Orion AQUAfast
1. Turn colorimeter on with the
mode
2. Fill clean dry cuvette to mark with sample and cap.
3. Place cuvette in colorimeter and press
zero•test
4. Remove cuvette, remove cap, and add tablet or
reagent. Crush and stir with stir rod. Recap
the cuvette.
5. Allow sample to react for any required wait time.*
6. If additional tablet or reagent is required, add,
crush, and stir with stir rod. Allow sample to react
for any required wait time.
7. Wipe any moisture or fingerprints off of cuvette
and place in colorimeter.
8. Press
zero•test
9. Record reading from display.
See individual instructions for detailed sample and
standard preparation.
* NOTE: Some two reagent samples require a
reading be taken at this point, then proceed to
step 7. Perform reading then add additional
reagent in step 6.
Analyze • Detect • Measure • Control™
key to choose method for multimeters.
key.
key.
®
II
Quick Start Card
key. Press
power

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Orion AQUAfast II and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Thermo Electron Orion AQUAfast II

  • Page 1 Orion AQUAfast ® Quick Start Card 1. Turn colorimeter on with the key. Press power key to choose method for multimeters. mode 2. Fill clean dry cuvette to mark with sample and cap. 3. Place cuvette in colorimeter and press key.
  • Page 2 Orion AQUAfast ® Calibration Colorimeters come pre-calibrated, however a user can check and adjust calibration if desired. Always use standard in recommended range: 1. Hold the key while turning on the mode colorimeter with the key. power 2. Fill clean dry cuvette to mark with standard and cap.

Table of Contents