Summary of Contents for MPB PIRS
Page 1
Tel +39 0641200744 Via Giacomo Peroni 400/402 Fax +39 0641200653 00131 Rome - Italy info@gruppompb.com www.gruppompb.uk.com Quick Start Manual PIRS Quick start Manual Version: 1.1 Updated to the Software Version: 1.0.0 Updated to Firmware Version: 1.01 Quick start Manual PIRS...
Page 2
To ensure the correct use of the product, these general instructions must be read and applied before and for any use of the instrumentation. The PIRS is made for industrial environments and should be used by authorized staff only. MPB disclaims any responsibility for a use of the device different from explained in the manual.
Page 3
D e c l a r a t i o n o f c o n f o r m i t y (according to EMC 89/336/EEC directive and low voltage 73/23/EEC) This is to certify that the product: PIRS complies with the following European standards:...
CW signals with calibration feedback. 1.2. System description The PIRS can be used for Pre compliance Radiated Immunity Test (in accordance to the IEC 61 000-4-3) and to generate the electric fields with intensity among 1 V/m and 30 V/m inside shielding rooms, TEM and GTEM Cells, anechoic and semi anechoic chambers.
PIRS Options: Tripod support NMR-01 Tower up to 4m NMR-03 Log periodic antenna LP-02 Biconical antenna BC-01 Biconical log periodic antenna BL-01 PIRS converter RF Power meter Broadband meter 1.4. Technical specifications 10 kHz … 3 GHz Frequency Range Resolution...
The field generated will be a function of the distance of the object to be radiated, the type of antenna and the environment. By the PIRS the operator creates a quite field area in which to immerse the instrument he wants to test.
3. Software 3.1. Installation Launch the setup file (setup.exe) and eventually click YES/Execute in the pop up window. Please note that the PIRS SW can require to be “executed as administrator”. Follow the steps as in the following figures. Figure 3...
Figure 6 3.2. Connecting the instrument 1) Start the PIRS Software, connect the USB cable to the computer and to the instrument: at the first connection, wait for the driver installation, at the end of which a notification of the operating system will appear...
6. Set the duration of the signal generation 7. Click “Add row” to insert a new frequency list 8. Click “Send to PIRS” to save the list in the instrument 3.4. Edit list To modify a frequency list, double click on the list in the frequency list box and change the parameters in the right panel called “List number ##”...
QUICK START MANUAL PIRS 3.7. Add frequency To add a frequency click on the add row button and modify the value you want to insert. 3.8. Delete all frequency from list To delete all frequency list, click on the clear all button.
The calibration procedure is essential before making the measurements. To carry out the generation of the chosen field target it is first necessary for PIRS to associate all the frequencies with the necessary dBm of power that it must be supply according to the chosen setting.
Fiber optic (with or without converter according to the type of meter connected) tripods meter sample probe. Place the PIRS whit the antenna at the desired distance you need to follow the step below: Turn on the PIRS Connect the optical fiber Turn on the meter...
Page 15
For the generation of a known field to a stored frequency list it is necessary to maintain exactly the same positioning of the various elements as in the previous step of calibration (see chapter 4). Turn on the PIRS Press button GEN (F2) Press button LIST (F4)