Water quality monitor adjustment to alarm on rejection for nuisance alarms (3 pages)
Summary of Contents for Better Water PB2
Page 1
Better Water LLC Mediport Portable RO Operator Manual rev. Jul 2019 REA 2380...
Page 2
Better Water LLC. All rights reserved. The content of this manual is the intellectual property of Better Water LLC. It is furnished for the express use by Better Water LLC, their customers and dealers, for informational use only for operation, service, and internal training.
- 8” x 30” Water Softener with Brine Tank ……………………………………………………… 29 - Divert-to-Drain …………………………………………………………………………………… 30 OPERATION ……………………………………………………………………………………… 33 - Familiarization with Water Quality Monitor and PB2 Controls …………………………….. 34 - Setup for a Dialysis Treatment ………………………………………………………………… 35 - Flushing the PB2 ………………………………………………………………………………... 36 - Low Pressure Alarm ……………………………………………………………………………..
Page 4
- TSB2012001: PB1/PB2 Low PSI / Monitor Modification …………………………………… 56 - TSB2012002: PB2 Low Pressure Issues When In Flush ………………………………….. 62 - TSB2012003: PB2 Water Quality Monitor Adjustment to Alarm on %-Rejection for Nuisance Alarms ………………………………………………………………………………… 64 - TSB2012004: Portable RO Current Leakage Values ………………………………………. 67 - TSB2012005: Replacing the 10”...
PB2 – Mediport Portable RO Operator Manual Better Water LLC is a leading integrated manufacturer of water treatment equipment and components for the industrial, commercial and institutional Company markets. Located in Smyrna, Tennessee, Better Water LLC continues its history of manufacturing and distribution of equipment specifically designed for the renal dialysis market.
- Return Goods Authorization Request Form Introduction The Better Water LLC PB2 is manufactured to the utmost quality. With proper care, preventative maintenance, and proper use, it should provide you with a very effective means of treating water for dialysis treatments.
PB2 – Mediport Portable RO Operator Manual WARNINGS 1. It is unsafe to operate or service this device without first reading and understanding the entire Operator and Service Manuals. Keep this manual and other associated documentation for future reference. 2. Misuse, improper operation, and/or improper monitoring of this system could result in serious injury, death, or other serious reactions to patients undergoing hemodialysis treatment.
230 vac, single phase EQROPB01613UL 115 vac, single phase IMPORTANT INFORMATION FOR SUPPORT Adhered to the side of the control box of each PB2 is a label containing important information Model Serial Number relating to the specific unit, and details both the .
3. The Top Panel is hinged to fold forward to access the control box and the back of the Water Quality Monitor. 4. A hinged Cabinet Door is located on the front of the PB2 for easy accessibility to its inner components 5.
PB2 – Mediport Portable RO Operator Manual DETAILED VIEWS Water Quality Monitor Hinged Front Gauges: Panel to Filter-In, access Control Filter-Out, Box and Pump, Membrane Product Access Door to internal components Motor/Pump compartment Front View Hand Hold Power Cord Air Vent Right Side View Better Water LLC;...
Page 11
PB2 – Mediport Portable RO Operator Manual Accessories Connection & Electrical Panel Back View Hand Hold Water-In Flush On-Off Port Valve Drain Pre-Booster Port Port Disinfect Post-Booster Port Port Product-Out Port Left Side View Better Water LLC; rev. Jul 2019...
PB2 – Mediport Portable RO Operator Manual MONITOR and CONTROL SYSTEM The Control Box contains the electronic components needed to operate the RO. The front panel has gauges, as well as the following: Operate switch, Low Pressure Alarm Light, Low Pressure Alarm Reset Button, Water Quality Monitor Mute Button, as well as the Water Quality Monitor’s...
Page 13
The PB2 uses a stainless-steel Pump covered in a rubber boot to reduce noise. This works in conjunction with a 1/2" HP Motor to pump water within the PB2 loop. part#...
Operator Manual OPTIONAL ACCESSORIES: Short & Extended Carts Better Water offers two models of carts to hold the PB2 and various accessories. Both are strong and lightweight, and includes casters, two of which lock. The extended cart is also designed to transport the dialysis machine as well.
OPTIONAL ACCESSORIES: Divert-to-Drain The Divert-to-Drain will automatically divert the product water from the PB2 to drain when the water quality monitor detects a poor water condition. It includes a “Good Water Quality” Light, and connection ports for “Product Out”, “From RO”, and “To-Drain”.
