Labels and Descriptions About this Manual Indications of Use Contraindications About the Dabir System How it Works 2 Components of the Dabir Surfaces Overlay System Pneumatic Controller, Hose Assembly, and Accessories Overlays 3 Setting up the System Pneumatic Controller Pre-Assembly System Installation...
1.1 Important Before You Start • Please read entire manual carefully and completely before using. • As with any device, failure to carefully read and follow all instructions and safety information prior to use may result in decreased product performance or failure. 1.2 Safety Warnings •...
1.3 Labels and Descriptions The Symbols below appear on the Pneumatic Controller, Overlays, Hose Assembly, Accessories and/or packaging. Label Description Application UL Mark ANSI/AAMI ES60601-1 AMD 1 (2012), "Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance, Amendment 1.
Page 5
1.3 Labels and Descriptions - Continued The Symbols below appear on the Pneumatic Controller, Overlays, Hose Assembly, Accessories and/or packaging. Label Description Application Pneumatic Controller label, Box Follow instructions for use labels, Overlay labels, Hose Assembly label Defibrillation-proof Type BF Applied Part Pneumatic Controller label Indicates a defibrillation-proof type BF applied part complying with IEC 60601-1.
Page 6
1.3 Labels and Descriptions - Continued The Symbols below appear on the Pneumatic Controller, Overlays, Hose Assembly, Accessories and/or packaging. Label Description Application Serial Number Box labels, Pneumatic Controller Indicates the manufacturer’s serial label, Serial ID labels, Overlay labels number so that a specific medical device can be identified.
1.4 About this Manual This manual is your introduction to the Dabir Surfaces Patient Support System. Use it to initially set-up the DabirAIR Pneumatic Controller and Overlay. Keep it as a reference for day-to-day operation and as a guide for maintenance.
1.7 About the Dabir System The Dabir Surfaces Patient Support System is a low-profile, disposable, alternating pressure relief system with optional microclimate feature. The system consists of: • DabirAIR Pneumatic Control Unit - Easy Power Activation - Smart Interconnect System...
3.1 Pneumatic Controller Pre-Assembly Place Dabir Pneumatic Controller on a flat, stable surface, free and clear of loose articles which could block the air intake. CAUTION: Do not block air intake ports on the bottom of the Pneumatic Control Unit or the exhaust vent under the handle.
3.2 System Installation Overlay Installation Place Overlay flat onto patient support surface, mattress or pad with “THIS SIDE UP” facing up. The hose routing position is preferred to be at the foot end. CAUTION: Ensure there are no sharp objects between the Overlay and patient support surface, mattress or pad.
Page 13
3.2 System Installation - Continued Specific overlay usage details: DabirAIR Micropressure Overlays DabirAIR Overlays are intended to provide alternating support relief with reduced skin shear. Additional layers such as moisture barriers, incontinence pads, etc. may be used, but should be kept to a single layer only. Additional bulk or non-stretch layers placed between the covering bottom sheet and patient should be avoided.
Page 14
3.2 System Installation - Continued Connect Hose Assembly Connector (Female) to the Overlay Assembly Connector (Male). Orange Release Button Overlay Assembly Connector (Male) Hose Assembly Connector (Female) Hose / Overlay Connection: Connection Procedure With the system powered “OFF” or “STOPPED”, insert the Overlay Assembly Connector (Male) into the Hose Assembly Connector (Female)
4 Operating the Dabir System Powering Pneumatic Controller “ON” Plug the Pneumatic Controller Power Cord into an acceptable wall power outlet. CAUTION: To ensure proper device grounding, do not use an extension cord with this device. The Power Cord should be plugged directly into an acceptable wall power outlet.
5.2 Setting Overlay Firmness The Pneumatic Control Unit User Interface is equipped with three (3) different firmness settings: (Default Setting) MEDIUM HIGH Manually adjust comfort by touching the desired firmness option on the User Interface (see images below). NOTE: Microclimate Overlays will only have one firmness setting. The LOW, MEDIUM and HIGH User Interface buttons all provide a “LOW”...
6 Interrupting, Rapid Deflate and Resuming Operation Press the “START/STOP” key to pause or resume operation (see image below). NOTE: Unless unit is powered “OFF”, the Pneumatic Controller will remain at the previous settings when operation is resumed. To turn Pneumatic Controller off, toggle the main power switch to the “OFF” position. The Pneumatic Controller unit will return to default settings.
