System Status Signals; Maintenance & Support; Device Versions; Accessories - Novasina PascalMaxx +/-50Z Operating Instructions Manual

Differential pressure measurement instrument for climatic control systems
Table of Contents

Advertisement

7. System status signals

*S TST*
*DONE*
CODE?
*WRONG*
*SPACE*
*RANGE*
*POWER*
*FATAL*
OVERFL
UNDERFL
ZEROING
Error messages :
A severe error is always shown with a error code on the display. This code is important for a later failure
analysis. Please indicate in such cases always the failure code
8. Maintenance & support
Maintenance/Cleaning:
If required please clean the device with a humid cloth. Do not use aggressive detergents!
Technical support:
In case of trouble, please inform your local Novasina distributor. You can find more information on :

9. Device versions

Description
PascalMaxx +/-50 Z
PascalMaxx 500 Z
PascalMaxx 2000 Z

10. Accessories

• Filter-Set for external hose connection,
• Factory calibration at 3 calibration points incl. certificate,
• External power supply Europe plug,
• External power supply USA/JP plug,
Power ON reset, performing self test procedure
Action successfully accomplished
Password entry required
Protection code input: entry of invalid code
CAL points are to close (min. 10% of the full scale for gain adjustment)
Deviation is too big, calibration not possible
Power supply out of limits
Self test has encountered a serious problem
Measurement value is beyond the upper measurement range limit (+10 Pa)
Measurement value is below the lower measurement range limit (-10 Pa)
Device is currently carrying out the zero-point calibration while a gain
calibration is tried to be performed
www.novasina.com
Article no.
260 0955
260 0091
260 0083
Messbereiche dP :
-50 ... +50 Pa
0 ... 500 Pa
0 ... 2000 Pa
Art. no.
260 0097
Art. no.
111 7603
Art. no.
252 4210
Art. no.
252 4211