20” Carbon Block filter. The internal carbon block filter can either be used or replaced with a short bowl. These Carbon Tanks come in either 8”x30” or 9”x42” sizes. NOTE: 9”x42” tanks will not fit on the PB2 carts. part# part# EQASSYCFRO.8CF830PB2...
CAUTION The short bowl SHOULD NOT BE USED if there are no additional external Carbon Tanks and/or Carbon Blocks used in the pre-treatment of the feed water to the PB2. NOTE about O-ring: The short bowl uses the same O-ring as the 10” Big Blue Filter Housing. If replacing the standard filter housing with a short bowl the existing O-ring can be used but it is recommended that a new O-ring be installed.
PB2 – Mediport Portable RO Operator Manual OPTIONAL ACCESSORIES: False Tube the for Capsule Filter If the Capsule Filter is to be replaced with an alternative external filter of greater or equal capacity a false tube can be used in place of the filter. This false tube has the same type of connections as the filter and simply installed in place of the filter.
Electrical Requirements: - It is recommended that the PB2 be plugged into a dedicated G.F.I. plug, and be protected against electrical surges and be on a stable power supply. Do not use power strips or extension cords. Failure to follow these recommendations can cause the Water Quality Monitor and the Divert-to-Drain to operate improperly.
Page 20
PB2 – Mediport Portable RO Operator Manual Better Water LLC; rev. Jul 2019 Page 16 of 80...
PB2 – Mediport Portable RO Operator Manual INSTALLATION & SET- - Unpacking and Initial Start-Up - Diagram of Water Hose Connections and Water Flow - Order of Pre-Treatment and Diagram of Water Flow - Point-to-Point Connection Details - Connection Configurations With and Without a Booster Pump and/or a Divert-to-Drain - 8”...
UNPACKING AND INITIAL START-UP CAUTION The PB2 comes from the factory, tagged and packed in a preservative solution to prevent microbiological growth. Take the necessary precautions to prevent physical exposure to this preservative when following the steps below. Treat exposure by quickly flushing with water.
Page 23
5. Prep 8”x30” Carbon Tank(s) (if equipped) a. The 8”x30” Carbon Tanks must be prepared prior to use. See instructions detailed per the “Portable Carbon Tank Quick Start Guide” or the “PB2 Operator Manual”. 6. Prep 8”x30” Water Softener (if equipped) a.
Page 24
Pre-Treatment (if any), … otherwise if no more external Pre-Treatment, connect it to the PB2 WATER IN port. 11. Connect UV to PB2 (if equipped) a. Connect a 3/8” Clear-Braided Hose from the UV OUT port to the WATER-IN port.
Page 25
PB2 – Mediport Portable RO Operator Manual 12. Connect Booster Pump (if equipped) Booster a. Connect the Electrical Cord from the Pump Booster Pump to the back of the PB2, Connection into the PB2 BOOSTER PUMP connection. b. Connect a 3/8” Clear-Braided Hose...
Page 26
Product port * Take care not to contaminate this tubing while draining. (blue tubing) 16. Verify the following before proceeding: a. The Disinfect Switch on the back of the PB2 is in the Flush position. Valve b. The Flush Valve on the side of the PB2 is in the position.
Page 27
A minimum of 90% rejection should be maintained. b. If the PB2 does not maintain greater than 90% rejection, check the source water pH. If the pH is above 8.0 this may impact the rejection of the PB2 and a DI polishing cartridge may be required.
Page 28
IMPORTANT OPERATIONAL NOTE Do not use “hydrochloric acid” based products for disinfection and/or cleaning the PB2 as these can damage the internal components and will void the Limited Warranty. Use only the manufacturer recommended products as detailed in the “PB2 Operator Manual”.
PB2 – Mediport Portable RO Operator Manual ORDER of PRE-TREATMENT and DIAGRAM of WATER FLOW POINT-to-POINT CONNECTION DETAILS PRE-TREATMENT OPTIONS DESCRIPTION INLET CONNECTION OUTLET CONNECTION 1/2" Inlet Water Hose (gray 1/2” Hose: garden hose, style 5000) 1/2" Hose: 1/2" female quick...
Page 30
PB2 – Mediport Portable RO Operator Manual PB2 PORT CONNECTIONS 3/8” clear braided from last 3/8” female quick PB2 Water-In Port (INLET) piece of disconnect Pre-Treatment 3/8” blue tubing either… … Divert-to-Drain FROM RO, PB2 Product Port 3/8” John Guest...
