The Series BCHP Calibration Test Pump is able to generate pressure
and vacuum for adjusting or calibrating pressure gauges, transmitters, or
switches. The pump is hand operated and has a pneumatic pressure
range of -28 in Hg to 870 psi (-0.95 bar to 60 bar). Dual pressure ports
allow two instruments to be connected to the pump without additional
fittings. The Series BCHP can be used in laboratories and production
areas. The fine adjustment valve on this pump ensures precise
measurements.
PRODUCT OVERVIEW
Figure 1
1. Pressure connector
for reference instrument
1/2˝ BSP female rotating
2. Fine adjustment knob
3. Pressure relief knob
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
Series BCHP Low Pressure Hand Pump
Specifications - Installation and Operating Instructions
4. Change-over switch for
pressure/vacuum generation
5. Pressure connector for test
instrument 1/4˝ BSP female
rotating
6. Pump handle
TestEquipmentDepot.com
4-27/64
1/2˝ FEMALE BSP
[112.40]
W/ O-RING
REFERENCE PORT
11/32
[8.67]
3-3/16
[81.02]
3-1/64
[76.57]
1-13/32
[35.83]
61/64
29/64
[24.26]
[11.43]
SPECIFICATIONS:
Media: Air.
Output Ranges: -28 in Hg to 870 psi (-0.95 to 60 bar).
Process Connections: 1/4˝ female BSP
Gauge Connection: 1/2˝ female BSP.
Materials: Anodized aluminum, brass, and ABS.
Weight: 8.4 lb (3.8 kg).
MOUNTING INSTRUCTIONS
The reference instrument is to be mounted on the top connection of the
pump. Fastening with the knurled nut by hand is sufficient. An integral O-
ring seals the reference instrument pressure port without the use of
teflon tape.
The test instrument is attached to the connection at the end of the hose.
The optional NPT adapter set includes sealing gaskets and different
adapters to best fit the instrument being tested. The tube can also be
removed to attach the test instrument directly to the pump using the
same adapter.
NOTICE
Do not use tape sealant to seal the ports, as it may damage the
pump.
Bulletin TE-BCHP
2-3/64
1-25/32
[51.85]
[45.16]
1/4˝ FEMALE BSP W/
O-RING TEST GAGE
CONNECTION
APPROX. 18˝
3-3/16
11/32
[81.02]
[8.91]
9/64
[3.61]
29/64
7/8
[11.54]
[22.07]
43/64
[16.88]

Advertisement

loading
Need help?

Need help?

Do you have a question about the BCHP Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Dwyer Instruments BCHP Series

  • Page 1 Bulletin TE-BCHP Series BCHP Low Pressure Hand Pump Specifications - Installation and Operating Instructions 2-3/64 1-25/32 [51.85] [45.16] 4-27/64 1/2˝ FEMALE BSP [112.40] W/ O-RING 1/4˝ FEMALE BSP W/ REFERENCE PORT O-RING TEST GAGE CONNECTION 43/64 11/32 [16.88] APPROX. 18˝ [8.67] 3-3/16 [81.02] 3-3/16...
  • Page 2 Contact customer service to receive a Return Goods Authorization number before shipping the product back for repair. Be sure to include a brief description of the problem plus any additional application notes. ©Copyright 2014 Dwyer Instruments, Inc. Printed in U.S.A. 7/14 FR# R6-444183-00...

This manual is also suitable for:

Bchp-1