2020-05-13 Revision: X Notice Butterfly Network, Inc. (BNI), shall not be liable for errors contained herein or for incremental or consequential dam- ages in connection with the furnishing, performance, or use of this material. Proprietary Information This document contains proprietary information, which is protected by copyright.
This chapter provides an introduction to the Butterfly iQ™ Personal Ultrasound System. Overview Butterfly iQ™ personal ultrasound is designed to be easy to use, portable, and battery powered. Its commercial off- the-shelf mobile platform (mobile device) provides a simple interface for the user.
Contraindications for Use Butterfly iQ™ should not be used for indications other than the ones approved by the applicable governing agency. Training In order to safely and effectively operate Butterfly iQ™, the user shall meet the following: •...
This user manual is intended to assist in the safe and effective operation of Butterfly iQ™. It is important that all users review and understand all instructions in this user manual before operating the device, paying careful attention to the warnings and cautions throughout the manual.
• Movement of the patient during scanning may impact results. User should exercise clinical judgement in the interpretation of results. • Do not use Butterfly iQ™ until the materials present in this manual have been reviewed and fully understood. Do not operate Butterfly iQ™ for purposes other than intended in this manual.
Page 9
Contact Support at: support@butterflynetwork.com • Do not use Butterfly iQ™ in the presence of flammable gases or anesthetics. Doing so can re- sult in a possible fire or explosion. • Butterfly iQ™ has not been evaluated or approved for use in hazardous locations as defined in the National Electric Code standard.
Electrical Safety Electrical Safety WARNINGS! • Before use, carefully inspect the probe. Always inspect the probe before and after cleaning, dis- infection, or use. Check the lens face, cable, housing, seams, and connector for signs of dam- age such as cracks, chips, abrasions, or leaks. To avoid the risk of electrical hazards, do not use the probe if there is any sign of damage.
Ultrasound Safety” publication. This publication is available as a PDF link in the Butterfly iQ™ App. • If Butterfly iQ™ is contaminated due to exposure to Creutzfeldt-Jakob disease, there is no ade- quate disinfecting procedure. • Use the correct clinical application presets for the associated body part being examined. Some applications require lower acoustic output limits.
Doing so may result in injury to the patient or operator. • Do not use the Butterfly iQ™ Application on a mobile device that does not meet minimum re- quirements. Using the Butterfly iQ™ Application on a mobile device that does not meet the mini- mum requirements may affect performance and image quality, possibly resulting in misdiagno- sis.
This chapter provides an overview of Butterfly iQ™. It includes information about its features, the components that are included in the system, requirements necessary to download, install, and use the Butterfly iQ™ App, and an overview of the user interface.
Internet Connectivity An Internet connection is required to download, install, or update the Butterfly iQ™ App from the Apple App Store or Google Play Store. An Internet connection is also required to log in and to archive studies to Butterfly Cloud. Other- wise, no Internet connection or wireless connectivity is required to use the mobile device.
System Components Butterfly iQ™ App The primary function of the Butterfly iQ™ App is general-purpose diagnostic imaging, for use by qualified and trained health-care professionals to enable the visualization and measurement of anatomical structures within the human body. The App is a free download from the Apple App Store or Google Play Store. The App and Butterfly account are re- quired to use the Butterfly iQ™...
Do not connect third-party probes to the Butterfly iQ™ mobile device and do not attempt to use the Butterfly iQ™ probe with other ultrasound systems. The Butterfly iQ™ probe is only for use with the Butterfly iQ™ App. Do not attempt to connect the probe to other ultrasound systems.
Figure 2. Charging Pad Components 1. Charging Pad 2. Charging Cable 3. Wall Adaptor NOTE The latest Butterfly iQ™ charger has a matte black finish. If you have a previous model please visit support.butterflynetwork.com for additional information on charging your probe. System Overview...
Presets Presets are a predefined set of imaging parameter values. When selected, the Butterfly iQ™ App automatically op- erates in accordance with the corresponding set of imaging parameter values. Presets available correspond to the clinical applications details in Indications for Use [5].
