V. MAINTENANCE OF THE PRONEB
Tubing Maintenance
If there is any moisture or condensation in the tubing after the treatment, let the compressor run
with only the tubing attached for 10 minutes. The warm air from the compressor will dry the
moisture or condensation in the tubing. Alternatively, you can dry the tubing by removing it
from the compressor and hanging it up with the "ends down" (Fig. D) to allow any moisture or
condensation to drain out and evaporate. Use a damp cloth to wipe the exterior of the tubing.
Cleaning the PRONEB Housing
Use a damp cloth to wipe the exterior of the PRONEB housing and tubing (Fig. E). Never spray
fluid into the ventilation slots on the PRONEB Compressor. Fluid could cause damage to the
electrical and compressor parts and could lead to a malfunction. In the event that fluid enters
the unit, contact a PARI authorized service center.
Filter Replacement
Under normal conditions, the filter on the front panel of the PRONEB should be replaced after
500 hours of use, or sooner if it appears soiled (gray or brown), clogged or wet. Check the
condition of the filter after every 10-12 treatments. Do not attempt to wash and re-use filters.
Do not use substitute filter material. Compressor damage can result. Turn the round plastic
filter cover (Fig. F) counterclockwise (left) and remove filter (use a coin to help turn the cover
if tight) (Fig. G). Replace with part #41F64.
Maintenance and Repair
Never operate the PRONEB if it is damaged or improperly or incompletely assembled. Use
only original PARI parts and attachments authorized by PARI for the PRONEB. Use of other
parts and attachments may damage the unit or cause improper treatment. Do not open or
disassemble the compressor unit at any time. Repairs, service and maintenance should be
carried out only by authorized PARI service centers.
Note: There are no user serviceable parts inside the PRONEB compressor case. Service by
anyone other than authorized PARI Service Center voids warranty.
VI. WARRANTY OR REPAIR RETURN PROCEDURE
1. Contact PARI Patient Connection (Service Department) using the phone numbers listed
in Section VII (Proneb Limited Warranty) of this booklet.
2. In writing, describe the malfunction or symptoms of the PRONEB. Include your full name,
address and daytime telephone number.
3. Send written description and PRONEB in a protected box to the address provided by
Patient Connection.