Download Print this page

No Touch Thermometers THD2FE Manual

Non-contact forehead thermometer

Advertisement

Quick Links

THD2FE Non-contact Forehead Thermometer
Intended use:
The infrared forehead thermometer can measure the infrared heat generated by
central forehead and surrounding tissues to reflect patient's body temperature
accurately.
Intended operator: At least 11 years old (5 years intensive reading experience), no maximum.
Specifications
 Temperature measurement range:
Forehead mode: 34~42.2°C (93.2~108°F)
Surface mode: -22~80°C (-7.6~176°F)
 Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
 Storage temperature range: It should be stored at room temperature between -20~+50°C ,
RH≦85%
Transportation temperature shall be less than 70°C, RH≦95%
 Atmospheric pressure: 800~1013 hPa
 Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC),
IEC/EN60601-1(Safety) standards, ISO10993, RoHS.
 Accuracy:
Forehead mode: ±0.2
Surface mode: ±0.3°C (0.5°F) within 22~42.2°C (71.6~108°F),
 Fever indication, memory and °C/°F switch function
 Battery: AAA x 2 pcs
 Battery life: around 3,000 continuous readings.
 Expected Service Life: 4 years
 This thermometer converts the forehead temperature to display its "oral equivalent."
(according to the result of the clinical evaluation to get the offset value)
 E: Blue LED Backlight
Blue LED backlight will be automatically turned on after measurement, and automatically
turned off after 5 seconds.
 Enclosure Rating: IP22
 Dimensions: 158.0 x 48.0 x 40.2 mm
Weight: 100 grams including battery
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using the infrared thermometer.
This is not an AP or APG product.
Functions
The thermometer has been designed for home use. It's not meant to replace a visit to the
doctor. Please also remember to compare the measurement result to your regular body
Forehead temperature
temperature. Please consult with doctor if you have health concerns.
 Please see the "Use of the thermometer" section to learn how to measure the body
temperature.
The surface mode shows the actual and unadjusted surface temperature which is different from
the body temperature. It can help you monitor if the object temperature is suitable for the baby
Surface temperature
or patient, for example the baby's milk.
 Please see the "Use of the thermometer" section to learn how to measure the object
temperature.
If the thermometer detects a temperature ≧37.5°C (or 99.5°F) under forehead mode, three
Fever indication
short beep sound will follow one long beep sound to warn the user for potential fever.
There are total 25 sets of measurement condition for body temperature.
Memory locations
 When power on, press the "ON/MEM" button to see the temperature records with
In "Power Off" mode, press and hold the "START" button, then press the "ON/MEM" button for
3 seconds, icon "°C" will be switched to icon "°F". You can also use the same process to
°C / °F switch
change the LCD display from °F to °C.
NOTE: Memory clear and Celsius/Fahrenheit switch are together. When you switch Celsius and
The device setting with buzzer is on, you can set buzzer on/off under Mute mode.
When power on, press and hold the "ON/MEM" button for 3 seconds. The icon
the LCD screen and then release the "ON/MEM" button to set MUTE. Thus you will not hear
Mute mode
beep sounds. You can also use the same process to turn off the Mute function.
NOTE: If keep pressing "ON/MEN" button for 5 seconds after
Use of the thermometer
Note: If there is any temperature difference between the places where the device is stored and where you are going to measure, subject and the device should
stay in the same room for at least 15 minutes before measurement.
1. Always make sure the probe lens is clean without any damage and the forehead is clean.
Warning: Choking from swallowing small parts and batteries by children or pets is possible, please keep small parts and batteries at places where
children and pets can't reach.
2. Power on:
Press the "ON/MEM" button (see figure 1).
o
o
C (0.4
F) within 35~42°C (95~107.6°F),
o
C (0.5
o
F) for other range.
±0.3
others ±4% or ±2°C (4°F) whichever is greater.
Fahrenheit, the Memory will be cleared.
be power off WITHOUT setting Mute.
icon flashing, the device will
Figure 1
icon.
will flash on

Advertisement

loading

Summary of Contents for No Touch Thermometers THD2FE

  • Page 1 THD2FE Non-contact Forehead Thermometer Intended use: The infrared forehead thermometer can measure the infrared heat generated by central forehead and surrounding tissues to reflect patient’s body temperature accurately. Intended operator: At least 11 years old (5 years intensive reading experience), no maximum.
  • Page 2 3. Measuring body temperature on the forehead: Press the “ON/MEM” button to power on the device. Forehead mode is the default mode. You can see the icon on the screen and hear two beep sounds (see figure 1). In this mode, you can hold the thermometer within 4 cm from the central forehead (Fig. 2) and press the "START"...
  • Page 3 Symbol Descriptions Indicates this device is subject to the Waste The CE mark and Notified Body Electrical and Electronic Authorized Registration Numbers, the Equipment Directive in the European Union. representative in the requirement of Annex II from Medical To protect the environment, dispose of European community Device Directive 93/42/EEC are met.
  • Page 4 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.