Advertisement

Quick Links

Content
Classification............................................................................................ 2
Symbol ..................................................................................................... 2
Precaution................................................................................................ 3
Instrument Components .......................................................................... 4
Theory and Function................................................................................ 4
Characteristics ......................................................................................... 5
Environment Requirement ....................................................................... 6
Install and Connection ............................................................................. 7
Introduction of the Interface and Setting ................................................. 8
1. Interface Introduction.................................................................... 8
2. Setting........................................................................................... 9
2.1 Input ID and Setting Parameters................................................ 9
2.2 Input the Name of Patient......................................................... 10
2.3 Eye mode.................................................................................. 10
2.4 Left/right eye setting ................................................................. 10
Measure ..................................................................................................11
1. Preparation ..................................................................................11
2. Auto Mode................................................................................... 13
3. Manual Mode.............................................................................. 14
4. IOL Calculation ........................................................................... 14
5. Print ............................................................................................ 15
6. Case Review .............................................................................. 15
7. System Criteria Testing............................................................... 15
Change the Print Paper ......................................................................... 16
Package and Transportation.................................................................. 17
Daily Maintenance ................................................................................. 18
Trouble Shooting.................................................................................... 19
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SW-1000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Suder SW-1000

  • Page 1: Table Of Contents

    Content Classification.................... 2 Symbol ..................... 2 Precaution....................3 Instrument Components ................4 Theory and Function................4 Characteristics ..................5 Environment Requirement ............... 6 Install and Connection ................7 Introduction of the Interface and Setting ..........8 1. Interface Introduction..............8 2.
  • Page 2: Classification

    Introduction Dear User: SW-1000 is a precise ultrasound measure instrument, designed for detecting axial length, depth of anterior chamber, thickness of lens and calculating the IOL. This User’s Guide is designed as a training and reference manual. We recommend you carefully read and follow the steps in this guide to ensure that the machine would have optimum performance.
  • Page 3: Precaution

    Precaution This instrument requires the power supply of AC 90~230V , 50HZ±2%. The power socket indoors should have ground line. If the local power net could not fit the requirement, then we recommend you to use regulated power supply. Improper power supply may cause instrument damage or person injury.
  • Page 4: Instrument Components

    Instrument Components Touch Pen Gain Brightness Foot switch Buzzer Probe Printer Pic.1 Front View of SW-1000 Power switch A Probe socket Foot switch socket Power cable socket Pic.2 Back view of SW-1000...
  • Page 5: Theory And Function

    Theory and Function Ophthalmic A scan SW-1000 is composed of main unit, probe, pedal and so on. The probe contacts the cornea and sends out ultrasound of 10MHz. The echo wave from front surface, back surface of lens and retina is detected. The signal is processed and then we get the depth of chamber, thick of lens and axial length.
  • Page 6: Environment Requirement

    Output:IOL calculating table; Measure:Anterior chamber, lens, vitreous body and total length. Curve freezing:Manual/Auto mode, controlled by pedal; Output:Build-in fast thermal printer; Display:Color liquid-crystal screen; Input: touch screen. Power: 30W Environment Requirement 1. Work Environment Ambient temperature:+5℃ ~ +40℃ Relative moisture:≤ 80% Power Supply:AC 90~230V,50×(1±2%)Hz Keep away from strong magnetic or electronic field.
  • Page 7: Install And Connection

    Install and Connection 1. Check the package to see if anything missing. (Reference to Appendix: package list) 2. Insert probe, pedal and power plug to respective socket. (See Pic.3) Align the salient on the probe connector to the gap on the probe Pic.3 Pull and plug the probe socket, and plug it in.
  • Page 8: Introduction Of The Interface And Setting

    Introduction of the Interface and Setting 1. Interface Introduction Setting Menu Left or right Eye Type Gain Time Threshold Measurement Data Reading Window Scale Mode Pic.4 measure interface After switch on, “Measure” mode and “Contact” is default. See Pic.4. Press ”Contact” to switch to “Immersion”. On the upper part is the setting menu.
  • Page 9: Setting

    farsighted ; while DR = -1.00 means the recovery eyesight is expected to be 100 nearsighted. 1.2 On the right is reading window displaying the result of measurement. Readings:the eye axial length of each measure. 5 groups can be displayed. AC:Depth of anterior chamber L:Thickness of Lens V:Thickness of vitreous body...
  • Page 10: Input The Name Of Patient

    The touch screen is very sensitive. In order to prevent damage of touch screen, don’t press it with much strength. If you want to re-enter the items, press other item or press ”Enter”, then press the item you want to correct. Now you can reset the item. Use Caps and Shift to change between Capitalization and lowercase.
  • Page 11: Measure

    LENS VITREOUS THICKNESS NORMAL 1641 1532 1532 DENSE 1629 1532 1532 APHAKIC 1532 PMMA 2718 1532 1532 SILICONE 1049 1532 1532 ACRYLIC 2200 1532 1532 Normal & High: It’s easier to take a reading under ”Normal”, but with less precision, “High” can get more precise reading, but more difficult. AVG only can be used under eye mode APHAKIC Thickness only used under eye mode PMMA, SILICONE, ACEYLIC Measure...
  • Page 12 Cornea A scan probe Cornea A scan probe Proper contact Improper Contact Immersion: 1. In order to release the pain in the eye, the cornea needs anaesthesia. 2. Choose suitable size eye cup and Put it in the eye lid. 3.
  • Page 13: Auto Mode

    2. Auto Mode Click ‘Measure’ below after setting the parameters to choose measure mode. When ‘AUTO’ is displayed lower-right corner of the screen, it shows that it is under AUTO mode. (see Pic.7 Auto mode pic.7) Make preparation as above-mentioned. The instrument will catch and freeze the waveform automatically, and beep to indicate the completion of a measure.
  • Page 14: Manual Mode

    the convenience of next measure. 3. Manual Mode In measure mode, press the “auto” to switch to manual mode. And “manual” will be shown on the screen.(See pic.8) Preparations before measurement are the same as AUTO mode. Put the probe perpendicular to the Pic.8 Manual Mode center of patient’s cornea.
  • Page 15: Print

    value between the * is most close to the DR setting value. Click AC on the left upper part, use small keyboard to revise the value. It can be used only in the formula HAIGIS-STD. Click the formula by touch pen to apply it. And the value will change accordingly.
  • Page 16: Change The Print Paper

    Fill the test block with distilled water ( No bubble in the water) Eye Type: Aphakic Eye Speed: 1520 2. Gain: adjust the gain to a lower lever, around 50 Put the probe straight against first step, on the edge, move the probe slightly, make sure the ultrasound can get reflection from all the 4 steps.
  • Page 17: Package And Transportation

    Pic.14 Press S/L Pic.15 Refill roller, see paper end Pic.16 Press S/L to roller paper Pic.17 Close the cover Package and Transportation The packed instrument can be transported by transportation vehicles. It must not be transported together with goods that are easy to burn, explode, or corrode.
  • Page 18: Daily Maintenance

    Daily Maintenance 1. In moist season, if the machine hasn’t been put into use for a long time, it should be powered up for about one hour each month to drive the moisture away. 2. When the ambient temperature is under 10℃, the machine should be warmed up for 5 minute before measuring.
  • Page 19: Trouble Shooting

    Trouble Shooting PHENOMENA REASONS METHODS Failure of Re-insert the plug to Bad connection of lightening connect the power plug the power indicator it well with the socket Choose the mode The mode of the instrument again to make the is not correct. instrument under Failure of freezing measuring mode.