Archimedes

Maintenance

Other than routine cleaning, the Archimedes has been designed to be maintenance free, however, we recommend
that your Archimedes is inspected annually by either your supplier or other suitably qualified person.
NOTE: If the Archimedes is used with the assistance of a professional carer, the Archimedes becomes classed as
'work equipment' under the 'Lifting Operations and Lifting Equipment Regulations 1998 (LOLER)' and we would
recommend that the Archimedes undergoes a thorough examination at intervals not exceeding 24 months, in
accordance with the 'Thorough Examination Schedule' - available through Mangar Health.
If required, Mangar Health can provide the above inspection within an annual service contract. Please contact our
Customer Service Department for further details.
There are no user-serviceable parts in the Archimedes. Repairs or servicing must only be carried out by your
supplier or other authorised person.
66.5cm
72.5cm
61cm
44.5cm
29.5cm
17cm
39.5cm
8
14
Seat Height
45cm (Max)
8cm (Min)

Fault Finding

If the seat fails to rise:
• Check that the hand control plug is FULLY pushed into
the actuator lead socket (fig.14).
If the seat fails to lower:
• Check to see if the battery state indicator light is
showing RED. If it is, put the hand control on charge
overnight.
• Check that the hand control plug is FULLY pushed into
the actuator lead socket (fig.14).

Specifications

Archimedes
Max.user weight:
Total weight:
Seat section only:
Base section only:
Actuator:
Seat height (Min/Max):
Actuator power input:
Environmental protection:
Hand Control (Battery)
Type:
Output (Max):
Internal protection:
Environmental protection:
Battery Charger
AC Input:
DC Output:
Internal protection: Self resetting thermal cut-out
Environmental protection: IP50
Double insulated Class II Equipment
This product is suitable for transporting in aircraft.
165kg (26st/364lbs)
10kg
5,2kg
2,8kg
2kg
8cm/45cm
12V: 7,5A
IP66
Ni-Cd
14V: 10A
Self resetting fuse
IP67
230V: 1,22VA: 50-60Hz
17.4V: 70mA