PB2 – Mediport Portable RO Operator Manual CONNECTION CONFIGURATIONS WITH and WITHOUT a BOOSTER PUMP and/or a DIVERT-to-DRAIN Better Water LLC; rev. Jul 2019 Page 27 of 80...
4. Connection to Devices a. Device Placement - In general it should be placed between the Source Water Inlet and the PB2. - If equipped with any external Sediment Filters it should be placed after these. - If equipped with an external Carbon Block, Water Softener, or UV Light, it should be placed before these.
4. Connection to Devices a. Device Placement - In general the Softener should be placed between the Source Water Inlet and the PB2. - If equipped with any external Carbon Tanks or Carbon Blocks it should be placed after these.
Verify that the PB2 is powered OFF, and unplugged from its electrical receptacle. b. Remove the two corner screws on top of the PB2’s cabinet, and swing open the top panel. c. Remove the two screws securing the control box cover and remove the cover.
Page 35
PB2 – Mediport Portable RO Operator Manual 3. Verify that all air has been purged from the PB2’s pre-treatment as detailed in the “Unpacking and Initial Start-Up” section. 4. Turn the OPERATE Switch to the OPERATE position. - If the PB2 goes into Low Pressure Alarm, press the LOW PRESSURE ALARM RESET Button.
Page 36
PB2 – Mediport Portable RO Operator Manual Better Water LLC; rev. Jul 2019 Page 32 of 80...
PB2 – Mediport Portable RO Operator Manual OPERATION Before you start using this device, operators must read and understand this manual in its entirety. This manual of Operator's Instructions describes in considerable detail all of the steps and procedures required to safely operate this device. With proper operation, maintenance, and care, this device should give you years of reliable service.
FAMILIARIZATION with WATER QUALITY MONITOR and PB2 CONTROLS The Water Quality Monitor is located on the front panel of the PB2’s control box, and contains the control switches and buttons for operating the PB2. The Water Quality Monitor monitors and displays vital information concerning the quality of water.
- If external carbon blocks or tanks are used, the internal carbon block should have been removed and replaced with a false cartridge. - Open the PB2’s cabinet door and obtain a sample of water from the Labcock located on the inlet solenoid valve.
… If the PB2 has not been used at all during any given day, simply connect and flush for 15 minutes. This is not required but will increase the life of the membrane.
If a low pressure condition persists, it will alarm again and additional measures should be taken to correct the issue. Water pressure must be above 20 psi before the PB2 will reset. Below are a few solutions for this condition which should be tried before contacting Technical Support.
PROCEDURE 1. Disconnect the blue product hose from the PB2’s Product Port. 2. Cut a small piece off of the blue product hose long enough to go between the PB2 and the John Guest “Y” Fitting. 3. Connect one end the short-cut piece of blue product hose to the PB2’s Product Port and the other end to the single leg of the John Guest “Y”...
- Quarterly: Low pH Clean with BWI-1000, High pH Clean with BWI-2000, and Disinfection with Minncare (in that order). - If the PB2 has been in storage and should have been loaded with preservative, then after performing the initial setup the following should be done: Low pH Clean with BWI- 1000, High pH Clean with BWI-2000, and Disinfection with Minncare (in that order).
Page 44
AAMI/ISO Standards to determine the current compatibility of the PB2 with the feed water and the suitability of the system for providing product water meeting the AAMI requirements. This should be performed annually or more often if needed but, is it is ultimately at the discretion of the physician or Medical Director.
PB2 – Mediport Portable RO Operator Manual CLEANING AND DISINFECTING PROCEDURE REQUIRED MATERIALS Part# Description Notes/Usage EQASSYDISJUG Disinfect Jug - Used to mix water and chemicals for the cleaning and disinfecting procedure. SUMCOO00572 BW-1000, Acid - Replaces MinnClean AC Cleaner, low pH - Application: For removing mineral scale in membrane applications.
Page 46
HOSE NOTE: When Product Hose and Drain Hose are mentioned these are from the PB2 unless it is equipped with a Divert-to-Drain and in that case these hoses will be from the Divert-to-Drain.
Page 47
- If the PB2 continues to go into Low Pressure Alarm, see “Low Pressure Alarm” Section. 19. Allow the product water to run to drain for 30-45 minutes to rinse solution from the PB2. Better Water LLC; rev. Jul 2019...
