Technical Data Operating Environment Transport and Storage Environment Description on device Safety measures Shipping contents Set-up Operation The TCM Endo III unit The footswitch Automatic Limiter (AL) Automatic Protection (AP) Memory Disinfection, Cleaning and sterilization Control unit and footswitch Mikromotor Endo...
TCM Endo III NOUVAG AG 1 Description 1.1 Operation and functions The TCM Endo is a microprocessor controlled endodontic electronic motor system. The slow speed, high torque rotary instrument uses an electronic torque controlled motor to achieve a faster and easier root canal preparation. The speed is held constant under varying loads until the selected torque level is reached.
TCM Endo III 2 Safety measures Your safety, the safety of your team, and, it goes without saying, the safety of your patients is for Nouvag AG the first priority. It is therefore vital that the following measures be strictly observed: •...
TCM Endo III NOUVAG AG 5 Operation 5.1 The TCM Endo III unit Starting the device: Press green “ON/OFF“ switch (rear of unit). Display: The digital display indicates the current drill speed or the selected torque level. Keypad/Control functions: „Motor“: Switch On/Off for micromotor (green light illuminates when the motor is in operation) „▲“:...
NOUVAG AG TCM Endo III 5.2 The footswitch The TCM Endo III can be operated with or without footswitch. Press footswitch to start the motor. Footswitch not pressed:....Motor off Footswitch fully pressed: ....Motor on The tool speed corresponds to the speed adjusted on the control unit.
TCM Endo III NOUVAG AG Adjusting the AP: Press the “ATC” key. In the display appears the torque mode set (for example [AL 20]). Pressing the keys “▲” and“▼” will increase and decrease the torque level respectively. By pressing the “ATC SET” briefly once more the digital display switches back automatically.
NOUVAG AG TCM Endo III 6.2 Mikromotor Endo The micromotor, cable and plug should be cleaned after each use to prevent the build up of deposits and debris that can destroy it, if not removed. Wipe it with a clean cloth dampened with disinfectant solution.