Butterfly iQ™ App, and charging the probe for use. Downloading and Installing the App You can download and install the Butterfly iQ™ App by visiting the Apple App Store or Google Play Store on your mobile device. Once in the App Store, search for "Butterfly iQ".
Charging the Probe Charging the Probe WARNINGS! • Use only cables, probes, chargers and accessories specified for use with Butterfly iQ™. Substi- tution of non-approved accessories may cause the system to perform improperly or cause injury to the patient or operator.
Page 21
Charging the Probe NOTE • The latest Butterfly iQ™ charger has a matte black finish. If you have a previous model please visit support.butterflynetwork.com for additional information on charging your probe. • Your exact charging pad may vary. • Butterfly iQ™ uses a wireless charging system. Do not attempt to insert your probe’s cable into the charging pad.
2. If the first button flashes, it indicates that the probe battery charge is too low to perform the study. To check the probe battery level using the Butterfly iQ™ App: • The probe battery status is displayed in the upper portion of the imaging screen.
Using the System 5. Using the System This chapter provides information and instructions for using Butterfly iQ™ to begin and end studies. It also provides information and instructions for freezing and unfreezing during live imaging, for performing measurements, and other imaging tools.
Uploading to Butterfly Cloud Uploading to Butterfly Cloud NOTE Depending on your platform, hardware, country and membership type, certain presets, modes and features may not be available. To archive a study: 1. When you finish capturing ultrasound images, tap the Capture Reel in the upper right corner of the screen. The Study screen is displayed.
Using Modes 6. Using Modes This chapter provides information and instructions for using modes when performing an ultrasound study. NOTE Advanced imaging capabilities may vary depending on the selected preset and status of paid sub- scription. Please visit support.butterflynetwork.com for the latest details on which preset has ac- cess to which modes.
Page 26
Using Spectral Pulsed-Wave Doppler Mode • Adjust gain of spectral trace by dragging finger left and right on the trace while trace is live. • Adjust the Scale to optimize for high or low flow by touching the Low Flow/High Flow control in the middle of the screen.
Page 27
Using Spectral Pulsed-Wave Doppler Mode NOTE Annotations and measurements may only be added to the Spectral trace region. Velocity measurements will be represented in cm/s as Peak Systolic Velocity (PSV), the absolute value of the vertical distance from the baseline first caliper dot, and End Diastolic Velocity (EDV), the absolute value of the vertical distance from the baseline of the second caliper dot.
Annotations 7. Annotations This chapter provides information and instructions for performing annotations on images in the Butterfly iQ™ App. Annotations can include linear measurements, ellipse measurements, and text annotations. Adding Annotations You can add annotations on any frozen image or images and clips in the capture reel.
Multiple calculation tools are available on the Butterfly iQ™ device depending on your membership status and geo- graphical location. The Butterfly iQ™ App helps you collect inputs to a calculation and process those inputs into clin- ically relevant outputs. When the calculation tool is initialized, the report may be viewed at any time by swiping up from the bottom of the screen.
The Auto Bladder Volume tool allows you to calculate bladder volume when capturing scans and visualize the blad- der in 3D. • The Butterfly iQ™ has the capability to acquire a 3D sweep while you hold the probe steady. A volume estimate is then calculated from this 3D sweep.
Butterfly Cloud is a web-based application that allows users to upload and review ultrasound exams from the Butter- fly iQ™ App. Users of the cloud can also document, bill and integrate Butterfly iQ™ into existing hospital systems PACS, VNA, EMR, and Modality Worklist. Butterfly Cloud also supports the acceptance of images from third-party ultrasound devices.
11. Using Butterfly TeleGuidance This chapter provides information regarding Butterfly TeleGuidance. The service enables users to call one of your available connections through your Butterfly iQ App for remote collaboration while scanning. WARNING! TeleGuidance is adjunctive and should not be solely relied upon to diagnose or treat COVID-19.
