HDM97
Pocket
User Manual
Reference Meter Series
HDM97
BL
HDM97
BH
HDM97
BN
HDM97
BO
HDM97
BP
HDM97
BQ

Advertisement

Table of Contents
loading

Summary of Contents for IBP HDM97 Pocket Series

  • Page 1 HDM97 Reference Meter Series Pocket User Manual HDM97 HDM97 HDM97 HDM97 HDM97 HDM97...
  • Page 2 Preface The information contained in this is subject to change without notice. IBP Medical GmbH, its distributors and subsidiaries take no responsibility for any errors or omissions in this document. The contained software is being delivered on the basis of a general licence contract or in single license.
  • Page 3 SAFETY INSTRUCTIONS For your and your patients safety read and consider the safety instructions below carefully • Read the entire manual carefully before using the HDM97Pocket. • Keep the instrument away from unauthorised persons. • Only use the HDM97Pocket as reference instrument to measure Conductivity, Temperature, Pressure, Flow and pH depending to it's equipped features.
  • Page 5: Table Of Contents

    Table of content Introduction ........................... 3 Application ............................ 3 License ............................4 Europe ................................4 USA ................................4 Delivery Content ........................... 5 Warranty ............................6 Battery and Electronic Recycling ....................6 Product Details..........................7 Appliance Part ............................... 7 The options at a glance ..........................8 Measuring Channels .............................
  • Page 6 Inhaltsverzeichniss Assigning graphs to hardware channels ..................... 45 List of channels ............................46...
  • Page 7: Introduction

    Delivery Introduction meter series are reference instrument designed for the calibration of medical devices such HDM97 Pocket as hemodialysis machines. Please read this entire manual carefully before the first use of the instrument. Application Depending on the version the may be equipped with up to 5 acquisition channels: HDM97 Pocket Channel...
  • Page 8: License

    Delivery License Europe EC DECLARATION OF CONFORMITY We hereby declare, that the following product(s) conform to the below mentioned European Directive. Manufacturer IBP Medical GmbH Ikarusallee 15 30179 Hannover Product HDM97BL HDM97BH HDM97BN HDM97BO HDM97BP HDM97BQ Relevant EC-Directives 2004/108/EG EMV-Richtlinie...
  • Page 9: Delivery Content

    Delivery Delivery Content Pos. Description Version Standard Option  HDM97 BL, BH, BM, BN, BO, BP, BQ Pocket  Conductivity/Temperature Probe BN, BO, BP, BQ  Dialyser Line Adapter BN, BO, BP, BQ  Hose and adapter for pressure measurement BH, BO, BP, BQ ...
  • Page 10: Warranty

    Delivery Warranty IBP Medical GmbH warrants that it will repair or replace, at its option, any defective or malfunctioning part without charge for the terms listed below. Parts used for repair or replacement are warranted for the remaining warranty period only.
  • Page 11: Product Details

    General Product Details Appliance Part HDM97 HDM97 Pressure Flow HDM97 HDM97 Pressure Conductivity Conductivity Temperature Temperature HDM97 HDM97 Pressure Conductivity Flow Pressure Conductivity Temperature Temperature HDM97 BQ with Option C (Conductivity Probe Connector) Flow Pressure Conductivity Temperature Depending on the configuration the appliance part consists of a BNC inlet for pH electrode connection, a male connector for pressure transducer inlet, a connectorfor the flow probe and a directly attached cable which leads to the conductivity / temperature probe or connector for the probe.
  • Page 12: The Options At A Glance

    General The options at a glance Your device might be equipped with one of the following options. Option H Option H is a higher accuracy in pressure measurement. Please note that the sensor is not fluid resistant with this option. Option identification is an H behind the product name, for example HDM97 BQH.
  • Page 13: Measuring Channels