Page 48
OPERATE Switch to the position to restart the PB2. - If the PB2 goes into Low Pressure Alarm, press the LOW PRESSURE ALARM RESET Button. If it continues to go into Low Pressure Alarm, see “Low Pressure Alarm” Section. 23. Test product water to determine if any residual solution still present.
The PB2 booster pump comes with a twist-lock power connection that plugs into the back of the PB2. The adaptor allows the pump to be plugged directly into the wall. If a PB2 City Booster Pump is not available, both it and the adaptor must be ordered from Better Water.
Page 50
Water-In Port. 6. If NOT equipped with a Divert-to-Drain… - Connect the opposite end of the Product Hose from the PB2 to the Disinfect Jug’s Product Port. - Connect the Drain Hose from the PB2 to the Disinfect Jug’s RO Drain Port.
Page 51
18. Disconnect the PB2 Booster Pump… … Disconnect the Disinfect Hose from the PB2 Booster Pump’s suction side (“IN”), and from the Disinfect Jug’s End-To-End Port … Disconnect the Hose from the PB2 Booster Pump’s Discharge side (“OUT”), and from the PB2’s Water-In Port...
Page 52
Divert-to-Drain’s solenoid valve will NOT divert to drain in a poor water quality condition. 29. If the PB2 is normally equipped with a Booster Pump, then remove the PB2 Booster Pump Adaptor from its power cord, reconnect its IN and OUT hoses and plug back into the back of the PB2.
If the PB2 will not be used for a temporary period of greater than 7 days, it should be Preserved and Packed for storage, which is good for up to 6 months. It should be noted that the BWI-3000 Membrane Preservative is NOT a disinfecting agent.
Page 54
… Remove the 10” Carbon Block Filter Housing, discard the filter, and allow to air day 18. Tag the PB2 to indicate that it is packed in preservative including the date it was packed. 19. The PB2 should be stored in a climate controlled environment, recommended to be between 60°-80°F.
- A minimum of 90% rejection should be maintained. - If the PB2 does not maintain greater than 90% rejection, check the source water pH. If the pH is above 8.0 this may impact the rejection of the PB2 and a DI polishing cartridge may be required.
Page 56
PB2 – Mediport Portable RO Operator Manual Better Water LLC; rev. Jul 2019 Page 52 of 80...
Buyer shall be responsible for all travel expenses and freight charges at all times. f. Better Water LLC shall have no obligation to repair or replace any defective part if buyer fails to follow the procedure set forth in “HOW TO OBTAIN A REPLACEMENT PART UNDER LIMITED WARRANTY”.
Page 58
Damage to or caused by filters/membranes or other replacement parts not purchased from Better Water LLC or damage caused by modification, alteration, repair or service of the equipment or any of its parts by anyone other than Better Water LLC or its expressly authorized representatives.
*** Should you wish Better Water LLC to make these wire relocation modifications, verification and testing, please call Better Water LLC to receive a RGA, then issue a Purchase Order for Labor / Testing charge of $125.00 (as of 1/11/12).
Page 61
Part # EQPB2RETROFIT 1) Unplug portable RO from outlet 2) On PB1 models, remove 4 screws in top cover of control box to access relay block. On PB2 models, remove 2 bolts in top corners of cabinet to access relay block.
Page 62
PB2 – Mediport Portable RO Operator Manual TECHNICAL SERVICE BULLETIN PB1/PB2 Low PSI / Monitor Modification TSB2012001 01/11/12 Page 3 of 6 TSB# Date Note: There are 2 brown wires on 21NC currently. To locate the correct brown wire, a volt meter must be used to determine continuity between #5 on the twist lock connector and the target brown wire on 21NC.
Page 63
Part # EQPB2RETROFIT 1) Unplug portable RO from outlet. 2) On PB1 models, remove 4 screws in top cover of control box to access relay block. On PB2 models, remove bolts in top corners of cabinet to access relay block.
Page 64
PB2 – Mediport Portable RO Operator Manual TECHNICAL SERVICE BULLETIN PB1/PB2 Low PSI / Monitor Modification TSB2012001 01/11/12 Page 5 of 6 TSB# Date Note: There are 2 brown wires on #7 currently. To locate the correct brown wire, a volt meter must...
Page 65
PB2 – Mediport Portable RO Operator Manual TECHNICAL SERVICE BULLETIN PB1/PB2 Low PSI / Monitor Modification TSB2012001 01/11/12 Page 6 of 6 TSB# Date Verifying Modifications: After the modification is complete, BEFORE the cabinet is closed and put into service, you must...
