Download Print this page

Advertisement

Quick Links

KELLY STAND
LOCO - 124
USER MANUAL
Page1
ENSURE THESE INSTRUCTIONS ARE FULLY READ AND
UNDERSTOOD BEFORE USE.
PLEASE KEEP IN A SAFE PLACE FOR FUTURE
REFERENCE.
HEALTH AND SOCIAL SERVICES
THESE INSTRUCTIONS SHOULD BE LEFT WITH THE USER
OF THE LOCOMOTOR KELLY STANDTO CONSULT AS
REQUIRED.
CONTENTS
PACKING LIST
SPARE PART LIST
Always charge battery when not in
use.
BEFORE USE READ 1.3 NOW
Northants
NN10 0BU Tel:
01933 411729 Fax: 01933 410213
Designed and Developed by: Proto Magic Innovations Ltd Page 1 ISSUE 4
KELLY STAND INSTRUCTIONS 07 23/02/2010

Advertisement

loading

Summary of Contents for Locomotor Kelly Stand 124

  • Page 1: Table Of Contents

    UNDERSTOOD BEFORE USE. PLEASE KEEP IN A SAFE PLACE FOR FUTURE REFERENCE. HEALTH AND SOCIAL SERVICES THESE INSTRUCTIONS SHOULD BE LEFT WITH THE USER OF THE LOCOMOTOR KELLY STANDTO CONSULT AS REQUIRED. CONTENTS PACKING LIST TECHNICAL SPECIFICATION APPROVED SLING LIST...
  • Page 2 * The TEST CERTIFICATE is valid for ONE year and can be used for your insurance records. Once the first year has expired, the Locomotor KELLY STAND will require a new inspection and re-testing for the next year; you can arrange this through your supplier. In cases of any difficulty, please contact the manufacturer directly for the name and telephone number of your nearest service dealer.
  • Page 3 KELLY STAND LOCO - 124 USER MANUAL Page3 Fitting and adjustment of Handle bars and Shin pad Remove by unscrewing the handle bar lock handwheels Refit handle bars in this position and fully tighten bar lock handwheels Handle bars are fitted like this for transit do not attempt to use the Kelly stand with the handle bars in this position...
  • Page 4 KELLY STAND LOCO - 124 USER MANUAL Page4 ELECTRICAL REQUIREMENTS The following details are required for the battery charger that is supplied with the KELLY STAND:- FOR APPLIANCES SUPPLIED WITH FITTED PLUG. If the socket outlets in your home are not suitable for the plug supplied with your battery charger, it should be removed and the appropriate plug fitted.
  • Page 5 (9) will indicate that battery is charging. When the battery is fully charged the amber light will go out. Assembly of your Locomotor KELLY STAND (See diagram) 2.1. Place the base chassis (4) on a level floor. 2.2. Locate the mast assembly (5) onto the chassis spigot (10), then lock with the mast lock hand wheel (11).
  • Page 6 3.1. Leg Opening. To open the legs of the Locomotor KELLY STAND press the leg open button (21) on the hand control (3) or (22) on the battery compartment(28). The legs can be set in any position depending on how long the button is held for. To close the legs press the leg close button (23) on the hand control (3) or (24) on the battery compartment(28).
  • Page 7 USER MANUAL Page7 3.5. Moving the Locomotor KELLY STAND. Always make sure the brakes are off, when moving the KELLY STAND. Try to pull the KELLY STAND (rather than push), either in a circular motion by pulling sideways on one handle, or backwards by pulling on both handles (37).
  • Page 8 6.1. DO ensure that you have read these instructions and understood them fully, including the glossary of symbols in the back of the instruction manual. 6.2. DO ensure that you carry out an inspection of the Locomotor KELLY STAND on a regular basis.
  • Page 9 KELLY STAND LOCO - 124 USER MANUAL Page9 Trouble Shooting. IF KELLY STAND FAILS TO LIFT OR LOWER CHECK THE FOLLOWING: YES NO a) Is the lift actuator's (18) power plug (35) plugged into the lift actuator power socket (6) ? If NO plug in power plug (35) to connect lift actuator (18). If YES GO TO (b). b) Has emergency stop switch been activated? If YES pull out emergency switch (2) to reset.
  • Page 10 KELLY STAND LOCO - 124 USER MANUAL Page10 IF KELLY STANDLEG OPENING WILL NOT WORK CHECK THE FOLLOWING YES NO a)Is the leg actuator power plug (12) plugged into the leg actuator power socket on battery compartment (28) ? If NO plug in power plug (12) to connect leg actuator. GO TO (b). b) Make sure legs are not obstructed GO TO (c).
  • Page 11 KELLY STAND LOCO - 124 USER MANUAL Page11 CHARGER WILL NOT CHARGE a) Check green power on lamp (7) is on. If NO replace fuse in mains plug with 3 amp fuse. If YES GO TO (c), if NO GO TO (b). b) Check mains socket is switched on then try another mains socket.
  • Page 12 KELLY STAND LOCO - 124 USER MANUAL Page12 MAINTENANCE SCHEDULE Check the sling for any damage or wear, paying particular attention SLINGS * to the lifting straps, and the integrity of the stitching. MAST ASSEMBLY (5) * Check for any damage or distortion in the handlebar, also check the welded joints for any wear or tear.
  • Page 13: Technical Specification

    KELLY STAND LOCO - 124 USER MANUAL Page13 TECHNICAL SPECIFICATION STANDARDS BS5724 Class 2 (Electrical Components) SAFE WORKING LOAD 203Kg (32 Stones) TESTED CAPACITY 220 Kg MINIMUM INTERNAL WIDTH 554mm 21.8" MINIMUM EXTERNAL WIDTH 605mm 23.8" MAXIMUM INTERNAL WIDTH 900mm 35.4"...
  • Page 14: Approved Sling List

    KELLY STAND LOCO - 124 USER MANUAL Page14 APPROVED SLING LIST Issue 1 Product Code Description Size Loco-055 Locomotor Stander Sling Small Loco-056 Locomotor Stander Sling Medium Loco-057 Locomotor Stander Sling Large Loco-058 Locomotor Stander Sling Extra Large APPROVED SPARE PARTS LIST...
  • Page 15 KELLY STAND LOCO - 124 USER MANUAL Page15 Diagram Select Healthcare (UK) Ltd Unit 2 Albert Road Rushden Northants NN10 0BU Tel: 01933 411729 Fax: 01933 410213 Designed and Developed by: Proto Magic Innovations Ltd Page 15 KELLY STAND INSTRUCTIONS 07...
  • Page 16 KELLY STAND LOCO - 124 USER MANUAL Page16 GLOSSARY Attention consult accompanying Full battery. documents before use. Refer servicing to Low battery. servicing personnel only. Recyclable product. Flat battery. Overload - check Lead acid battery. weight and make sure sling is not trapped. Do not dispose Hand controller of in the bin.
  • Page 17 KELLY STAND LOCO - 124 USER MANUAL Page17 GLOSSARY If sounder bleeps when lifting please change battery. Emergency stop. Emergency manual lowering release lever. Type B equipment (Protection against electric shock). Select Healthcare (UK) Ltd Unit 2 Albert Road Rushden Northants NN10 0BU Tel: 01933 411729 Fax: 01933 410213...