    General Measuring Channels Only use accessories provided with the . Use of other devices may damage the instrument and HDM97 Pocket will void the warranty. Conductivity/Temperature Probe The conductivity/temperature transducer is a quadropole design for greater accuracy and longevity than other designs.
  • Page 14: Charging The Battery

    General Charging the battery is internally equipped with a high performance Lithium-Ion-Polymer battery. It can be HDM97 Pocket charged using the external power supply, a PC or a laptop which is connected via the USB cable to the instrument. The battery of the HDM97 can only be charged from USB ports which are capable of Pocket...
  • Page 15: Keypad

    General Keypad Main functions The keypads main functions of the are listed below: HDM97 Pocket for switching the instrument on and off for selecting the measuring channel for selecting the unit to display or temperature coefficient leads to channel dependent special functions to tare the pressure transducer The keypad layout is shown in the left picture.
  • Page 16: Switching On The Instrument

    General Switching on the instrument The instrument is switched on and off by pressing the key. Display test during start up Immediately after switching on the instrument the display shows all possible segments dark for two seconds. Compare the display with the picture on the right side.
  • Page 17: Battery Gauge Display During Operation

    , and the main PCB itself contains the battery protection circuit. If the battery is not handled in HDM97 Pocket the right way, there is extremely danger of fire and explosion. Therefore, this battery can only be replaced from IBP personnel who have run through a special training in handling Lithium-Ion batteries.
  • Page 18 Base Installation Basic settings To invoke the installation menu please follow the steps below: 1. Switch on the meter using the key. 2. Press during the display test. 3. After this procedure the instruments starts the basic settings dialog instead of entering any measurement functions.
  • Page 19 Base Installation → CAL-RES → Press 5 seconds until the counter in the display reaches zero. Then a calibration dataset reset will be performed. This takes a couple of seconds. The instrument performes a restart afterwards. → LANGUAGE → → DEUTSCH german language →...
  • Page 20: Value Adjustment

    Base Installation → PR H-Res only available with version BH, BM, BO, BP and BQ → → High Resolution reading enabled → → High Resolution reading disabled Attention: The higher resolution does not give a higher measuring accuracy. Then to apply and save this setting, to discard this setting.
  • Page 21: Measuring

    Adjustment and Calibration Measuring By pressing the key the user can select between all available channels of the HDM97 Pocket These channels are • Conductivity and temperature Display "CD/TEMP" only Version BN, BO, BP, BQ • Temperature Display "TEMP" only Version if enabled BN, BO, BP, BQ •...
  • Page 22: Background Of Temperature Coefficients

    Adjustment and Calibration Background of temperature coefficients This is a topic that is frequently misunderstood and often neglected in hemodialysis. A solution’s conductivity will change according to temperature. With increasing temperatures, the measured solution’s conductivity will increase, too. To achieve meaningful measurement results, the conductivity value displayed is compensated to 25°C.
  • Page 23: Temperature Compensation

    Adjustment and Calibration Temperature compensation Listed below are the temperature coefficients used by most major hemodialysis machine and reverse osmosis manufacturers. We recommend that you double check this data with documentation from your machine’s manufacturer Display To use for Temperature Temperature coefficient coefficient...
  • Page 24: Selecting The Temperature Coefficient

    Adjustment and Calibration Selecting the temperature coefficient set To select the current temperature coefficient set press the function key while in conductivity mode. The temperature coefficients are shown with their respective temperatures alternating every 3 seconds. The upper line shows the name of the temperature coefficient set, the middle line shows the temperature and the lower line shows the temperature coefficient associated with the temperature above.
  • Page 25 Adjustment and Calibration Adjustment of customer specific temperature coefficient "USER TC" Pressing the navigation key while adjusting the temperature coefficient to "USER TC" the enters a second dialog to adjust the value itself instead of directly going back to measuring HDM97 Pocket mode.
  • Page 26: Temperature Measurement

