Digital Image Receptor
The PaxScan 2530W is a radiographic
digital x-ray imaging sub-system
82510-000
Rev A
11/2014
© Varian Medical Systems | X-Ray Products

Advertisement

Table of Contents
loading

Summary of Contents for Varian PaxScan 2530W

  • Page 1 Digital Image Receptor The PaxScan 2530W is a radiographic digital x-ray imaging sub-system 82510-000 Rev A 11/2014 © Varian Medical Systems | X-Ray Products...
  • Page 2 + 1 801 973 5023 Fax Notice Information in this manual is subject to change without notice. Varian is not liable for errors or omissions contained in this manual or for any damages incurred in connection with furnishing or use of this material.
  • Page 3: How To Reach Us

    Chapter 11 of this manual and email to: PAXSAN.RMA@varian.com before contacting your Varian representative. To speak with our technical support personnel call:  Call (800) 432-4422 dial 8. For Warranty and Returns please refer to:  http://www.varian.com/us/xray/services_support.html © Varian Medical Systems | X-Ray Products...
  • Page 4 Sales, Technical Service and Warehouse  Germany Karl-Arnold-StraBe 12 + 49 2154-92 4980 Phone + 49 2154-92 4980 Phone D-47877 Wilich, Germany + 49 2154-92 4994 Fax Sales, Technical Service and Warehouse E-mail: sales-xray@varian.com © Varian Medical Systems | X-Ray Products...
  • Page 5: General Safety Information

    To reduce the potential of electrical shock, the operator should not Warning: simultaneously touch the patient, cable connections, ex-sync serial connector, fuse holders, and the power supply. Equipment must be connected to an uninterruptible power source for Warning: which it was designed. © Varian Medical Systems | X-Ray Products...
  • Page 6 OEM system level to control panel surface temperature so that it does not exceed 41 Note: There are no contraindication situations. © Varian Medical Systems | X-Ray Products...
  • Page 7: Explanation Of Symbols

    Authorized Representative in the Consult Instruction European Manufacturer for Use Community/European Union Moisture Resistance IPX1 Heated Surface Type B Applied Part Level Tested Load Weight Temperature Limits Non-ionizing radiation Restriction © Varian Medical Systems | X-Ray Products...
  • Page 8: Chapter Summary

    Chapter Summary Introduction Getting Started Additional Features Paxscan Software Modes of Operation Calibration Procedures Image Acquisition Safety Maintenance Trouble Shooting Problem Report Appendix © Varian Medical Systems | X-Ray Products...
  • Page 9: Table Of Contents

    Connecting the Cables – I/O Box Setup Only ....................17 Table 1-0 Cable Connection Details ......................17 Figure 4-0 Cable Connection Varian I/O Box Mode - Overview ............. 18 Figure 5-0 Cable Connection I/O Box Mode - Detail ................19 Power on Sequence –...
  • Page 10 Table 4-0 Guidance and Mfgr Declaration - Electromagnetic Emissions ..........48 Table 5--0 Guidance and Mfgr Declaration - Electromagnetic Immunity ..........49 Table 6-0 Recommended Separation Distance Between ................51 Portable and Mobile RF Communications and the PaxScan 2530W ............51 Electrical Protection ............................52 Environment Limits ............................52 Temperature &...
  • Page 11 Cleaning and Disinfection ..........................62 Repairs................................62 Proper Disposal ..............................63 Troubleshooting ..............................64 PaxScan 2530W Problem Report Customer Information .................. 65 Appendix A ............................... 66 Interfacing Information ..........................66 Figure A.1 Diagram of the 2530W Imager System ................66 Figure A.2 Schematic for “Expose Ok”...
  • Page 12: Introduction

    Chapter 1 Introduction The PaxScan 2530W model family of radiographic digital image receptors is commonly referred to as a flat panel detector (FPD). The detector together with image processing and command software called Virtual Command Processor (VCP) is designed for integration into a complete X-ray system. The imaging system has two main system components: The flat panel sensor and VCP software.
  • Page 13: Optional Parts

    (801) 972-5000. Intend Use The PaxScan 2530W model family is designed to meet the needs of general radiography diagnostic for medical and veterinary use in portable applications utilizing multiple sensitive and extend dynamic range modes. This family model will acquire image over a wide range of dosage, while providing maximum access to the patient, with a minimum possible border on the active imaging area.
  • Page 14 Figure 1-0 Patient Contact Surfaces – 2530W © Varian Medical Systems | X-Ray Products...
  • Page 15 Figure 2-0 Patient Contact Surfaces – 2530W © Varian Medical Systems | X-Ray Products...
  • Page 16: Getting Started

