 or 270 instruments can also be found in the menu screens.
Warranty SDT International guarantees the SDT200 unit against manufacturing faults for a period of 2 (two) years, with the exception of the battery and accessories (charger, headphones, sensors, etc.) these are guaranteed for a period of 6 (six) months. The warranty covers all material supplied and implies the free replacement of all parts that contain a manufacturing fault.