    Adjustment and Calibration Temperature measurement This mode is only available if enabled in the basic settings menu. This mode exists to save power in case conductivity measurement is not needed. → TEMP (available in versions BN, BO, BP and BQ, if enabled in the Basic Settings) →...
  • Page 27: Pressure Drop Measurement

    Adjustment and Calibration Offset balancing (tare) Caused by drift of the pressure probe, the pressure display may sway slightly around zero. The HDM97 Pocket is equipped with a tare function which sets the value to zero if drift occurs. While performing this function there must not be any pressure on the pressure transducers inlet.
  • Page 28: Ph Measurement

    Adjustment and Calibration pH measurement → (available only in version BP, if enabled in the Basic Settings) → selection between compensation 20°C, 25°C, 37°C and measured temperature → → reading min. value reading max. value Returns automatically back to measuring mode after 5 sec →...
  • Page 29: Flow Measurement

    Adjustment and Calibration Flow measurement This mode is available only if a flow sensor is connected. → Flow (available only in version BL and BQ, if enabled in the Basic Settings) → Attenuation → → reading min. value reading max. value Returns automatically back to measuring mode after 5 sec →...
  • Page 30: Timer

    Adjustment and Calibration TIMER → TIMER (if enabled in the Basic Settings) → → Timer start → → Timer stop afterwards → → Timer reset afterwards the cycle "Start, Stop and Reset" can started again → no function → no function →...
  • Page 31: Adjustment And Calibration

    Adjustment and Calibration Adjustment and Calibration Precautions Use the adjustment functions only if you are fully aware of their consequences. Wrong adjustment values will lead to false measurement values and in consequence these false values lead to hazards for patients. Definitions Adjustment An adjustment is a process in which properties of a measuring device are changed in that way that its scale or...
  • Page 32: Handling Of Reference Solutions

    Adjustment and Calibration Handling of Reference Solutions IBP Medical standard solutions are produced under ISO13485:2003 quality management. They are traceable to NIST and PTB Standards Reference Materials and are sealed with tamper-evident packaging. To ensure standard solution and calibration/verification accuracy •...
  • Page 33 Adjustment and Calibration Pressure The pressure transducer has no nonlinear drift so an adjustment is normally not necessary. Use the tare function to balance the pressure transducers environmental pressure change drift. The pressure channel is adjusted using 7 sampling points. This improves the accuracy of the pressure measurement, because it recalculates slightly nonlinearities of the pressure transducer.
  • Page 34 Adjustment and Calibration Conductivity The adjustment of the conductivity channel is done by adjusting the cell parameter. To achieve this only one adjustment point is necessary. Point Display Cell parameter Initial Conductivity "CD xx.xx°" 0.xxxx 14.000 mS/cm First of all set the displayed adjustment value to the value of your reference solution. Then make sure that the solution has a stable temperature of exactly 25.00 °C.
  • Page 35: Maintenance Of Hdm97Pocket

    HDM97 which you can repair yourself. Contact our technical support team in Pocket the event of any malfunction. If the device should become damaged, or malfunction, send it to IBP or your distributor for repair. Storage Keep the device in a dry place. Suitable is, for example, the original packaging in which you have received , or the carrying case offered by IBP as accessory.
  • Page 36 In other words the accuracy of the references to adjust or calibrate the HDM must have at least a tree times higher accuracy. For liability reasons you should not carry out these tests and calibration and verification yourself. If IBP performs the calibration, you will receive a calibration certificate in accordance to ISO9001 which...
  • Page 37 Maintenance Tolerances For calibration, the instrument must meet the deviations listed below. Channel Reference Value Allowed deviation Notes  0,1 25,00  0,1 34,00  0,1 37,00 Temperature  0,1 40,00  0,1 55,00  0,1 80,00  0,60 µS/cm 74,00 µS/cm ...
  • Page 38: Troubleshooting

    Troubleshooting Troubleshooting Problem: USB cable connected, but the battery is not being charged Reason: USB port of laptop does not provide enough power Solution: Connect the meter to the battery charger or a USB port which is able to supply at least 500 mA.
  • Page 39: Technical Data