    331.5mm x 279.8mm x 16.5mm. Figure 3-0 Imager Configuration (Receptor 1 – Wireless, Receptor 2 – Wired) Receptor Optional Config File I/O Box Image Data Workstation *Prep Receptor *Exp_Req *Exp_OK © Varian Medical Systems | X-Ray Products...
  • Page 17 Win32 DLL. The control of the Receptor is platform- independent. The I/O Box (Optional) provides all hardware interfaces for the PaxScan 2530W using an external hardware interface connection. The laptop style power supply, which is optional equipment, provides the I/O Box with +24V DC power.
  • Page 18: Connecting The Cables - I/O Box Setup Only

    Action / Description If the imager is integrated with Varian I/O Box there are four (4) cable connections for the 2530W Flat Panel Receptor: (a) The laptop style power supply cable , (b) external sync cable, (c) Category 5 or better Ethernet cable, and (d) tether cable.
  • Page 19 Figure 4-0 Cable Connection Varian I/O Box Mode - Overview Use minimum 18 AWG wire for connection to the external Important: PE terminal on the I/O Box © Varian Medical Systems | X-Ray Products...
  • Page 20 Figure 5-0 Cable Connection I/O Box Mode - Detail © Varian Medical Systems | X-Ray Products...
  • Page 21: Power On Sequence - Without I/O Box

    The receptor will keep trying to connect until a successful association takes place. The Orange LED, when lit, indicates a Ready-to-Expose State. See Figure 6.1 for details. For wireless mode, Pleora eBUS Universal Pro driver must be Note: used © Varian Medical Systems | X-Ray Products...
  • Page 22 Figure 6-0 2530W Receptor Power/LED Figure 6-1 2530W Power/LED - Details © Varian Medical Systems | X-Ray Products...
  • Page 23 Wait at least 1 (one) minute after powering up the I/O box before any operations. Figure 7-0 I/O Box Power On Connection Figure 7-1 I/O Box LED – Details © Varian Medical Systems | X-Ray Products...
  • Page 24 PaxScan 2530W Moisture Resistance Level Tested, horizontal position, x-ray window face up, without Tether Cable attached and rubber connecter cover in place. Warning: IPX1 Receptor protected against falling water equivalent to 3-5mm rainfall per minute for duration of 10 minutes.
  • Page 25: Additional Features

    Chapter 3 Additional Features Auto Trigger PaxScan 2530W auto trigger feature allows the user to acquire images without the necessity of any hardware synchronization; the receptor automatically reacts to X-ray exposure and delivery an image to the workstation. Wireless ●...
  • Page 26: Paxscan System Software

     OK – will launch the PaxScan ViVA System Software Install Shield Wizard. Step Action / Results For a normal install, follow example shown in below screenshots. © Varian Medical Systems | X-Ray Products...
  • Page 27 Select the install eBUS Universal Driver and click install button, verifty that the eBUS Universal Pro Driver is installed and click Close It is critical to turn OFF any firewall on the workstation otherwise Note: the system will not connect. © Varian Medical Systems | X-Ray Products...
  • Page 28 Step Action / Results Set Network card IP address as follows Right click on Network Icon and Select Properties Select “Change adapter settings” Select “Properties” © Varian Medical Systems | X-Ray Products...
  • Page 29 Click on “Internet Protocol Version 4 (TCP/IPv4)” and click “Properties” The subnet mask will automatically be set to the one described below. Click “Ok”. If a network interface card is already using 192.168.2.0/255.255.255.0, contact your Varian representative for instructions. © Varian Medical Systems | X-Ray Products...
  • Page 30: Receptor Files Installation

    Follow through the Install Shield Wizard screens to complete the PaxScan Receptor installation. You must restart  your computer for installation to take effect. Step Action / Results  2530W – follow the prompt sequence Install Receptor installation Disk © Varian Medical Systems | X-Ray Products...
  • Page 31 Run Setup.exec and follow the instructions. When you get the prompt to verify where the receptor data is located, use the default shown below or specify the file location; then click install. Select “Finish” © Varian Medical Systems | X-Ray Products...
  • Page 32: Modes Of Operation

    Modes of Operation The PaxScan 2530W supports two modes of operation as defined in Table 2-0. In general, there is a tradeoff between varying operation modes of resolution, or field of view, or cycle time, or noise. The sensitivity of the imager is optimized to match the X-ray dose used in each mode.
  • Page 33: Default Mode

