  • Page 1 Smart Remote Contact User Guide 331676H03 • Revision A • 05/19/2020...
  • Page 2 IMPORTANT Read this instruction manual. Failure to follow the instructions in this manual can result in damage to the unit, injury to operating personnel, and poor equipment performance. CAUTION All internal adjustments and maintenance must be performed by qualified service personnel. Material in this manual is for informational purposes only.
  • Page 3: Table Of Contents

    IC Statements ............3 Introduction ..............5 Package Contents............ 5 Key Features ............6 What is Smart Remote Contact? ......7 Architecture Overview ..........7 Requirements and Recommendations...... 8 Where to Find More Information? ......8 Using the Smart Remote Contact ........ 9 Placement for Best Wireless Performance ....
  • Page 4 Smart Remote Contact Set Up procedure in Web app. 18 Alert Configuration for Smart Remote Contact in Web app ................18 WEEE Compliance............20 Contact Information ............. 24...
  • Page 5: Safety Instructions

    Use only batteries recommended by Thermo disassemble or modify internal components in any Scientific. Do not use a different type of battery, manner. Thermo Scientific products do not contain such as rechargeable, alkaline and magnesium, or any internal components that require user use batteries of different brands, or even different intervention or repair.
  • Page 6: Device Maintenance

    Thermofisher’s accessory specifications may invalidate the warranty. Non-replaceable Battery The Smart Remote Contact runs on AC power (via the provided 12 V adapter). The battery is only used to maintain the wireless connection to send alerts, in case of an outage on main power.
  • Page 7: Regulatory Information

    European Regulations FCC RF Radiation Exposure This paragraph pertains to 868 MHz Statement wireless. The Thermo Scientific product (contains Bluetooth component BMD- 300, LoRa component CMWX1ZZABZ) is This equipment complies with FCC radiation exposure limits compliant with the essential requirements set forth for an uncontrolled environment.
  • Page 8 CAUTION: Any changes or modifications not expressly approved by Thermo Scientific could void the user's authority to operate the equipment. RoHS Compliance The wireless device is in compliance with the restriction of the use of certain hazardous substances in electrical and electronic equipment RoHS Directive 2011/65/EU and the amendment of (EU) 2015/863.
  • Page 9: Introduction

    This guide describes how to prepare your device up for use with the Smart-Vue Pro wireless monitoring system. The Smart Remote Contact relay offers the ideal interface between your Smart-Vue Pro LoRa-based monitoring system and your building’s alert management system.
  • Page 10: Key Features

    Key Features Table 1. Key Features • Integrates Smart-Vue Pro alarms into existing alert systems with Smart Remote Contact Applications inputs. • For use with Smart-Vue Pro LoRa-based monitoring solutions. Compatibility • Two simultaneous Smart Remote Contact outputs, configurable as “normal open” or Outputs “normal closed”.
  • Page 11: What Is Smart Remote Contact

    What is Smart Remote Contact? Smart Remote Contact are used to raise alerts if alarm conditions is detected in your Smart-Vue Pro wireless monitoring system. Alarms can be configured in the Smart- Vue Pro web application for any or all your end-point modules and/groups.
  • Page 12: Architecture Overview

    Your access to the Smart-Vue Pro web platform and application is configured and you have appropriate access rights to configure devices in the system. The use and operation of your Smart Remote Contact device • LoRa wireless coverage is available at the location where is described in the Smart-Vue Pro web application user guide.
  • Page 13: Using The Smart Remote Contact

    Placement for Best AC Adapter Wireless Performance 1. Plug the AC adapter into the Smart Remote Contact as shown below. The plug should slide firmly into the socket For optimal operation, follow these recommendations when on the device. physically placing your device: •...
  • Page 14: Battery

    Terminal Wiring Battery The terminal is wired as follows (the images are shown in the The battery in the Smart Remote Contact is used to send an Smart-Vue Pro Web application): “AC unplugged” event to the Smart-Vue Pro Web Application.
  • Page 15: Cleaning Instructions

    3. Do not use any aggressive cleaning agents or scratching cleansers that might damage your Smart Remote Contact. 4. Do not submerse the unit in any liquid, as the casing is not waterproof. Smart Remote Contact Using the Smart Remote Contact |...
  • Page 16: Device Maintenance

    When Smart-Vue Pro Webapp detects a new alarm, then webapp sends the information to Smart remote contact via Lora, if the smart remote contact is configured in alerts rules in webapp. Can we configure the relay outputs to specific alarms? Yes, this configuration will be done in Smart-Vue Pro webapp.
  • Page 17 ON option from ON/OFF to enable LoRaWAN connectivety. 3. Then in Menu  Advanced  LoRaWAN  Network, select the Thermo Scientific option and for on premise installation select “custom”. Smart Remote Contact Smart-Vue Pro Solution Installation Procedure |...
  • Page 18 • To save the settings, save and restart to be selected. Figure 11. Create Data Logger Figure 9. WiFi Configuration Page 14 | Smart-Vue Pro Solution Installation Procedure Smart Remote Contact...
  • Page 19 7. Ensure data logger is added. 11. To add a Smart Remote contact, in serial number field, add the required serial number. 8. In Sensor page, verify the associated sensor serial number shown. Figure 12. Sensor Serial number display 9. In Infrastructure, add the LoRaWAN details.
  • Page 20 15. From Equipment page, click start data logging. Note: For more detail refer the user guides of the module. Figure 20. Watch mode in progress Figure 18. Logging from Equipment Page 16 | Smart-Vue Pro Solution Installation Procedure Smart Remote Contact...
  • Page 21 Configuration  Time slots  Add Time slot. In web app, Alerts can be configured as required, so that user would receive Emails/SMS/Voice calls based on the Smart Remote Contact Smart Remote Contact set up procedure in Web app |...
  • Page 22 Days off. Users that are selected from the Available users list under particular Day off tab while creating Alert rule would only receive the alert during the Day off period. 18 | Smart Remote Contact set up procedure in Web app Smart Remote Contact...
  • Page 23 Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito www.thermofisher.com in Servizi e Supporto.
  • Page 24 If You Need Assistance: Thermo Fisher Scientific products are backed by a global technical support team ready to support your applications. We offer cold storage accessories, including remote alarms, temperature recorders, and validation services. Visit www.thermofisher.com or call: Countries Sales Services North America +1 866 984 3766...

Table of Contents