4 - Safety Instructions

 If the calibrator is turned on, do not leave the room without an identification or warning
about the high temperature hazard.
 Before turning the calibrator off, return the block temperature to values close to the
ambient temperature.
 Never remove the insert from the dry block or the thermo-elements from the insert,
while they are in temperatures far from the ambient. Wait until they reach the ambient
temperature so that the heterogeneous cooling of the parts does not jam each other.
If, by chance, a jamming situation happens, refer to item 8.2 - Instructions for Insert
Jamming to proceed properly.

5 - Calibrator Warning Messages

Warning
RAM ERROR
READ MANUAL
EEPROM ERROR
READ MANUAL
LOW BATTERY
UNDER / OVER
????.??°C
Page 18
Meaning
Problem in RAM
memory
Problem in EEPROM
memory
Level of battery voltage is
low
Input signal out of
specifications or scaling
range
Input sensor is open
Procedure
Turn the calibrator off and
on. If the error persists,
contact the Presys
technical support dept.
Same as the previous item
Check the calibrator power
supply
See item 1.1.1 on Input
Specifications
Check input connections
and sensor
T-650PH