User Manual Thermostat E5s with Timer - vs. 1.51
User Manual
Bath-/ Immersion
Thermostat
E5s
with Timer
medingLab
Temperiertechnik
Saalhausener Str. 29
D – 01705 Freital

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the E5s and is the answer not in the manual?

Questions and answers

Summary of Contents for medingLab E5s

  • Page 1 User Manual Thermostat E5s with Timer - vs. 1.51 User Manual Bath-/ Immersion Thermostat with Timer medingLab Temperiertechnik Saalhausener Str. 29 D – 01705 Freital...
  • Page 2: Table Of Contents

    User Manual Thermostat E5s with Timer - vs. 1.51 Table of Contents Safety Instructions ______________________________________________________________ 3 Operational Safety__________________________________________________________ 4 Conditions for safe operation________________________________________________________ 4 Thermostat liquids and hose materials ________________________________________________ 5 Unpack and Check _________________________________________________________ 5 Technical Data ____________________________________________________________ 5...
  • Page 3: Safety Instructions

    User Manual Thermostat E5s with Timer - vs. 1.51 Safety Instructions The device has been produced after the valid rules of technology. It was tested extensively before leaving the production and has left the work in safety technical and function-technical faultless condition. To guarantee safe and faultless application, the user has to take the following notes and the existing regulations for accident prevention into account.
  • Page 4: Operational Safety

    The necessary operational safety is achieved by high temperature and low-level cut- off. The thermostats type E5 und E5s are supplied with an on-off temperature control, the type E20 comes with a self-adjusting control-system and a program for 2 sets of 10 temperature-time-segments each, an RS 232- interface and an external sensor connection.
  • Page 5: Thermostat Liquids And Hose Materials

    User Manual Thermostat E5s with Timer - vs. 1.51 Please use the fuses described by the manufacturer. Short circuit of the fuse holder is forbidden. Attention: You should not run the device if strong damages are visible or load running noises are perceived.
  • Page 6: Description

    5. Description Immersion Thermostat Type E5s The Immersion Thermostat type E5s can be used for upheating liquids in any bath vessels with a maximum volume of about 20 litres. The device can be attached to containers with a clamp. Type E5s is equipped with an electronic PID-controller with a digital display for the temperature. In addition it has an independent safety controller (of class 3 / DIN 12876) which avoids excessive upheating temperatures and prevents dry-running below minimum liquid volume.
  • Page 7: Bath Thermostat Typ E5S

    User Manual Thermostat E5s with Timer - vs. 1.51 Bath Thermostat Typ E5s A complete unit of bath thermostat - ready to operate - consists of the following units: a thermostat of the "E"-series a bath a bath bridge where a cooling coil and a connecting piece can be fastened for temperature regula- tion of external vessels.
  • Page 8: Setting Into Operation

    User Manual Thermostat E5s with Timer - vs. 1.51 6. Setting into operation The following preparations are necessary for setting the thermostat into operation. Please proceed as follows: Bring the device into a firm position . Execute the mounting work required.
  • Page 9: Setting Of The Control Parameters (Second Operation Level - P-Level)

    User Manual Thermostat E5s with Timer - vs. 1.51 Key: START/STOP With this key the selected heating time is started. The display indicates the remaining time. After course of time set the timer gives a signal for 5 seconds. The timer can be switched off with the DOWN key.
  • Page 10 User Manual Thermostat E5s with Timer - vs. 1.51 Control range limitation – lower setpoint limitation Control range limitation – upper setpoint limitation The adjustment range of the setpoint can be limited in both directions. Proportion range at PID regulation With approximation of the actual value to the setpoint value the variable is reduced linearly from +-100% to 0%.
  • Page 11: Third Operation Level (A-Level)

    User Manual Thermostat E5s with Timer - vs. 1.51 Third operation level (A-level) Attention: The following control parameters are adjustable with support of the manufacturer only. Please get in touch if necessary! Access to the third control level is granted when selecting the last P-parameter on the second control level.
  • Page 12: Maintenance And Repair

    User Manual Thermostat E5s with Timer - vs. 1.51 A31: Special function at boundary or alarm With this parameter it can be selected, whether, in the case of an alarm, the indication to flash and/or the buzzer is to start. Sensor alarm is indicated independently by flashing display and the buzzer.
  • Page 13 Declaration of Confirmity medingLab Temperiertechnik We : Inhaber Hans –Jörg Kupka Saalhausener Strasse 29 D – 01705 Freital/Germany hereby declare under our sole responsibility that the products : Thermostat E5 Thermostat E5 S to which this declaration relates, are in conformity with the following Council directives of the...