7 Information Center: Error Codes and Alarms The User Interface panel includes an “INFORMATION CENTER” to aid users in diagnosing potential problems (see image below). Medium High Medium High Medium High Medium High Information Information Center Center CYCLE SPEED FIRMNESS FIRMNESS CYCLE SPEED Information...
Page 20
7 Information Center: Error Codes and Alarms - Continued Information Center Buzzer Meaning Indicator Check Frequency: 2kHz The Pneumatic Controller is not properly communicating Connections with the Overlay. Check the Hose Assembly to Overlay 2sec ON connection. Light will clear when proper connection is Continuous after touching made and the unit will return to normal operation.
Dispatch™ Hospital Cleaner – Clorox, Virex 256™ - Diversey, Isopropyl Alcohol. Ensure there are no sharp objects between the Overlay and patient support surface, mattress or pad. 9 Routine Maintenance The Dabir Surfaces Patient Support System It is recommended that the Pneumatic Controller be returned to the manufacturer for annual maintenance and performance check after every 12 months of use.
Pneumatic Controller If fuse is blown - contact Dabir technical support for service (DO NOT REPLACE). How to reach us: Dabir Surfaces maintains regular business hours from 9:00 a.m. to 5:00 p.m. Eastern time. Technical and Warranty Support support@dabir-surfaces.com Tel: +1 (885) 478-0694...
11 Specification Sheets Controller Specifications Model: CA-1001 Supply Voltage: 100-240 VAC Supply Frequency: 50-60 Hz Power Input: 170VA Max Size: 356 x 178 x 127 mm (14in. x 7in. x 5 in.) Weight: 3.95 kg (8.7 lb) Case Material: Plastic Fuse Ratings: 2A, 5x20mm, Fast Acting Type of protection against electric shock:...
Page 24
11 Specification Sheets - Continued Accessory Specifications Replacement Hose Assembly Model: CA-9001 Weight: 0.73kg (1.6lb) Input: 5VDC 250mA RF Transmitter Separate collection for electronic waste Replacement Power Cord Model: CA-9002 Weight: 0.34 kg (0.75lb) Length: Wall Plug: NEMA 5-15P 3P Hospital Grade Green Dot Controller Plug: IEC 60320-C13 No Latex is used in the manufacture of this product.
Page 25
11 Specification Sheets - Continued Model CA-1001 Manufacturer’s Declaration – Electromagnetic Immunity The Model CA-1001 system is intended for use in the electromagnetic environment specified below. The customer or the user of the Model CA-1001 system should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level...
Page 26
11 Specification Sheets - Continued Model CA-1001 Manufacturer’s Declaration – Electromagnetic Immunity The Model CA-1001 system is intended for use in the electromagnetic environment specified below. The customer or the user of the Model CA-1001 system should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level...
Page 27
11 Specification Sheets - Continued Model CA-1001 Manufacturer’s Declaration – Electromagnetic Emissions The Model CA-1001 system is intended for use in the electromagnetic environment specified below. The customer or the user of the Model CA-1001 system should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment/Guidance...
Page 28
11 Specification Sheets - Continued Changes or modifications not expressly approved by Dabir Surfaces, Inc. could void the user's authority to operate the equipment. This device complies with Part 15 of the FCC Rules and with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause inter- ference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Surfaces, Inc. option, (1) to replace the Dabir Product, or (2) to refund the price paid, provided that the Dabir Product is returned to a location as specified by Dabir Surfaces, Inc. with a copy of the sales receipt, or dated itemized receipt, from an authorized reseller, and with a return authorization number provided by Dabir Surfaces, Inc.
Page 30
ER DIRECT OR INDIRECT) OR COMMERCIAL LOSS FOR BREACH OF ANY EXPRESS OR IMPLIED WARRANTY ON YOUR DABIR PRODUCT OR FOR ANY OTHER CLAIM, EVEN IF DABIR SURFACES, INC. HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. IN ADDITION, TO THE EXTENT PERMITTED BY LAW, THE MAXIMUM LIABILITY OF DABIR SURFACES, INC.
Revision: AA Print Date: MM/DD/YYYY Manufacturing 7447 West Wilson Ave Harwood Heights, IL 60706 USA Corporate Headquarters Dabir Surfaces, Inc. 24585 Evergreen Road Southfield, MI 48075 USA Tel: +1 (248) 796-0802 Fax: +1 (248) 354-6667 Sales sales@dabir-surfaces.com Technical and Warranty Support support@dabir-surfaces.com...