TECHNICAL DATA
·
l RIPlE OUTPUT LABORATORY POWER SUPPLY
T
2
L,
COS/MOS AND ANALOG APPLICATIONS
MODEL TP343
A
electrical specifications:
Sources
A
&
B (1)
0·20V, 0·2.SA
Parameter
each
output
INPUT:
105·125,
58-440 Hz
OUTPUT
0·20V, 0·2.5A
10·25V,
0·1A
LOAD
REGULATION:
0.01% +
1
mV per amp
LINE
REGULATION:
0.01% +0.5 mV
STABILITY
(2)
0.02%
+1 mV
RECOVERY TIME
(3)
100
us
CURRENT
LIMIT
2%·105%
of
rated
current
TEMPERATURE
0-50'
C
TEMP. COEFFICIENT
0.02% + 300
uVl'C
OV
CROWBAR
Adj.
~
3·30V (4)
TRACKING
0.1%
+10mVby
front
panel switch
VOLTAGE CONTROLS
10 turn potentiometer
FAULT
INDICATORS
Lamp indicates short
circuit
and
overload
METERING
Dual
range volt/ammeter
(1) Sources A &
B have
one
common terminal
*SOURCE
A
*SOURCE
B
SOURCE
c
IMPROVED SPECIFICATIONS
-(
0-20
VDC,0-2.5A/
0-25 VDC,0-lA
0-20
VDC,0-2.5A/
0-25 VDC,0-lA
0-6 VDC,0-5A I
0-15 VDC,0-2.5A
-
j>_
'Independent or 0.1
%
tracking with panel switch
NOISE & RIPPLE
LESS THAN lmV P-P
source C
0·6V,
0-SA
C).1SV, 0-2.SA
0-6V,
0·5AI
0-15V,
0-2.5A
0.01
%
+ 1 mV per amp
0.01
%
+ 0.5 mV
0.02
%
+
1
mV
100
uS
2%-105%
of
rated
current
0·50'
C
0.02% + 300 uV/'C
Adj. 3·20V
10
turn potentiometer
Lamp
indicates short
circuit
and
overload
Dual
range volt/ammeter
features:
• No
derating.
All
outputs may
be
operated simul·
taneously at
full
capacity for a maximum of 140W.
• Outputs are floatin'Q
with
respect to each other
and chassis except for an
_
i
nternal
connection
between
one negative terminal and one
positive
terminal of the
20 volt
sources.
The
two 20 volt
sources may
be
operated in series
to
provide 0-40
VDC, 0·2.5A
in any
polarity
with
respect
to chassis
or left floating. The
0·1
5V
source may
be
operated
in
any
polarity
with
respect
to chassis
or
left
floating
.
• Independent
adjustment of each output voltage
utilizing 10·turn potentiometers with
.05%
reso-
lution
.
• Separate current
limiting
adjustment for
each
output.
• Separately adjustable overvoltage
crowbars
on
each output. Overvoltage
crowbars on
0·20V
out·
puts
trip
simultaneoulsy
if
either
one
operates.
(2) 24 hours at constant
line,
load and ambient temperature
•
Three taut
band
suspension dual range
meters
individually
selectable
for voltage or current
o
f
each source.
(3) To
recover
to
within 15
mV
on
nominal
for a 10%·100% load
change
(4)
Both
sources will
crowbar
if
either
one
operates
mechanical specifications:
DIMENSIONS:
8
'1•'
H
x
73A'
W
x
13
V2'
D.
WEIGHT:
20
lbs.
FINISH:
Natural anodized aluminum
panel;
blue
vinyl enamel
cabioet,
with
carrying
handle.
• A "FAULT
"
indicator
lamp
on each output
signals
overload,
short circuit,
crowbar
trip or
current
limit
operation.
•
Automatic voltage
tracking of the
0-20
VDC
sources to
:1:
0.1 %
by
means of a panel toggle
switch
.
This
switch is equipped with a locking
lever
to prevent
accidental
operation. Single
potentiometer
control of tracking outputs.
•
Outputs may be
shorted into each other in any
polarity without damage.
POWER DESIGNS
POWER DESIGNS INC.
1700 SHAMES DRIVE • WESTBURY, N
.
Y
.
11590
Tel
:
516-333-6200 •TWX 510-222-6561
POWER DESIGNS PACIFIC INC
.
3381 MIRANDA AVENUE • PALO ALTO
.
CALIF
.
94304
Tel
:
415-493-6111 •TWX 910-373-1251
Need help?