Do you have a question about the Easidew PRO and is the answer not in the manual?
Questions and answers
Subscribe to Our Youtube Channel
Summary of Contents for Michell Instruments Easidew PRO
Page 1
Intrinsically Safe Field Display User’s Manual Easidew PRO I.S. Field Display 97321 Issue 2.3 May 2016...
Page 2
Please fi ll out the form(s) below for each instrument that has been purchased. Use this information when contacting Michell Instruments for service purposes. Instrument Code Serial Number Invoice Date Location of Instrument Tag No Instrument Code Serial Number Invoice Date...
I.S. Field Display User’s Manual Contents Safety ..........................vi Toxic Materials ......................vi Repair and Maintenance ....................vi Safety Conformity ......................vi Abbreviations ........................vi Warnings ..........................vi INTRODUCTION ....................1 Operation ......................1 1.1.1 Controls ....................... 2 INSTALLATION ....................3 4-20 mA Input ....................3 System Design for Gas Hazardous Areas ............. 4 2.2.1 Transmitter Loops ..................
Page 5
Figures Figure 1 Indicator Block Diagram ................2 Figure 2 Loop-Powered by a Galvanic Isolator (Easidew PRO I.S.) .......4 Figure 3 Loop-Powered by a Galvanic Isolator (Easidew TX I.S.) .........4 Figure 4 Alternative Circuits for Remote Indication in a Hazardous Area ......5 Figure 5 Installation Procedure .................6...
Repair and Maintenance The instrument must be maintained either by the manufacturer or an accredited service agent. Refer to www.michell.com for details of Michell Instruments’ worldwide offi ces contact information. Safety Conformity This product meets the essential protection requirements of the relevant EU directives. Further details of applied standards may be found in the product specifi...
After fi ve seconds the instrument displays the input current using the calibration information stored in the instrument memory. If the loop current is too low to power the instrument it will display the error message LPLo Michell Instruments...
I.S. Field Display User’s Manual INTRODUCTION Less than 1.2 V 88888 Analog 4-20 mA to digital input Power converter Supply alarm 1 Common Optional Optional Loop powered Alarms Backlight alarm 2 Separately powered Indicator Block Diagram Figure 1 1.1.1 Controls The I.S.
1 & 3 will not exceed those specifi ed by clause 5.7 of EN 60079-11 for simple apparatus. This simplifi es the application and documentation of intrinsically safe loops containing the I.S. Field Display. Michell Instruments...
5 V if the backlight is loop-powered. See Section 3.4.1. Figures 2 and 3 illustrate typical applications in which an I.S. Field Display is connected with an Easidew PRO I.S. and an Easidew TX I.S. transmitter powered by a galvanic isolator.
9999 driven via an intrinsically safe interface from a 4-20 mA safe area signal. The type Easidew PRO I.S. Field Display of intrinsically safe interface is not critical. Either a Zener barrier or a galvanic isolator may be used, providing that the Ui, li and Pi of the I.S.
/ cable entries. Installation Procedure Figure 5 illustrates the instrument installation procedure. Easidew PRO I.S. Field Display Remove the instrument terminal cover by unscrewing the two captive 'A' screws. Mount the instrument on a fl at surface and secure with screws or bolts through the two 'B' holes.
I.S. Field Display User’s Manual INSTALLATION Two M6 clearance 141mm holes for surface (5.5”) mounting 9999 Easidew PRO I.S. Field Display 72mm (2.8”) 117mm (4.6”) Separate Control terminal cover compartment Three M20 x 1.5 tapped cable entries. 91mm The FM approved version (3.6”)
I.S. Field Display User’s Manual OPERATION OPERATION The I.S. Field Display will be supplied already confi gured to order and only requires installation in the user's equipment. If it is necessary to re-range, change the resolution or set the alarms (if fi tted), see the sections below: Confi...
See Section 3.1.5. Calibration of display using internal references Enables the zero and span of the I.S. Field Display to be adjusted without the need for an accurate input current or disconnection from the 4-20 mA loop. See Section 3.1.6. Michell Instruments...