1. Insufficient flow. The PB2 has a minimum incoming flow requirement of approximately 2.2 gpm and a minimum filter-in pressure of 20 psi during flush. The PB2 is set at 36 gph (0.60 gpm) or 2270 ml/min and 110 to 125 psi pump pressure during test. When the PB2 is put into full flush, the water volume required to flush the membrane should be approximately 2.2 gpm, product and reject flow combined.
Page 67
1/2" HOSE CLAMP qty=2 5. If backflow preventers are used at the tap source it is recommended that a WATTS MODEL 9D DOUBLE-CHECK (BACKFLOW PREVENTER) be used with the PB2. - For specifications go to http://www.watts.com/pages/products_details.asp?pid=886. Better Water LLC; rev. Jul 2019...
TDS changes and fluctuating pH. The PB2 is set at the factory to alarm on %-Rejection. In this setting the water quality is more prone to a nuisance alarm due to the above mentioned changes in feed water quality. The PB2 can alternately be set to allow the alarm set-point to be set on Product TDS instead of %-Rejection.
Page 69
5. Change the set-point by pushing the CAL-UP or CAL-DOWN buttons until the desired set-point is displayed on the Water Quality Monitor display. 6. On the Water Quality Monitor on the front of the PB2, push the MODE SWITCH button until CAL is displayed.
Page 70
PB2 – Mediport Portable RO Operator Manual TECHNICAL SERVICE BULLETIN PB2 Water Quality Monitor Adjustment to Alarm on %- Rejection for Nuisance Alarms TSB2012003 04/27/12 Page 3 of 3 TSB# Date 8. Replace the cover on the internal Control Box, and close the outer External Top Panel cover.
Addition (Reaffirmed January 2012) Standards. The PB and PB2 RO units are a class 2 or type B Device. All PB2’s are tested to these standards at the factory before ship out. Beginning January 2012 the complete check off data sheet for each individual PB2 in the information packet included with the PB2.
Date OVERVIEW: If additional, external Carbon Blocks and/or Carbon Filters are used Better Water's PB1 and PB2 portable ROs, the internal, 10” Carbon Block Filter can be eliminated. A False Cartridge can be used in-place of the filter, inside the standard Filter Bowl.
Page 73
PB2 – Mediport Portable RO Operator Manual TECHNICAL SERVICE BULLETIN Replacing the 10” Carbon Block Filter Bowl with Replacement Short Bowl on Portable ROs TSB2012005 07/17/12 Page 2 of 2 TSB# Date REPLACEMENT INSTRUCTIONS: 1. Turn the power to the RO unit OFF.
USE of the NEPHROS SSU CAPSULE FILTER: When using the Nephros SSU capsule filter, it should be installed on the product hose between the PB2 and the dialysis machine. A sample port should also be installed after this filter to draw water to mix bicarb and...
… product water from the PB2 is diverted to drain … the Good Water Quality LED indicator is not lit * The Divert-to-Drain is set to divert based on the PB2’s Water Quality Monitor set-point setting * If the Water Qualify Monitor fails, the Divert-to-Drain will...
Page 76
Divert-to-Drain onto the bracket. WARNING To avoid electrical shock, turn the power to the PB2 OFF, and unplug it from the electrical outlet. A lock-out tag should be placed on the unit to prevent accidental use while this conversion is underway.
Page 77
REMOVE OLD CONTROL BOX top of the control box removed 7. Open the top cover of the PB2 to reveal the control box. 8. Remove the two screws that hold the top of the control box on, and remove that top.
Page 78
Green Tube from the DTD PRODUCT OUT Port to the DIALYSIS MACHINE 22. Remove the lock-out tag from the PB2 and reconnect it to the electrical Top view of Divert-to-Drain Outlet. showing tube connections * PB2 with Divert-to-Drain is ready to use.
Page 79
PB2 – Mediport Portable RO Operator Manual TECHNICAL SERVICE BULLETIN PB2 Conversion for New Control Box and Divert-to-Drain TSB2012011 12/26/12 Page 5 of 5 TSB# Date 5. Verify that once the Water Quality Monitor gets above the set-point and there is a good water quality condition that…...
Page 80
PB2 – Mediport Portable RO Operator Manual Better Water LLC; rev. Jul 2019 Page 76 of 80...