    Interface Technical data Conductivity 0 to 30,0 mS/cm, 4-Pole probe, Resolution and accuracy: Range Resolution Accuracy  0,6 µS/cm 0.0 – 200 µS/cm 0,1 µS/cm  6µS/cm 201 - 1999 µS/cm 1 µS/cm  0,03 mS/cm 2,00 – 30 mS/cm 0,01 mS/cm Reference temperature 25°...
  • Page 40 Interface HDM97 BH30 Range: -12 to 30 psi -0.85 to 2 bar -650 to +1550 mmHg Resolution: 0.01 mmHg Accuracy: General 0.05% full scale Media Compatibility: Gasses HDM97 BH50 Range: -12 to 50 psi -0.85 to 3,5 bar -650 to +2585 mmHg Resolution: 0.1 mmHg Accuracy:...
  • Page 41 Interface Internal power source Internal Li-Ion-Polymer battery, rechargeable with USB 5V DC / max. 500mA Operating time, depending in channel, up to 60 hours. Nominal Voltage 3,7 V Max. charge voltage 4,2 V (internally generated an monitored) Capacity 850 mAh or 3,1 Wh Battery charge time approx.
  • Page 42: Pc Interface

    Interface PC Interface USB Interface The USB interface of the is electrically isolated from the appliance part. Two drivers are HDM97 Pocket needed for installation. One driver is made for invoking the interface like any other USB device and the other driver is used to emulate a serial COM port device.
  • Page 43: Driver Installation, Example For Microsoft Windows Xp Service Pack 2

    Interface ® ® Driver installation, example for Microsoft Windows XP Service Pack 2 Insert the Installation-CD in the CD or DVD drive of your computer. Connect the via the USB-cable with a HDM97 Pocket free USB port of the running computer. It does not matter whether the instrument is switched on or off.
  • Page 44 Interface The driver installation is in progress. This can take a couple of minutes. The driver installation was successful, click on button "Finish" complete procedure. After you have finished this procedure another similar Wizard will popup. Do exactly the same you did in the first procedure as described earlier.
  • Page 45: Example

    Interface Data transfer To receive measurements from the via the USB interface, the following abbreviations must be HDM97 Pocket sent via the interface. The entire communication takes place in readable ASCII-characters. Command Reaction of the HDM97 HDM97 Pocket Pocket HDMCMO activates the data transfer HDMCMF deactivates the data transfer...
  • Page 46: Ibpview

    IBPView IBPView What is IBPView IBPView can be obtained optionally for the and can be used as well for other HDM HDM97 Pocket instruments. A demo version is included in the folder \Software\ on the driver CD. This demo can not save, export or print and the acquisition time is limited to 30 seconds.
  • Page 47: Installation

    Software's help function. IBPView is intentionally not limited to HDM instruments. It can be configured independently and therefore achieves its maximum flexibility to be ready for future IBP products. Hardware Interface IBPView uses serial COM ports for communication. This can be common RS232 serial interfaces like many...
  • Page 48: Selecting Hardware

    IBPView Template projects Project files are named in the way "[Name].ibp". To get an easy start with IBPView there are preconfigured version specific templates for each type of instrument. These templates include all common setting for start working with the application. Only the number of the COM port must be adjusted to the number the driver installation has chosen if it is not coincidentally the same number is used.
  • Page 49 IBPView Assigning graphs to hardware channels In the main window the list of channels is located left. The assignment of a particular graph to a specific hardware channel is done by performing a double click on the channel's name in the list. This double click leads to a dialog "Cannel configuration"...
  • Page 50 IBPView List of channels Depending on the version the is equipped with up to 4 channels HDM97 Pocket HDM97 HDM97 HDM97 HDM97 HDM97 HDM97 Ch. Description Range Unit     Conductivity 0.000 .. 30.00 mS/cm    ...