1. Warnings and Contraindications 1.1.Warnings WARNING: ALTHOUGH THE LUMA IS CERTIFIED EYE SAFE WITHOUT THE NEED FOR PROTECTIVE EYEWEAR AVOID POINTING THE LUMA AT A PERSON’S EYES. WARNING: THIS DEVICE IS INTENDED FOR VETERINARY USE ONLY. THIS DEVICE HAS NOT BEEN CLEARED FOR HUMAN USE. 1.2.Contraindications Before using the Luma Device, please notify your veterinary care provider if your pet has been diagnosed with any of the following conditions:...
2. Parts Identification 1. Power Button / Set Time Button 2. Start/Stop Treatment Button 3. Low Battery Indicator 4. Bluetooth Connectivity Indicator 5. Error Indicator 6. Device Charging Indicator 7. Optical Comb 8. Attachment: Plastic Spacer - Front end 9. DC in Jack (not shown)
3. Operating Instructions 3.1. Charging the battery Before usage, connect the AC adapter power cord to a wall socket. Connect the DC plug of the charger to the DC in jack of the device (No. 9). The charging Indicator Light (No. 6) will blink during charging. When battery is fully charged the charging light will stop blinking and stay on, until disconnected from the AC adapter.
The Luma device contains a rechargeable battery, which should be replaced only by MedcoVet service. Replacing the battery may damage the Luma device 3.2.Treatment ● The Luma device can be used with or without the spacer attachment. ● For surface wounds that exude liquids, snap the spacer front end (No. 8) onto the device.
● A location where medication is injected locally ● Thyroid gland 3.2.2.Before using the Luma Device Before using the Luma Device review the MedcoVet Safety Video at MedcoVet.com/LumaSafety Only operate the Luma in a warm and dry environment (0-35ºC). If the Luma has been stored at a temperature outside of normal operation, allow the Luma 20 minutes to acclimate to a normal operating temperature.
3.2.3.Treatment Instructions Please refer to the image below. If you have any questions, please contact your veterinary care provider or email support@medcovet.com.
WARNING: ALTHOUGH THE LUMA IS CERTIFIED EYE SAFE WITHOUT THE NEED FOR PROTECTIVE EYEWEAR AVOID POINTING THE LUMA AT A PERSON’S EYES. WARNING: THIS DEVICE IS INTENDED FOR VETERINARY USE ONLY. THIS DEVICE HAS NOT BEEN CLEARED FOR HUMAN USE. 3.2.4.Power Off To Power Off, Press and hold the Power Button (No.
Bluetooth (BT) Device is paired with the Luma app via Bluetooth Error Device is malfunctioning. Connect the app to share the issue with MedcoVet support or email support@medcovet.com for further assistance. Charging When blinking, device is charging. When on...
Prior to using the Luma, watch the treatment overview video included in the app or on the MedcoVet website. Make sure that the comb is gently massaged into the fur and sits flush against the area to be treated, so that the light exiting the comb reaches the skin.
Page 12
Treatment time will be prescribed by your veterinary care provider. Carefully follow the protocol instructions for where, how long and how often to treat. For treatment areas larger than the device spot size, divide the treatment area into sections and treat each section separately. Choose sections that do not overlap and include both healthy and treatment skin in each treatment section.
7.IMPORTANT PRECAUTIONS The Luma is classified as exempt group, or risk group 0, per the International Standard IEC 62471. This means that the device does not cause eye or skin damage. However, 1- NEVER POINT THE LUMA AT A PERSON’S EYES, INCLUDING YOUR OWN. 2- IF YOU HAVE ANY CONCERNS ABOUT SAFETY, OPERATIONS, OR EVEN A SUSPICION THAT SOMETHING DOESN’T SEEM RIGHT, PLEASE DISCONTINUE USE AND CALL YOUR VETERINARY CARE PROVIDER.
The warranty applies when the unit is employed for normal home use, in compliance with what is outlined in this User Manual. In no case can MedcoVet be held responsible for any particular damages, inherent, direct or consequent to the purchase or use of this product or for costs higher than the original cost of the product.
11.Device Labels 11.1.Manufacturers label This label displays the manufacturer, model number, serial number, and regulatory compliance declarations 11.2.For Veterinary Use Only Label This label displays a reminder that the Luma is intended for veterinary use only and is not approved for human use.
43mm Aperture 75 mW/cm Power Density IEC 62471 exempt group Device Eye Safety Class 13. Company Information MedcoVet is the veterinary brand name of Medical Coherence LLC. Address: 277 Fairfield Road Suite 336 Fairfield, NJ 07004 Phone Number: (408) 663-5351...