Maintenance 12. Maintenance This chapter provides information and instructions for storing, transporting, cleaning, and disinfecting the probe. Maintaining the Probe Storing and Transport: CAUTIONS! • Avoid storing the probe where the probe or its cable could be easily damaged. • Avoid transporting the probe unless it is well supported and secured. Secure the cable tightly to the probe when transporting it or carrying it.
While the Cleaning and Disinfection guidance contained here has been validated for effectiveness, a list of cleaning and disinfection products that are compatible with the Butterfly iQ™ probe, but not tested for effectiveness by Butter- fly, can be found in the "Compatible Cleaning and Disinfection Products" article available at support.butterflynetwork.com.
Cleaning and Disinfecting the Probe Disinfecting the Probe WARNING! Always inspect the probe before and after cleaning, disinfection, or use. Check the lens face, ca- ble, housing, seams, and connector for signs of damage such as cracks, chips, abrasions, or leaks.
Page 36
Cleaning and Disinfecting the Probe High Level Disinfection ® It is recommended that you use Cidex OPA by Ethicon US, LLC. Making sure your probe is compatible with HLD: 1. Tap to display the Settings screen. 2. Tap My iQ to display the My iQ screen. 3.
Perform the diagnostic test periodically. With normal use, monthly testing is best practice. The diagnostic test is only for the Butterfly iQ™ ultrasound probe. The App does not have the ability to assess the mobile device’s screen integrity.
Troubleshooting 13. Troubleshooting This chapter provides information and instructions for troubleshooting the system. WARNING! Do not use the probe if there is any sign of damage. Contact Support. See Getting Support [40] for more information. Troubleshooting CAUTION! Ignoring the App alerts and messages may result in the system becoming inoperable. Table 3, “Troubleshooting”...
Page 39
Hardware Recall alert appears instructions. This indicates that your mobile device no longer has a passcode. The Butterfly iQ™ App requires that the mo- Forced Log Out alert appears bile device to have a passcode for patient data security. Tap Settings to enable and configure the passcode for your mobile device.
Australia Contacting Support through the Butterfly iQ™ App You can contact Butterfly Support directly through the Butterfly iQ™ App and submit a request for help. To access support: 1. From the imaging screen, tap your user avatar (user uploaded photo or your initials) in the upper left corner.
Using the Butterfly iQ™ App on a mobile device that does not meet the minimum requirements may affect performance and image quality, possibly resulting in misdiagnosis. Butterfly iQ™ works on many Apple and Android devices. For the latest list of compatible mobile devices please visit support.butterflynetwork.com.
Table 6, “Environmental Operating Conditions” [42] lists the environmental conditions for the Butterfly iQ™ probe only. For details on the mobile device on which you run the Butterfly iQ™ App, refer to the accompanying documen- tation for your mobile device.
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Butterfly iQ™ Ultrasound System is used exceeds the applicable RF compliance level above, the Butterfly iQ™ Ultrasound System should be observed to verify normal operation.
Separation Distances Devices such as cellular/mobile phones, radio transmitters, and transceivers transmit radio waves (RF), which can create disturbances. The Butterfly iQ™ is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. If radiated and conducted electromagnetic disturbances are observed and performance is affected, the user or cus- tomer should take measures to mitigate, including relocation or reorientation of the system.
Page 45
ALARA training is provided by the American Institute of Ultrasound in Medicine (AIUM) in a booklet, “Medical Ultra- sound Safety.” This booklet is provided as a PDF link in the Butterfly iQ™ App and the Butterfly Cloud web interface. It provides training and educational information on ultrasound bio effects and biophysics, prudent use and imple- menting ALARA.
Page 46
Acoustic Output Transducer: Butterfly iQ™ Ultrasound System transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation General (Track 1 Only) Specific (Tracks 1 & 3) Power Color Doppler...