    The Virtual Command Processor software supports multiple connections to two or more receptors of the same type; however, ViVA controls one receptor at a time. The receptor selection is changed from the  Acquisition drop down menu. This feature is typically useful in a testing environment. © Varian Medical Systems | X-Ray Products...
  • Page 34 Select Open link Receptor under the Acquisition drop down menu bar to establish connection to the PaxScan imaging system. For multiple receptors, completion of the single receptor setup is required before additional receptors are available for setup in the serial # drop down. © Varian Medical Systems | X-Ray Products...
  • Page 35: Calibration Procedures

    Ensure required receptor appears in the Mode drop down. The 2530W currently supports Rad or Auto Trigger 1x1 4pf. Click Offset Calibration button or select from the menu bar under Acquisition. © Varian Medical Systems | X-Ray Products...
  • Page 36: Gain Calibration

    To reduce the effects of noise, the average of each pixel in the flat field image is calculated by accumulating a number of frames into an internal memory buffer, then dividing the sum of each pixel by the number of frames acquired. © Varian Medical Systems | X-Ray Products...
  • Page 37 However, the actual number of calibration frames used must be determined solely by the system integrator depending upon their specific performance requirements. Gain calibration requires the production of X-rays and therefore Important: certain precautions must be taken by the human operator. © Varian Medical Systems | X-Ray Products...
  • Page 38: Connect To The Panel As Access Point

    Detailed instructions on performing gain calibrations are covered in the ViVA Online help documentation. Table3-0 Gain Calibration - All Modes Step Action Results To ensure proper warm up, the PaxScan 2530W Receptor Warm Up must be operational for a least two (2) hours prior to Gain calibration. Offset Calibration Software performs a new Offset calibration referred to as dark field acquisition.
  • Page 39 If the WiFi connection is lost on the workstation the green LED will stay illuminated. Before an exposure is allowed, it is the Warning: responsibility of the system integrator to verify the WiFI connection via Windows and also battery level vial Virtual CP (VCP). © Varian Medical Systems | X-Ray Products...
  • Page 40 If Viva is launched and only one receptor is installed, Viva will automatically open the link to the receptor. If multiple receptor files are on the workstation the following steps are needed to select the proper receptor by serial number. © Varian Medical Systems | X-Ray Products...
  • Page 41 7. Select Open link Receptor under the Acquisition drop down menu bar to establish connection to the PaxScan imaging system. For multiple receptors, completion of the single receptor setup is required before additional receptors are available for setup in the serial # drop down. © Varian Medical Systems | X-Ray Products...
  • Page 42: Radiographic Or Auto Trigger Mode Gain Calibration

    Step Action / Results Ensure the desired receptor and imaging mode appears in the Mode drop down. Click Gain Calibration from the menu bar under Acquisition invokes hardware handshaking for dark field calibration. © Varian Medical Systems | X-Ray Products...
  • Page 43 The software will accumulate 2 frames right after the panel is exposed to an X-ray, when the current image process is completed the “Gain Calibration” window will display “X-ray exposure in progress… Frame count: 0: © Varian Medical Systems | X-Ray Products...
  • Page 44 Step Action / Results Once all X-ray frames have een acquired, click “Finish” to complete the calibration. The Gain Media should be 9000 +/- 1000 counts, click “Close” to complete gain calibration.: © Varian Medical Systems | X-Ray Products...
  • Page 45 Note: Replacement of the X-ray tube will require a new gain calibration to be performed. Varian recommends accumulating 32 frames for gain calibration for optimal image quality. However, the actual number of calibration Note: frames used must be determined solely by the system integrator depending upon their specific performance requirements.
  • Page 46: Image Acquisition

    Once Offset and Gain Calibration is performed, you are ready to acquire images. Radiography or Auto Trigger Mode The Radiography mode provides the technician with superior single-shot, higher resolution images, for  diagnosis. Step Action / Results Select required receptor from Mode drop down menu. © Varian Medical Systems | X-Ray Products...
  • Page 47 Select “System Settings” from the menu bar under Acquisition to turn ON corrections for a calibrated image. Only select “Offset Corrections”, “Gain Corrections”, and “Pixel Corrections” Select the Acquire Image button invokes imager to begin acquiring images. © Varian Medical Systems | X-Ray Products...
  • Page 48 When “Viva – Radiograph Acquisition Progress” windows appears Depress operator control fully to “Expose” position to complete image capture and retrieval. Acquired image can be saved in the desired file format by selecting File / Save As. © Varian Medical Systems | X-Ray Products...
  • Page 49: Safety

