2.
COMPONENT NAMES AND BASIC PACKAGE
① ② ③ ④ ⑤ ⑥
24
① PRESTO
Control Unit
⑧ Plumbing Hose
② Pressure Gauge
⑨ PRESTO Handpiece
③ Regulator Knob
⑩ Handpiece Tubing
④ Filter Inspection Window
⑪ Foot Pedal Connector
⑤ Handpiece Connector
⑫ Hose Stopper
⑥ Tube Stopper
⑬ Plumbing Connector
⑦ Foot Pedal
⑭ Unit Mounting Screws
3.
WARRANTY
We provide a limited warranty for our products. We will repair or replace the products if the cause of failure is due to the following
manufactures defects. Please contact us or your local distributor for details.
① Defect in manufacturing.
② Any shortage of components in the package.
③ Where damaged components are found when initially opening the package.
(This shall not apply if the damage was caused by the negligence of a customer.)
4.
CONTACT US
For your safety and convenience when purchasing our products, we welcome your questions.
If you have any questions about operation, maintenance and repair of the product, please contact us.
Contact Us
Contact Us
For U.S. Market
Company Name
:
Industrial Div.
Business Hours
: 8:30am to 17:00pm (CST)
(closed Saturday, Sunday and Public Holidays)
U.S. Toll Free No.
: 800-585-4675
Telephone No.
: 847-843-7664
: 847-843-7622
Fax No.
Web Address
: www.nskamericacorp.com
For Other Markets
Company Name
:
Business Hours
: 8:00am to 17:00pm
(closed Saturday, Sunday and Public Holidays)
Telephone No.
: +81 (0) 289-64-3520
e-mail Address
: webmaster-ie@nsk-nakanishi.co.jp
5.
SPECIFICATIONS
-1
speed
320,000min
(rpm)
Recommended Air Pressure
0.2 - 0.25MPa (29 - 36.3psi)
Handpiece Dimensions
16.6 (D) x 130 (H) mm
Handpiece Weight
71g
Noise Level at 1m distance
Less than 70dB (A)
23
21
22
⑮ Handpiese Tubing Clamps
㉒ Wrench Flats
⑯ Bur Push-in Tool
㉓ Chuck Release Ring
⑰ Cartridge Wrench
㉔ Tube Clamp Ring
⑱ Rotor Shaft Nut Wrench
㉕ Operation Manual
⑲ Handpiece Stand
⑳ Air Supply Connector
㉑ Cartridge
Control Unit Dimensions W120 x D102 x H165 (mm)
Control Unit Weight
720g
Proper Air Pressure
40Nℓ / min
Inside Diameter
1.6mm Bur Only
Operation Environment
Transportation and Storage Environment
6.
SYSTEM ASSEMBLY PROCEDURES
6 - 1 Mounting of the Control Unit
The Control Unit can be secured on a wall, on a work-bench, or under a work-bench.
When securing on a wall surface, etc., ¿x the Control Unit with unit ¿xing screws in the two screw holes on its back.
6 - 2 Installation of Plumbing Hose
CAUTION
25
・ Push the plumbing hose into the plumbing connector until it is securely set.
Otherwise, air may leak.
・ Pushing the white ring, on the plumbing connector, gently remove the tube.
Push the plumbing hose into the plumbing connector located at the regulator on the left-
hand side of the Control Unit until it is securely set to make connection as shown in Fig. 1.
Check if the hose is securely installed by tugging it after connection.
Connect the other end of the plumbing hose to the air line. At this time, use the attached
connector if necessary.
6 - 3 Installation of foot Pedal
Loosen and remove the hose stoppers (two) from the foot pedal connectors on the back
of the Control Unit, and insert them into the foot pedal hoses. At this time, insert the hose
stoppers so that their screws come outside (Control Unit side) (Fig. 2). For wall-hanging
use, pass the hoses through the hole shown in Fig. 2.
Insert the "1" - marked end of each hose according to the instructions on the label. After
insertion of both hoses, securely tighten the hose stoppers.
6 - 4 Installation of Handpiece
Loosen and remove the tube stopper from the handpiece
connector on the right-hand side of the Control Unit, and insert
it into the handpiece tubing. At this time, insert the tube stopper
so that its screw comes outside (Control Unit side) (Fig. 3).
Insert the handpiece tubing into the handpiece connector, and
securely tighten the tube stopper.
Insert the tube clamp ring into the other end of the handpiece
tubing as shown in Fig. 4, and insert it into the handpiece
tubing socket as shown in Fig. 5.
6 - 5 Installation of Handpiesce Tubing Clamps
Locate a Handpiece Tubing Clamp (two pieces supplied) at a proper location to conveniently position the handpiece tubing.
OPERATION OF CONTROL UNIT
7.
7 - 1 Set the Drive Air Pressure
Supply air, and adjust to 0.25MPa (36.3psi) by pulling the Regulator Knob upward and turning.
When adjustment is made, push the knob down to lock (Fig.6).
7 - 2 Operation
Stepping on the Foot Pedal starts rotation.
7 - 3 For Finishing Work
Close the main air valve.
8
.
OPERATION OF HANDPIECE
① As the chuck of the PRESTO handpiece is a push-in type, set the
1.6mm burs and push in as shown in Fig.7.
If the end of the bur is tapered push in using the provided bur-
inserting wrench as shown in Fig.8.
② When removing the bur, turn the chuck release ring in the direction
of the arrow in Fig.9.
Temperature
Humidity
MAX.75%
0 - 40°C
(No condensation)
-10 - + 50°C
10 - 85%
White Ring
White Ring
Plumbing Hose
Plumbing Hose
Hose Stopper
Hose Stopper
When using by Wall-hanging
When using by Wall-hanging
way, this hole can also be used.
way, this hole can also be used.
Handpiece Connector
Handpiece Connector
Handpiece Connector
Handpiece Connector
Tube Stopper
Tube Stopper
Handpiece Tubing
Handpiece Tubing
Fig. 3
Fig. 7
Atmospheric Pressure
700 - 1,060hPa
500 - 1,060hPa
Plumbing connector
Plumbing connector
Plumbing connector
Fig. 1
Foot Pedal Connector
Foot Pedal Connector
Fig. 2
Tube Clamp Ring
Tube Clamp Ring
Fig. 4
Fig. 5
Regulator Knob
Regulator Knob
Fig. 6
Fig. 8
Chuck Release Ring
Chuck Release Ring
Fig. 9