The ultrasound system maintains acoustic outputs below the appropriate limits for each application listed below. Non ophthalmic applications: System Probe TI Type TI Value SPTA.3 PA.3 Butterfly iQ™ 44.9 mW/cm 0.289 0.49 54.6 W/cm Ophthalmic applications: System Probe TI Type TI Value SPTA.3...
Page 48
Acoustic Output Table 13. B-Mode Non-Scan Index Label Non- Scan ≤1 >1 Scan aprt aprt Maximum Index Value 0.485 0.02 Pr.3 (MPa) 0.718 (mW) 4.40 min of [W (mW) TA.3 (cm) (cm) (cm) 5.83 (cm) (MHz) 2.19 2.41 X (cm) Dim of A aprt Y (cm)
Page 49
Acoustic Output Table 14. B-Mode + Color Non-Scan Index Label Non- Scan ≤1 >1 Scan aprt aprt Maximum Index Value 0.485 0.13 0.29 Pr.3 (MPa) 0.718 (mW) 17.4 min of [W (mW) 0.74 TA.3 (cm) (cm) 2.76 (cm) 5.83 (cm) 1.84 (MHz) 2.19...
Page 50
Acoustic Output Table 15. B + M-mode Non-Scan Index Label Non- Scan ≤1 >1 Scan aprt aprt Maximum Index Value 0.485 0.013 0.012 Pr.3 (MPa) 0.718 (mW) 2.64 0.63 min of [W (mW) TA.3 (cm) (cm) (cm) 5.83 (cm) (MHz) 2.19 2.41 1.56...
Page 51
Acoustic Output Table 16. B-Mode + Power Non-Scan Index Label Non- Scan ≤1 >1 Scan aprt aprt Maximum Index Value 0.485 0.13 0.29 Pr.3 (MPa) 0.718 (mW) 17.4 min of [W (mW) 0.74 TA.3 (cm) (cm) 2.76 (cm) 5.83 (cm) 1.84 (MHz) 2.19...
Page 52
Acoustic Output Table 17. B-Mode + Ophthalmic Non-Scan Index Label Non- Scan ≤1 >1 Scan aprt aprt Maximum Index Value 0.162 0.0015 0.0021 Pr.3 (MPa) 0.354 (mW) 0.0657 0.0657 min of [W (mW) TA.3 (cm) (cm) (cm) 1.65 (cm) (MHz) 4.80 4.80 4.80...
Page 53
Acoustic Output Table 18. Color + Ophthalmic Non-Scan Index Label Non- Scan ≤1 >1 Scan aprt aprt Maximum Index Value 0.162 0.00145 0.0167 0.0477 0.0340 Pr.3 (MPa) 0.355 (mW) 0.0633 0.719 0.719 0.719 min of [W (mW) TA.3 (cm) (cm) (cm) 1.65 1.22...
Page 54
Acoustic Output Table 19. Pulsed Wave Doppler Below Index Label At Sur- Below Surface At Surface Sur- face face Maximum Index Value 0.21 9.15E-02 0.14 0.10 Index Component Value 9.15E-02 4.08E-02 9.15E-02 0.14 (MPa) 0.44 (mW) 4.67 4.67 4.67 (mW) 4.48 4.48 (cm)
Butterfly Network is deeply invested in the preservation of the natural environment. Equipment may contain materi- als that pose a risk to the environment if proper disposal procedures are not followed. Recycle the Butterfly iQ™ probe and accessories at the end of their useful life and in accordance with local, state, provincial, and/or national regulations.
Symbols 16. Symbols This chapter lists and describes the symbols and icons that may be used on the Butterfly iQ™, its accessories, and packaging. Symbols Table 20, “Symbols” [56] lists and describes a set of symbols for medical electronic equipment that classify a con- nection or warn of potential hazards.
Page 57
Symbols Refer- Symbol Standard Title Description ence Model name Model name for the device. Indicates a medical device that has not been subjected to a steriliza- ISO 15223-1 5.2.7 Non-sterile tion process. Refer to instruction ISO 7010 M002 To signify that the instruction manual/booklet must be read manual/ booklet Operator’s manual;...
Need help?