    Table 4-0 Guidance and Mfgr Declaration - Electromagnetic Emissions The PaxScan 2530W is intended for use in the electromagnetic environment specified below. The customer or the user of the PaxScan 2530W should assure that it is used in such an environment.
  • Page 50 Table 5--0 Guidance and Mfgr Declaration - Electromagnetic Immunity The PaxScan 2530W is intended for use in the electromagnetic environment specified below. The customer or the user of the PaxScan 2530W should assure that it is used in such an environment.
  • Page 51 To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in whih the PaxScan 2530W used exceeds the applicable RF compliance level above, the PaxScan 2530W should be observed to verify normal operation. If abnormal...
  • Page 52: Portable And Mobile Rf Communications And The Paxscan 2530W

    Table 6-0 Recommended Separation Distance Between Portable and Mobile RF Communications and the PaxScan 2530W The PaxScan 2530W is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the PaxScan...
  • Page 53: Electrical Protection

    100ºC, incinerate, or disassemble the battery. Charge only with the Caution: receptor or battery charger supplied. Please follow local governing ordinances and recycling plans regarding proper disposal or recycling of the lithium-ion rechargeable battery. © Varian Medical Systems | X-Ray Products...
  • Page 54: Lithium-Ion Battery Handling, Storage, & Shipping

    ● Always check all applicable local, national, and international regulations before transporting a Lithium-Ion battery. ● It is customers responsibility to ship battery according to local and international shipping regulation for Lithium-Ion battery in effect at the time of shipment. © Varian Medical Systems | X-Ray Products...
  • Page 55: Regulatory

    A Declaration of Conformity has been filed for this product and available upon request by contacting Varian Medical Systems - X-Ray Products. ● The Varian Battery Charger is a Class 1, continuous operation device and meets the following: IEC 61000-4-2 Electro-Static...
  • Page 56: Radio Frequency (Rf) Compliance Information

    Modifications not expressly approved by the manufacturer could void the user's authority to operate the equipment under FCC rules. In the 5150 to 5250 MHz frequency range this transmitter is restricted to indoor use only. © Varian Medical Systems | X-Ray Products...
  • Page 57: Taiwan Ncc Warning Statement

    在 5.25-5.35 秭 赫 頻 帶 內 操 作 之 無 線 資 訊 傳 輸 設 備 , 限 於 室 內 使 用 。 前項合法通信,指依電信規定作業之無線電信。低功率射頻電機須忍受合法通信 或工業、科學及醫療用電波輻射性電機設備之干擾 Japan Telecom Certification – JATE © Varian Medical Systems | X-Ray Products...
  • Page 58: Singapore Compliance

    Interference is likely to occur during operation of the equipment Korean Class A The devices are for business (A grade) EMC equipment distributors and users of this point as a caution and outside the home for the purpose of the use. © Varian Medical Systems | X-Ray Products...
  • Page 59: Industry Canada Notice

     EN 300 328  EN 301 893  EN 301 489-17  EN 60950 Marking by the symbol: ! indicates that usage restrictions apply in countries listed on this product’s packaging. © Varian Medical Systems | X-Ray Products...
  • Page 60 Europe - Declaration of Conformity in Languages of the European Community. Česky [Czech] Varian Medical Systems, Inc. tímto prohlašuje, že tento Radiolan je ve shodě se základními požadavky a dalšími příslušnými ustanoveními směrnice 1999/5/ES. Dansk [Danish] Undertegnede Varian Medical Systems, Inc. erklærer herved, at følgende udstyr Radiolan overholder de væsentlige krav og øvrige relevante krav i...
  • Page 61 [Portuguese] os requisitos essenciais e outras disposições da Directiva 1999/5/CE. Slovensko Varian Medical Systems, Inc. izjavlja, da je ta Radiolan v skladu z bistvenimi zahtevami in ostalimi relevantnimi določili direktive 1999/5/ES. [Slovenian] Varian Medical Systems, Inc. týmto vyhlasuje, že Radiolan spĺňa základné...
  • Page 62 If operation is occurring outside of the allowable frequencies for outdoor use, as listed above, the user must contact the applicable national spectrum regulator to request a license for outdoor operation. © Varian Medical Systems | X-Ray Products...
  • Page 63: Maintenance

    No user serviceable parts. If repairs are necessary, please see How To Reach Us. The least replaceable units (LRU) are:  Receptor Assembly  I/O Box  External Power Supply  Tether Cable  Varian Battery  Varian Battery Charger (if applicable) © Varian Medical Systems | X-Ray Products62...
  • Page 64: Proper Disposal

    Proper Disposal The 2530W receptor should be returned to Varian Medical Systems for disposal. We request that you obtain an RMA number using the same procedure for warranty/returns of products. Contact: PAXSCAN.RMA@VARIAN.COM Do not dispose of the lithium-ion rechargeable battery in the garbage. Please consult local governing ordinances and recycling plans regarding proper disposal.
  • Page 65: Troubleshooting

