kenall MEDMASTER ENVELA MRSDL4L Series Installation Instructions

Regressed downlight patient room & surgical suite luminaires

Advertisement

Quick Links

MEDMASTER ENVELA
Patient Room & Surgical Suite Luminaires
MRSDL4L / M2SRDL4L SERIES
LED
THE ENVELA DOWNLIGHT CONTAINS SENSITIVE ELECTRONICS. TAKE CARE TO AVOID DAMAGE BY IMPROPER HANDLING OR STATIC
ELECTRICITY DISCHARGE (ESD). EITHER TYPE OF DAMAGE COULD RENDER THE SYSTEM INOPERABLE OR CAUSE LATENT FAILURE.

NEW CONSTRUCTION- DRYWALL CEILING

Note- 28L or 28T fixtures must be installed minimum
24" center to center, 12" from wall and have 1/2"
clearance above the trim.
1. Loosen locking screws to extend hanger bars. See
Figure 1.
2. Align bottom of hanger bar tabs to bottom of joist.
3. Secure luminaire hanger bars using nails or screws.
4. Position luminaire as required and lock position by
tightening locking screws.
5. Remove junction box cover.
6. Remove appropriate 1/2" knockout(s) and connect
to electrical service.
7. Wire per wiring diagram using approved wire
connectors. See Figure 2.
8. Replace junction box cover.
www.kenall.com | P: 800-4-Kenall | F: 262-891-9701 | 10200 55th Street Kenosha, Wisconsin 53144, USA
This product complies with the Buy American Act: manufactured in the United State swith more than 50% of the component cost of US origin It may be covered by patents
found at www.kenall.com/patents. Content of speci cation sheets is subject to change; please consult www.kenall.com for current product details. ©2019 Kenall Mfg. Co.
REGRESSED DOWNLIGHT
®

INSTALLATION INSTRUCTIONS

IMPORTANT SAFEGUARDS

When using electrical equipment, basic safety precautions
should always be followed, including the following:
THIS PRODUCT MUST BE INSTALLED IN ACCORDANCE
WITH THE APPLICABLE INSTALLATION CODE BY A PERSON
FAMILIAR WITH THE CONSTRUCTION AND OPERATION OF
THE PRODUCT AND THE HAZARDS INVOLVED.
CE PRODUIT DOIT ÊTRE INSTALLÉ SELON LE CODE
D'INSTALLATION PERTINENT, PAR UNE PERSONNE QUI
CONNAÎT BIEN LE PRODUIT ET SON FONCTIONNEMENT
AINSI QUE LES RISQUES INHÉRENTS.
DISCONNECT POWER TO ALL CIRCUITS BEFORE WIRING FIXTURE.
INSTALL IN ACCORDANCE WITH ALL NATIONAL, STATE, AND
LOCAL CODES. DO NOT CONNECT TO AN UNGROUNDED
SUPPLY. READ ALL FIXTURE MARKINGS AND LABELS TO
ENSURE CORRECT INSTALLATION OF FIXTURE. SUPPLEMENTAL
INSTRUCTIONS MAY BE LOCATED ON THE FIXTURE, IN ADDITION
TO THIS INSTRUCTION SHEET, REGARDING ORIENTATION, OR
MOUNTING RESTRICTIONS.

SAVE THESE INSTRUCTIONS

– RECOMMENDED CEILING CUTOUT- 4.125" SQUARE
LOCKING SCREW
ALIGN TO BOTTOM OF JOIST
CONNECT (+)10V FROM DIMMER TO VIOLET WIRE (OPT)
CONNECT (-)10V FROM DIMMER TO GREY WIRE (OPT)
CONNECT HOT SUPPLY WIRE (120-277V) TO BLACK WIRE
CONNECT COMMON SUPPLY WIRE TO WHITE WIRE
CONNECT GROUND WIRE TO GREEN
AND COPPER GROUND WIRES
FIGURE 2
FIGURE 1
LED
DRIVER
VIOLET
GREY
BLACK
WHITE
GREEN
COPPER GROUND WIRE
F-4826-090418
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MEDMASTER ENVELA MRSDL4L Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for kenall MEDMASTER ENVELA MRSDL4L Series

  • Page 1: Installation Instructions

    This product complies with the Buy American Act: manufactured in the United State swith more than 50% of the component cost of US origin It may be covered by patents found at www.kenall.com/patents. Content of speci cation sheets is subject to change; please consult www.kenall.com for current product details. ©2019 Kenall Mfg. Co.
  • Page 2 This product complies with the Buy American Act: manufactured in the United State swith more than 50% of the component cost of US origin It may be covered by patents found at www.kenall.com/patents. Content of speci cation sheets is subject to change; please consult www.kenall.com for current product details. ©2019 Kenall Mfg. Co.
  • Page 3 This product complies with the Buy American Act: manufactured in the United State swith more than 50% of the component cost of US origin It may be covered by patents found at www.kenall.com/patents. Content of speci cation sheets is subject to change; please consult www.kenall.com for current product details. ©2019 Kenall Mfg. Co.
  • Page 4 This product complies with the Buy American Act: manufactured in the United State swith more than 50% of the component cost of US origin It may be covered by patents found at www.kenall.com/patents. Content of speci cation sheets is subject to change; please consult www.kenall.com for current product details. ©2019 Kenall Mfg. Co.

This manual is also suitable for:

Medmaster envela m2srdl4l series

Table of Contents