Advertisement

Quick Links

CC-04
Calibrateur Volt/mA
Notice d'utilisation
ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CC-04 and is the answer not in the manual?

Questions and answers

Summary of Contents for Adel CC-04

  • Page 1 CC-04 Calibrateur Volt/mA Notice d’utilisation ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 2: Table Of Contents

    Simulate a Transmitter (SIMULATE mA) 3.2.3 Output Voltage dc (SOURCE V) 3.2.4 Output Millivolt dc (SOURCE mV) 3.2.5 Auto Ramp Function Ⅰ ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 3 (SOURCE mV dc) LOOP Power General Specifications 5、 Maintenance and Calibrati Repair Cleaning Replace the Battery and Fuse 6、 Accessories Included Ⅱ ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 4: Safety Information

    If charge the Calibrator, must be use the power adapter that is provided from Calibrator. Disconnect the power immediately and return it to detect and maintain if discover abnormal situations or excessive local temperature on charging. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 5: Symbols And Marks

    0-24mA , it can measures 0-20mA or 4-20mA current loop as well as measure voltage dc 0-100mV/0-10V. It is widely used to calibrate various loop current or voltage/current equipment. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 6: Pushbutton Functions

    It would storage the current setting status before your turn off the calibrator and keep it on when you using it again. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 7 When at full scale point, press key again to decrease the current or voltage with stepping of 25% of full scale (0-20mA/4-20mA)/(0-100mV) /(0-10V). ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 8: Display

    • 4 mA-20 mA =0%-100% (default value) • 0 mA-20 mA =0%-100% (optional) The selection will be saved until next change. 2.2 Display ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 9 Check of the full scale (0%-100%) Loop Loop power The combination of unit symbol: mA /mV/V Auto ramp type Stepping direction indicate symbol Low battery ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 10: Using

    If the voltage over 10V the LCD will show “-OL-”. The lower display area ,shows the voltage value percentage. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 11: Measure Millivolt Dc (Measure Mv)

    If the voltage over 100mV the LCD will shows “-OL-”. The lower display area ,shows the current value percentage. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 12: Measure Current Dc (Measure Ma)

    3) If the current over 24mA, the LCD will shows “-OL-”. The lower display area ,shows the current value percentage. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 13: Measure Ma With Loop Power

    Touch test lead probes to the circuit across the load or power source as shown below. If the current over 24mA, the display will show “-OL-”. The lower display area ,shows the current value percentage. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 14: Output

    , stepping up/down 25% of full scale for output value. 6) Press key to select auto ramp output type. 7) Press key to select “SpanCheck” function. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 15 Volt/mA Calibrator Using ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 16: Simulate A Transmitter (Simulate Ma)

    6) Press key to select auto ramp output type. 7) Press key to select “SpanCheck” function. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 17: Output Voltage Dc (Source V)

    , stepping up/down 25% of full scale for output value. 6) Press key to select auto ramp output type. 7) Press key to select “SpanCheck” function. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 18: Output Millivolt Dc (Source Mv)

    , stepping up/down 25% of full scale for output value. 6) Press key to select auto ramp output type. 7) Press key to select “SpanCheck” function. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 19: Auto Ramp Function

    Smooth ramp with rate of change 0%-100%-0% within 15s. Press any pushbutton (except pushbutton) or turn off the Calibrator to exit this function. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 20: The "Spancheck" Function (0%-100%)

    4 mA-20 mA (0%-100% ) [default value] ; 0 mA-20 mA (0%-100%) [optional]; To change the output span, press simultaneously. The setting value for selection will be saved until next change. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 21: Power

    3) To disable this function, rotate the knob until the display shows “OFF”. 4) The Calibrator will exit automatically about 5s later or you press the knob. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 22: Charging Function

    If low light On-site environmental,this intrument has backlight function.In normal mode, Press button , The backlight open, repeat this operations can close backlight. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 23: Technical Index

    ± (0.02% of reading+4 Counts) 4.3 MEASURE mA dc Range Resolution Accuracy Sampling resistor 0 ---- +24 mA 1 uA 10Ω ± (0.02% of reading+4 Counts) ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 24: Source / Simulate Ma Dc

    4.6 SOURCE mV dc Range Resolution Accuracy 0 ---- +100 mV 0.01 mV ±(0.02% of reading+4 Counts) 4.7 Loop Power: ≥24V ,≤26V. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 25: General Specifications

    Weight: About 400g, (including the rubber cover). Conformity to safety standards: Meet the standard of IEC 61010-1-95 CAT I, 30V; Meet the standard of EN61010-1 EN61326. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 26: Maintenance And Calibrati

    6) Insert the battery cover make sure it is securely in place (check the polarity). 7) Restore the calibrator to its holster. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 27 3) Lift off the battery cover on the back of the Calibrator as shown in fifure. 4) Remove the damaged fuse. 5) Insert the replacement fulse and restore the battery cover make sureit is securely in place. ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...
  • Page 28: Accessories Included

    Test lead (red) Test lead (black) Bending test plug wire clip (red) Bending test plug wire clip (black) Adapter(DC 10V 1A) Instruction sheet Certification ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37...

Table of Contents