Before use Electrical installation Before connection please read and Thank you for purchasing this Stuart product. To get the best performance from understand this manual and ensure that the equipment, and for your own safety, the line supply is suitable. This water bath requires a supply rated at 230V, 50 or please read these instructions carefully before use.
Page 5
Safety advice before use RE300DB ❖ Before use please read and understand this Fill the Teflon coated bowl to the required level manual and familiarise yourself with the with cold water. To avoid scale build up and the layout of the controls – see figure 1. necessity for cleaning it is recommended that the bowl be filled with distilled or deionised water.
Cleaning The bath may contain water or oil (RE300OB only) at very high temperatures. No attempt Where used with water periodically de-scale the to move the bath should be made when it is hot. bath using a domestic kettle descaler suitable for metal kettles.
Warranty Bibby Scientific Ltd warrants this instrument to be free from defects in material and workmanship, when used under normal laboratory conditions, for a period of three (3) years. In the event of a justified claim, Bibby Scientific Ltd will replace any defective component or replace the unit free of charge.
Fault finding Problem Cause Solution Display does not work even No mains power Check mains power supply if the power is switched on Fuse has blown Replace the fuse, if fuse blows again immediately stop operation and consult a local service representative.
Page 10
This product meets the applicable EC and so we cannot guarantee that interference will not occur in practice. Where there is a possibility that injury, harmonized standards for radio frequency interference and may be expected not to damage or loss might occur if equipment malfunctions interfere with, or be affected by, other equipment with due to radio frequency interference, or for general advice before use, contact the Technical Department of Bibby...
INSPECTION REPORT ❏ ❏ MODEL RE300DB RE300OB ELECTRICAL SAFETY ❏ ✓ 1. Earth continuity ❏ ✓ 2. Insulation ❏ ✓ 3. Flash test FUNCTIONAL ❏ ✓ 1. Indicators ❏ ✓ 2. Temperature control ❏ ✓ 3. Visual acceptance QUALITY CONTROL INSPECTOR Bibby Scientific Italia Srl Bibby Scientific France SAS ZI du Rocher Vert - BP 79...