For Research Use Only For safety, this manual must be read carefully and the device be handled properly. FastGene ® Protein Electrophoresis Chamber Instruction Manual...
Page 2
Alteration of this device will create a potential safety hazard. Nippon Genetics Europe GmbH is not responsible for any injury or damage caused by use of this device for purposes other than those for which it is intended or by modifications of the device.
Page 3
Features o Unique wedge system to firmly and easily fix the Gel Cassette(s) o Buffer filled in the Chambers to cool all the surface of the Gel Cassettes(s) o Smaller quantity of buffer required o Easier sample loading visible through the transparent body Components o 1x Lid with power cables (1) o 1x Anode Buffer Chamber (2)
Page 4
Unpacking To remove the Lid, hold the device firmly and slide the Lid as indicated on the top. Pull out the blue Wedges to release the Inner Chamber (Cathode Buffer Chamber Frame). Take Out the Inner Chamber. Assembly and Operation Caution: Electrophoresis operates at a high voltage and Gel Cassetes are fragile by nature.
Page 5
Slide this sandwich into the Outer Chamber, matching respective colors of the plug (red and black) to those as indicated on the Outer Chamber. Once reaching the bottom of the chamber a slight resistance can be felt. Stabilize the Inner Chamber by inserting the Wedges straight down into the corner between the Gel Cartridges and the Outer Chamber.
Page 6
12. Turn the power supply on to start electrophoresis. Recommended conditions (for your reference) SDS-PAGE: 40mA DC constant for 1 hour a Gel Native-Page: 20mA DC constant for 2 hours a Gel DNA-PAGE: 20mA DC constant for 2 hours a Gel 13.