PortaCount
•
Set up PortaCount
®
making sure ambient particle count is over 3000 pcm
for testing half-face respirators and is over 10000 pcm
for full-face respirators.
•
Ask if the individual has smoked or vaped in the last
60 minutes or eaten in the last 15 minutes.
•
Ask if they are medically fit to take the test – it will
involve gentle continuous exercise for around 8 minutes.
•
Check that the individual is clean shaven and ensure
they are aware that they must always be clean shaven
when wearing RPE at work.
•
Ask if other PPE is worn during work (glasses, hearing
aids, visors etc). Any required PPE must be worn during
the Fit Test.
•
Explain why you are conducting a Fit Test.
•
Briefly explain the test exercises and reiterate that the
subject will be stepping/cycling/pacing throughout the
test in addition to these.
•
Probe the respirator and attach tubing.
•
Ensure a P3 or P3 combination filter is attached of a
similar size/weight/shape to what will be worn in the
work environment.
•
Instruct the subject how to don and doff the mask.
•
Visually check that their mask is fitted correctly, checking
that no hair is trapped in the face seal.
•
Fill out details for the individual and the make/model of
the respirator they are Fit Testing into the PortaCount
For more information call 01933 672180 email: info@fullsupportgroup.com
© Copyright 2019
Pre Fit Test Checklist
®
machine and run daily check,
www.fullsupportgroup.com
3
3
.
®
FSI-0141-B