Contents Note: This manual describes the features, functions and operation of the Manual XY Stage. Before use, carefully read this manual, directions, all precautionary information and specifications. 1.0 Product Introduction ..............1 1.1 Handling Scientifica equipment – precautions ......1 1.2 The Manual XY Stage ..............
The Manual XY Stage is a translation stage for moving microscopes in X and Y. The XY Stage provides 50 mm of travel in X and Y with 20 nm resolution. Due to its large load carrying capacity the microscope stage can be used with almost any Zeiss, Leica, Nikon or Olympus inverted or upright microscopes;...
1 x Stage Protective Membrane* 1 x Hex Key Set 1 x Screw Kit Please note: When purchased for use with other manufacturer’s microscopes the XY Stage will be shipped with the protective membrane already in place. 2.0 Packing List...
All stages will be shipped with red brackets to prevent component damage in transit. Before proceeding, remove the red shipping brackets from the sides of the XY Stage using the supplied hex key set. Important: Retain these brackets for future transit or return of the equipment.
3.2 Attach micrometers Loosen grub screw on the Y Axis and insert a micrometer (Fig A). For left- handed users, remove the blanking cap on the left-hand side and follow the same process. Figure A: Y Axis micrometer Loosen screws in pinch block on the X Axis and insert a micrometer (Fig B).
The XY Stage is fixed to an anti-vibration table using the mounting holes in its lower section, marked ‘A’ and ‘B’ in Figure C. The XY Stage is attached to the anti-vibration table using either ¼ x 20 or M6 x 12 screws, depending on the thread of the table.
Scientifica for help. 3.4.1 Conventional mounting of microscopes to the XY Stage Scientifica’s SliceScope can be secured to the XY Stage using the six mounting holes indicated in Figure D: Demonstrating holes for mounting a SliceScope to the XY Stage.
As an example, in the case of a Zeiss FS2 microscope, the following steps must be carried out: 1. Remove the top plate from the stage of the XY Stage by unscrewing the 6 x M6 screws securing it in place.
Do not use any type of abrasive pad or material to clean the unit. 4. 1 Troubleshooting 1.Movement not smooth • Remove micrometres from the XY Stage and check to see if the screw housing has been damaged by over tightening. • Check to see if the micrometre rotates smoothly outside the XY Stage.
15 to +40ºC Storage 0 to +60ºC Bearings Crossed roller Dimensional diagram: Figure A: XY stage top plate dimensions* *Dimensions and mounting holes will vary depending on the microscope being mounted, as each top plate is made specific to the microscope.
During the warranty period, Scientifica Ltd will repair or, at its option, replace parts which prove defective when the instrument is returned to Scientifica Ltd in the UK or to your local representative. The warranty will not apply if the instrument has been damaged by accident, misuse, or as a result of modification by persons other than Scientifica Ltd personnel.
7.0 About Scientifica Scientifica specialises in the manufacture and distribution of innovative, superior equipment for electrophysiology researchers, based on principles of careful design and solid engineering. By responding to the unique demands placed on the equipment of electrophysiologists and thinking laterally around these challenges,...