Confi guration FunC Menu and press . If the function is set as required, press to return to the Confi guration Menu, or press to change the setting, followed by to return to the Confi guration Menu. Michell Instruments...
I.S. Field Display User’s Manual OPERATION 3.1.2.1 ‘ ’ Linear (Default) Provides a linear relationship between the 4-20 mA input current and the I.S. Field Display. 3.1.3 Resolution: ' rESn This function defi nes the resolution of the least signifi cant display digit. Decreasing the display resolution can improve the readability of a noisy signal.
I.S. Field Display displays and aborts the calibration. FaiL • If the zero current is greater than the span current the instrument will be reverse acting, i.e. an increasing input current will cause the display to decrease. Michell Instruments...
I.S. Field Display User’s Manual OPERATION 3.1.6 Re-Range Display or Calibrate Using Internal Reference: ‘ ’ Using the function the I.S. Field Display can be calibrated without the need to know the value of the 4-20 mA input current, or to disconnect the instrument from the 4-20 mA loop.
Confi guration Menu. The revised security CodE code will be activated when the I.S. Field Display is returned to the display mode. Please contact Michell Instruments (see contact details at www.michell.com) if the security code is lost. 3.1.10 Reset to Factory Defaults: ‘...
I.S. Field Display User’s Manual OPERATION 3.1.11 Under and Over-Range If the numerical display range of the I.S. Field Display is exceeded, all the decimal points will fl ash as shown below: Under range -9.9.9.9 Over range 9.9.9.9 Although not guaranteed, most indicators will continue to function normally with an input current between 1.8 mA and 4 mA.
Alarm Alarm setpoint setpoint Output Output Output Output ALARM Open Closed Closed Open 4-20 mA Input 4-20 mA Input Alarm Outputs Figure 8 Confi gurable functions for each alarm include adjustable setpoint, hysteresis, alarm delay and alarm accept. Michell Instruments...
Press 5 times until is shown. Alr1 Press again. Summary of alarm confi guration functions: Display Description of function Alarm enable EnbL Enables or disables the alarm without changing the alarm parameters. See Section 3.3.4. Michell Instruments...
I.S. Field Display User’s Manual OPERATION Alarm setpoint 1 Adjusts the alarm setpoint. The alarm is activated when the display equals the setpoint. See Section 3.3.5. Alarm function Hi.Lo Defi nes the alarm function as High or Low. See Section 3.3.6. Normally open or normally closed output no.nC Sets the alarm output open or closed in the non-alarm condition.
For example, a Field Display calibrated to display 0 to 10000, with a high alarm set at 9000 and hysteresis of 200 will perform as follows: The high alarm will be activated when increasing display equals 9000, but will not reset until the display falls below 8800. Michell Instruments...
I.S. Field Display User’s Manual OPERATION 3.3.9 Alarm Delay: ' dELA This function delays activation of the alarm output for an adjustable time following the alarm condition occurring. The delay can be set in 1 second increments between 0 and 3600 seconds.
0000 in the display mode by pressing simultaneously. Unless otherwise requested, new instruments with alarms are supplied with this function disabled and the security code set to 0000 Michell Instruments...
I.S. Field Display User’s Manual OPERATION 3.3.13 Adjusting Alarm Setpoints from the Display Mode NOTE: With the I.S. Field Display in the display mode, direct access to the ACSP alarm setpoints is only available when the menu is enabled, see Section 3.3.12.
4-20 mA input as shown in Figure 13, which increases the maximum Field Display voltage drop from 1.2 to 5 V. Hazardous Area Safe Area 9999 Easidew PRO I.S. Field Display Zener barrier or 3 4 6 5 Galvanic Isolator 4-20 mA...
The optional backlight may also be powered from a separate safe area power supply via an intrinsically safe interface, as shown in Figure 14 . Hazardous Area Safe Area 9999 Easidew PRO I.S. Field Display Zener barriers or Galvanic Isolators +24 V 3 4 6 5...