    X-rays until the residual image is gone. ViVA error message. 1. Please complete PaxScan 2530W Problem Report. 2. Email the error log file generated to: paxscan.rma@varian.com. This log file is normally found at C:\users\{username}\AppData\Local\ crashdumps\viva.log...
  • Page 66: Paxscan 2530W Problem Report Customer Information

    Chapter 11 PaxScan 2530W Problem Report Customer Information Date: Your Name Company/Unit Name: Email: Phone Number: Fax Number: Product Information. PaxScan Part Number: Imager Serial Number: Software Revision #: Operation I was trying to perform (be as specific as possible: What happened (use additional sheets as necessary): PAXSCAN.RMA@varian.com...
  • Page 67: Interfacing Information

    Appendix A Interfacing Information All the interfacing connections for the PaxScan 2530W are at the panel itself and the Varian I/O box. The Gigabit Ethernet connection carries control information to the panel and supplies image data with diagnostic information to the customer supplied workstation. The Hardware Synchronization connection is a 9-pin D-sub type.
  • Page 68 If the Varian I/O box is used the synchronization interface to the panel consists of two inputs and one output, all isolated through opto-couplers. The expected inputs to the panel are Prepare and Exposure_Request. The output from the panel is Expose_OK, which can be used to trigger the generator. This active low signal is used to identify when the panel is ready for exposure.
  • Page 69 The typical timing of the synchronization interface is shown below. The panel is maintained in an idle state, until the asynchronous Expose_Request is received. When the panel receives the Expose_Request, the assumption is that the generator is ready to make an exposure when Expose_OK is issued. Figure A.3 2530W Timing © Varian Medical Systems | X-Ray Products...
  • Page 70 Figure A.4 2530W Timing © Varian Medical Systems | X-Ray Products...
  • Page 71: Use Cases

    The following are potential application configurations: Single Panel 29'-0" X-Ray 1 Table Prep/Exp_Req Exp_OK Host Bridge Generator Chest Stand X-Ray 2 Patient Receptor 1 User 16'-6" 4'-9" Figure B.1 Example of Single Panel Use © Varian Medical Systems | X-Ray Products...
  • Page 72: One Panel Per Room

    Exp_OK Exp_OK Host Bridge Host Bridge Generator Generator Chest Stand Chest Stand X-Ray 2 X-Ray 2 User Receptor 1 Patient Patient Receptor 2 User Figure B.2 Example of One Panel Per Room Use © Varian Medical Systems | X-Ray Products...
  • Page 73: Multiple Panel

    Multiple Panel X-Ray 1 Receptor 2 Table Prep/Exp_Req Exp_OK Host Bridge Generator Chest Stand X-Ray 2 Patient Receptor 1 User Figure B.3 Example of Multiple Panel Use © Varian Medical Systems | X-Ray Products...
  • Page 74: Battery Installation/Removal For 2530W

    If the button is already depressed from a previous battery ejection, proceed to step 3. Press down on the lifted side of battery snapping it into place in the battery compartment. Receptor is now ready for user. © Varian Medical Systems | X-Ray Products...
  • Page 75 Doing this will cause one side of the battery to lift out of the battery compartment. Grab the lifted side of the battery and finish removing. © Varian Medical Systems | X-Ray Products...
  • Page 76: Battery Charger

    The Varian 1 bay and 3-bay charger is intended for use with the Varian Battery. The Varian charger will fully charge a Varian battery in maximum 3 hours when operating at room temperature. The charger is used with the PaxScan wireless radiographic digital image receptors.
  • Page 77 Charging a fully discharged battery will take 3.5 hours maximum and 2.5 hours typical when operating at room temperature. Up to three batteries can be charged at a time in the Varian 3- bay battery charger.
  • Page 78 2. The output of the charger is 16.8V DC, 1.4A max (+/1 1% and 10 Hz).. Step Action / Results The Varian battery communicates with the Varian 1-bay charger when it is plugged into the charger. The battery monitors its readiness state and is mechanically keyed to charger for easy installation.
  • Page 79 Do not remove the charger cover. The Charger has no internal user Warning: serviceable parts. Do not use in operating room or other oxygen rich environment. Do not use in conjunction with flammable agents. Warning: Do not use in an environment with condensing moisture. © Varian Medical Systems | X-Ray Products...
  • Page 80 Do not use flowing liquid or immersion on the receptor, battery, battery compartment, or battery charger. Caution: Do not sterilize. Do not attempt to insert objects other than the Varian battery into the Caution: charger bay. Use the Varian Battery Charger only with the Varian supplied power Important: supply and power cord.