Download Print this page

Land Instruments International SYSTEM 4 M1 600/1600C L Installation Manual page 3

Fibroptic thermometer

Advertisement

TARGET SIZE TABLES - FIBROPTIC M1 AND M2
Target diameter
15.0mm
12.3mm
9.5mm
6.8mm
4.0mm
0.59in
0.48in
0.37in
0.27in
0.16in
0mm
25mm
50mm
75mm
100mm
0in
0.98in
1.97in
2.95in
3.94in
Target distance
Optic Head A10, Field of View 25:1 (Red spacer)
Target diameter
14.3mm
11.1mm
7.8mm
4.6mm
1.3mm
0.56in
0.44in
0.31in
0.18in
0.05in
0mm
25mm
50mm
75mm
100mm
0in
0.98in
1.97in
2.95in
3.94in
Target distance
Optic Head A10, Field of View 75:1 (Red spacer)
Target diameter
11.2mm
11.0mm
10.7mm
10.5mm
10.2mm
10.0mm
0.44in
0.43in
0.42in
0.41in
0.40in
0.39in
0mm
50mm
100mm
150mm
200mm
250mm
0in
1.97in
3.94in
5.91in
7.87in
9.84in
Target distance
Optic Head A25, Field of View 25:1 (Blue spacer)
Target diameter
11.0mm
9.5mm
7.9mm
6.4mm
4.8mm
3.3mm
0.43in
0.37in
0.31in
0.25in
0.19in
0.13in
0mm
50mm
100mm
150mm
200mm
250mm
0in
1.97in
3.94in
5.91in
7.87in
9.84in
Target distance
Optic Head A25, Field of View 75:1 (Blue spacer)
Target diameter
10.3mm
12.8mm
15.4mm
17.9mm
20.5mm
23.0mm
0.41in
0.50in
0.61in
0.70in
0.81in
0.91in
0mm
100mm
200mm
300mm
400mm
500mm
0in
3.94in
7.87in
11.81in
15.75in
19.69in
Target distance
Optic Head A50, Field of View 25:1 (No spacer)
Target diameter
10.1mm
9.4mm
8.7mm
8.1mm
7.4mm
6.7mm
0.38in
0.37in
0.34in
0.32in
0.29in
0.26in
0mm
100mm
200mm
300mm
400mm
500mm
0in
3.94in
7.87in
11.81in
15.75in
19.69in
Target distance
Optic Head A50, Field of View 75:1 (No spacer)
TARGET SIZE TABLES - FIBROPTIC M3
Target diameter
15.9mm
13.2mm
8.8mm
0.63in
0.52in
0.35in
0mm
25mm
125mm
0in
0.98in
4.92in
Target distance
Optic Head D10, Field of View 20:1 (Green spacer)
5.2in
0.20in
INSTALLATION INFORMATION
Laser Alignment Unit (optional)
Press button to illuminate target area with laser
light from optic head. Adjust the thermometer
mounting until optic head is aligned on required
target.
125mm
4.92in
Refer to User Guide for full instructions.
14.2mm
0.56in
300mm
11.81in
6.2mm
0.24in
View on Front
300mm
11.81in
26.1mm
1.03in
600mm
23.62in
10.1mm
0.38in
NOTE
The thermometer body must be earthed
600mm
electrically via it's mountings
23.62in
Target diameter
13.7mm
13.5mm
10.5mm
7.7mm
5.0mm
10.2mm
0.54in
0.53in
0.41in
0.30in
0.20in
0.40in
50mm
75mm
100mm
125mm
0mm
1.97in
2.95in
3.94in
4.92in
0in
Target distance
View on Side
Mounting Pad (Tripod Mounting)
2 holes, ¼" BSW, 7mm (0.28in) deep on
25mm (0.98in) centres
View on Bottom
Target diameter
13.2mm
13.0mm
12.7mm
12.5mm
17.7mm
0.52in
0.51in
0.50in
0.49in
0.70in
50mm
100mm
150mm
200mm
250mm
300mm
1.97in
3.94in
5.91in
7.87in
9.84in
11.81in
Optic Head D25, Field of View 20:1 (Purple spacer)
Target distance
View on Rear
Cable Schedule
Processor
Pin Nº
1
2
3
4
5
6
7
13.0mm
15.4mm
17.8mm
20.2mm
22.6mm
25.0mm
0.51in
0.61in
0.70in
0.80in
0.89in
0.98in
0mm
100mm
200mm
300mm
400mm
500mm
0in
3.94in
7.87in
11.81in
15.75in
19.69in
Optic Head D50, Field of View 20:1 (No spacer)
Lightguide
Ensure that the lightguide is not subjected to
tension or other undue force.
Minimum bend radius = 50mm (2.0in)
Electrical Connection
Keyway
Lugs
Locking Sleeve
Markers
Align the two red markers (i.e. align the lugs
on the plug of the pre-wired cable with the
keyway in the connector on the thermometer).
Push the plug into the connector.
Cable Colour
Function
Yellow
T+
Blue
T-
White
E+
Screen
Screen
Red
V+
Black
V-
Green
E-
FOT010098