Pass the hose clips around the pipe to which the instrument is to be mounted, re-engage both clips and tighten after turning the mounting plate into the required position. Fix the instrument to the mounting plate using four locally sourced 4mm x 12mm long, self-tapping screws. Michell Instruments...
fi tted. A single spare instrument may quickly be recalibrated to replace any instrument that is damaged or fails. No attempt should be made to repair instruments at component level. It is recommended that faulty instruments are returned to Michell Instruments or to a local Michell Instrument's agent for repair. Routine Maintenance The mechanical condition of the instrument and electrical calibration should be regularly checked.
Page 35
I.S. Field Display User’s Manual APPENDIX A Appendix A Hazardous Area Certifi cation Michell Instruments...
APPENDIX A I.S. Field Display User’s Manual Appendix A Hazardous Area Certifi cation ATEX Certifi cation The I.S. Field Display is available with ATEX Gas & Dust Certifi cation. Dust certifi cation is an option which must be specifi ed when the I.S. Field Display is ordered. Hazardous Area Certifi...
Maintenance and servicing of the product must only be carried out by suitably trained personnel or returned to an approved Michell Instruments’ Service Center. Michell Instruments...
Page 38
I.S. Field Display User’s Manual APPENDIX A 97321 Issue 2.3, May 2016...
Page 39
I.S. Field Display User’s Manual APPENDIX A Michell Instruments...
Page 40
I.S. Field Display User’s Manual APPENDIX A 97321 Issue 2.3, May 2016...
Page 41
I.S. Field Display User’s Manual APPENDIX A Michell Instruments...
Page 42
I.S. Field Display User’s Manual APPENDIX A 97321 Issue 2.3, May 2016...
Page 43
I.S. Field Display User’s Manual APPENDIX A Michell Instruments...
Page 44
I.S. Field Display User’s Manual APPENDIX A 97321 Issue 2.3, May 2016...
Page 45
I.S. Field Display User’s Manual APPENDIX A Michell Instruments...
Page 46
I.S. Field Display User’s Manual APPENDIX A 97321 Issue 2.3, May 2016...
Page 47
I.S. Field Display User’s Manual APPENDIX A Michell Instruments...
Page 48
I.S. Field Display User’s Manual APPENDIX B Appendix B EC Declaration of Conformity 97321 Issue 2.3, May 2016...
The directives’ aim is to reduce the impact that electronic devices have on the environment. Michell Instruments is in full compliance with the WEEE Directive and is registered with an approved recycler (Registration No. WEE/JB0235YW) and treats the requirement of the directive and the protection of the environment with the utmost importance.
Equipment product sold into the EU market place as 22nd July 2017. However, the careful design policy of all Michell Instruments’ products continues to attain compliance in the shortest practical timescales and strives to ensure that less than 0.1% of total mass per product, of all non-compliant materials, appear within them.
Our continued review of the SVHC Candidate List and latest additions is to ensure we remain compliant. Michell Instruments maintains a hazardous material register in which MSDS data sheets are collated, and we will check that our suppliers will comply to REACH requirements for all materials and substances we use in the processes of our manufacturing.
I.S. Field Display User’s Manual APPENDIX C Return Policy If a Michell Instruments’ product malfunctions within the warranty period, the following procedure must be completed: Notify a Michell Instruments’ distributor, giving full details of the problem, the model variant and the serial number of the product.
Page 55
I.S. Field Display User’s Manual APPENDIX D Appendix D Return Document & Decontamination Declaration Michell Instruments...
Has the equipment been cleaned and decontaminated? NOT NECESSARY Michell Instruments will not accept instruments that have been exposed to toxins, radio-activity or bio-hazardous Work will not be carried out on any unit that does not have a completed decontamination declaration.
Page 57
I.S. Field Display User’s Manual NOTES: Michell Instruments...
Need help?
Do you have a question about the Easidew PRO and is the answer not in the manual?
Questions and answers